Measurement of oxygen saturation in the retina with a spectroscopic sensitive multi aperture camera

被引:42
作者
Ramella-Roman, Jessica C. [1 ]
Mathews, Scott A. [1 ]
Kandimalla, Haryipria [1 ]
Nabili, Afshin [1 ]
Duncan, Donald D. [2 ]
D'Anna, Salvatore A. [3 ]
Shah, Syed Mahmood [3 ]
Nguyen, Quan Dong [3 ]
机构
[1] Catholic Univ Amer, Washington, DC 20064 USA
[2] Oregon Hlth & Sci Univ, Portland, OR 97201 USA
[3] Johns Hopkins Univ, Wilmer Eye Inst, Baltimore, MD 21218 USA
关键词
D O I
10.1364/OE.16.006170
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We introduce a new multi aperture system capable of capturing six identical images of the human fundus at six different spectral bands. The system is based on a lenslet array architecture and is well suited for spectroscopy application. The multi-aperture system was interfaced with a fundus camera to acquire spectroscopic sensitive images of the retina vessel and ultimately to calculate oxygen saturation in the retina in vivo. In vitro testing showed that the system is able to accurately reconstruct curves of partially oxygenated hemoglobin. In vivo testing on healthy volunteers was conducted and yielded results of oxygen saturation similar to the one reported in the literature, with arterial SO2 similar to 0.95 and venous SO2 similar to 0.5. (c) 2008 Optical Society of America.
引用
收藏
页码:6170 / 6182
页数:13
相关论文
共 25 条
  • [1] NONINVASIVE TECHNIQUE FOR OXIMETRY OF BLOOD IN RETINAL-VESSELS
    DELORI, FC
    [J]. APPLIED OPTICS, 1988, 27 (06) : 1113 - 1125
  • [2] Retinal venous oxygen saturation and cardiac output during controlled hemorrhage and resuscitation
    Denninghoff, KR
    Smith, MH
    Lompado, A
    Hillman, LW
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 2003, 94 (03) : 891 - 896
  • [3] Retinal large vessel oxygen saturations correlate with early blood loss and hypoxia in anesthetized swine
    Denninghoff, KR
    Smith, MH
    Chipman, RA
    Hillman, LW
    Jester, PM
    Hughes, CE
    Kuhn, F
    Rue, LW
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1997, 43 (01) : 29 - 34
  • [4] An instrument for che measurement of retinal vessel oxygen saturation
    Drewes, J
    Smith, M
    Denninghoff, K
    Hillman, L
    [J]. OPHTHALMIC TECHNOLOGIES IX, PROCEEDINGS OF, 1999, 3591 : 114 - 120
  • [5] DREWES JJ, THESIS U ALABAMA HUN
  • [6] OPTICAL-PROPERTIES OF OCULAR FUNDUS TISSUES - AN IN-VITRO STUDY USING THE DOUBLE-INTEGRATING-SPHERE TECHNIQUE AND INVERSE MONTE-CARLO SIMULATION
    HAMMER, M
    ROGGAN, A
    SCHWEITZER, D
    MULLER, G
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (06) : 963 - 978
  • [7] A review of methods for human retinal oximetry
    Harris, A
    Dinn, RB
    Kagemann, L
    Rechtman, E
    [J]. OPHTHALMIC SURGERY LASERS & IMAGING, 2003, 34 (02) : 152 - 164
  • [8] A STUDY OF RETINAL VENOUS BLOOD OXYGEN SATURATION IN HUMAN SUBJECTS BY PHOTOGRAPHIC MEANS
    HICKAM, JB
    FRAYSER, R
    ROSS, JC
    [J]. CIRCULATION, 1963, 27 (03) : 375 - &
  • [9] Snapshot hyperspectral imaging in ophthalmology
    Johnson, William R.
    Wilson, Daniel W.
    Fink, Wolfgang
    Humayun, Mark
    Bearman, Greg
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2007, 12 (01)
  • [10] ROLE OF BLOOD-FLOW AND IMPAIRED AUTOREGULATION IN THE PATHOGENESIS OF DIABETIC-RETINOPATHY
    KOHNER, EM
    PATEL, V
    RASSAM, SMB
    [J]. DIABETES, 1995, 44 (06) : 603 - 607