Ceramic Bearings for Total Hip Arthroplasty Have High Survivorship at 10 Years

被引:104
作者
D'Antonio, James A. [1 ]
Capello, William N. [2 ]
Naughton, Marybeth [3 ]
机构
[1] Greater Pittsburgh Orthopaed Assoc, Moon Township, PA 15108 USA
[2] Indiana Univ Sch Med, Indianapolis, IN USA
[3] Stryker Orthopaed, Mahwah, NJ USA
关键词
CROSS-LINKED POLYETHYLENE; FEMORAL COMPONENTS; FOLLOW-UP; WEAR; SQUEAKING; IMPLANTS; DISLOCATION; OSTEOLYSIS; FRACTURES; HISTORY;
D O I
10.1007/s11999-011-2076-7
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Ceramic bearings were introduced to reduce wear and increase long-term survivorship of total hip arthroplasty. In a previous study comparing ceramic with metal-on-polyethylene at 5 to 8 years, we found higher survivorship and no osteolysis for the ceramic bearings. We asked whether ceramic bearings have equal or superior survivorship compared with that for metal-on-polyethylene at longer followup; we also determined survivorship of the implant systems, the presence or absence of radiographic osteolysis, and incidence of device squeaking. Five surgeons at five sites have followed 189 patients (216 hips) for a minimum of 10 years and average of 10.3 years (range, 10-12.4 years) comparing alumina ceramic bearings (144 hips) with cobalt chrome-on-polyethylene bearings (72 hips). We determined Kaplan-Meier survivorship of the bearing surface and implant systems and collected radiographic and clinical data. We observed no difference between the control metal-on-polyethylene and the alumina-bearing couple cohorts with regard to bearing-related failures (98.9% versus 99.1%). Revisions for any reason occurred in 10.5% of the control patients and 3.1% of the patients with alumina bearings. All femoral implants remain well fixed (100%), whereas one acetabular component (1%) is unstable in the control group. Osteolysis occurred in 26% of the control patients and in none of the patients with alumina bearings. Squeaking occurred in two of 144 hips (1.4%) of the patients with ceramic bearings. Patients receiving the ceramic-on-ceramic bearings had fewer revisions for any reason and less osteolysis than the control metal-on-polyethylene at 10 years. Our data suggest ceramic bearings continue to provide an option for the young and more active patient and provide for a measure to compare other new alternative bearings that are currently available. Level I, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
引用
收藏
页码:373 / 381
页数:9
相关论文
共 38 条
[1]   Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty [J].
Berry, DJ ;
Von Knoch, M ;
Schleck, CD ;
Harmsen, WS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (11) :2456-2463
[2]   LONG-TERM RESULTS OF UNCEMENTED ALUMINA ACETABULAR IMPLANTS [J].
BOEHLER, M ;
KNAHR, K ;
PLENK, H ;
WALTER, A ;
SALZER, M ;
SCHREIBER, V .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (01) :53-59
[3]   Ceramic-on-Ceramic Total Hip Arthroplasty: Update [J].
Capello, William N. ;
D'Antonio, James A. ;
Feinberg, Judy R. ;
Manley, Michael T. ;
Naughton, Marybetb .
JOURNAL OF ARTHROPLASTY, 2008, 23 (07) :39-43
[4]   Continued Improved Wear with an Annealed Highly Cross-linked Polyethylene [J].
Capello, William N. ;
D'Antonio, James A. ;
Ramakrishnan, Rama ;
Naughton, Marybeth .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (03) :825-830
[5]  
Chevillotte C, 2010, INT ORTHOP 1221
[6]   The 2009 Frank Stinchfield Award "Hip Squeaking'' A Biomechanical Study of Ceramic-on-ceramic Bearing Surfaces [J].
Chevillotte, Christophe ;
Trousdale, Robert T. ;
Chen, Qingshan ;
Guyen, Olivier ;
An, Kai-Nan .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (02) :345-350
[7]  
CLARKE IC, 1992, CLIN ORTHOP RELAT R, P19
[8]   Alumina ceramic bearings for total hip arthroplasty - Five-year results of a prospective randomized study [J].
D'Antonio, J ;
Capello, W ;
Manley, M ;
Naughton, M ;
Sutton, K .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (436) :164-171
[9]   Five-year experience with Crossfire® highly cross-linked polyethylene [J].
D'Antonio, JA ;
Manley, MT ;
Capello, WN ;
Bierbaum, BE ;
Ramakrishnan, R ;
Naughton, M ;
Sutton, K .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (441) :143-150
[10]  
DELEE JG, 1976, CLIN ORTHOP RELAT R, P20