ChromAIX: A high-rate energy-resolving photon-counting ASIC for Spectral Computed Tomography

被引:15
作者
Steadman, Roger [1 ]
Herrmann, Christoph [1 ]
Muelhens, Oliver [1 ]
Maeding, Dale G. [2 ]
Colley, James [3 ]
Firlit, Ted [4 ]
Luhta, Randy [5 ]
Chappo, Marc [5 ]
Harwood, Brian [5 ]
Kosty, Doug [4 ]
机构
[1] Philips Res Europe, Weisshausstr 2, D-52066 Aachen, Germany
[2] Innovat Design, Dana Point, CA 92629 USA
[3] Zera Engn PC, Albany, OR 97321 USA
[4] Aeroflex, Colorado Springs, CO 80907 USA
[5] Philips Healthcare, Cleveland, OH 44143 USA
来源
MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING | 2010年 / 7622卷
关键词
Photon counting; Computed Tomography; Spectral CT; K-edge imaging; CZT;
D O I
10.1117/12.844222
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
In Computed Tomography applications a major opportunity has been identified in the exploitation of the spectral information inherently available due to the polychromatic emission of the X-ray tube. Current CT technology based on indirect-conversion and integrating-mode detection can be used to some extent to distinguish the two predominant physical causes of energy-dependent attenuation (photo-electric effect and Compton effect) by using dual-energy techniques, e. g. kVp switching, dual-source or detector stacking. Further improvements can be achieved by transitioning to direct-conversion technologies and counting-mode detection, which inherently exhibits a better signal-to-noise ratio. Further including energy discrimination, enables new applications, which are not feasible with dual-energy techniques, e. g. the possibility to discriminate K-edge features (contrast agents, e. g. Gadolinium) from the other contributions to the x-ray attenuation of a human body. The capability of providing energy-resolved information with more than two different measurements is referred to as Spectral CT. To study the feasibility of Spectral CT, an energy-resolving proprietary photon counting ASIC (ChromAIX) has been designed to provide high count-rate capabilities while offering energy discrimination. The ChromAIX ASIC consists of an arrangement of 4 by 16 pixels with an isotropic pitch of 300 mu m. Each pixel contains a number of independent energy discriminators with their corresponding 12-bit counters with continuous read-out capability. Observed Poissonian countrates exceeding 10 Mcps (corresponding to approximately 27 Mcps incident mean Poisson rate) have been experimentally validated through electrical characterization. The measured noise of 2.6 mV(RMS) (4 keV FWHM) adheres to specifications. The ChromAIX ASIC has been specifically designed to support direct-converting materials CdZnTe and CdTe.
引用
收藏
页数:8
相关论文
共 13 条
[1]  
[Anonymous], IEEE NSS MIC RTSD 20
[2]   Application of PILATUS II Detector Modules for High Resolution X-Ray Imaging Crystal Spectrometers on the Alcator C-Mod Tokamak [J].
Bitter, M. L. ;
Broennimann, Ch. ;
Eikenberry, E. F. ;
Hill, K. W. ;
Ince-Cushman, A. ;
Lee, S. G. ;
Rice, J. E. ;
Scott, S. .
2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, :3721-3723
[3]  
CARMI R, 2005, NUCL SCI S IEEE PISC
[4]   First performance evaluation of a dual-source CT (DSCT) system [J].
Flohr, TG ;
McCollough, CH ;
Bruder, H ;
Petersilka, M ;
Gruber, K ;
Süss, C ;
Grasruck, M ;
Stierstorfer, K ;
Krauss, B ;
Raupach, R ;
Primak, AN ;
Küttner, A ;
Achenbach, S ;
Becker, C ;
Kopp, A ;
Ohnesorge, BM .
EUROPEAN RADIOLOGY, 2006, 16 (02) :256-268
[5]   Photon counting energy dispersive detector arrays for x-ray imaging [J].
Iwanczyk, J. S. ;
Nygard, E. ;
Meirav, O. ;
Arenson, J. ;
Barber, W. C. ;
Hartsough, N. E. ;
Malakhov, N. ;
Wessel, J. C. .
2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, :2741-+
[6]  
Knoll G.F., 1989, Radiation Detection and Measurement, V2nd
[7]   Counting and integrating readout for direct conversion X-ray imaging: Concept, realization and first prototype measurements [J].
Kraft, Edgar ;
Fischer, Peter ;
Karagounis, Michael ;
Koch, Manuel ;
Krueger, Hans ;
Peric, Ivan ;
Wermes, Norbert ;
Herrmann, Christoph ;
Nascetti, Augusto ;
Overdick, Michael ;
Ruetten, Walter .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2007, 54 (02) :383-390
[8]  
OVERDICK M, IEEE T NUCL IN PRESS
[9]   First results of XPAD3, a new photon counting chip for X-ray CT-scanner with energy discrimination [J].
Pangaud, P. ;
Basolo, S. ;
Chantepie, B. ;
Clemens, J. C. ;
Delpierre, P. ;
Dinkespiler, B. ;
Menouni, M. ;
Bonissent, A. ;
Debarbieux, F. ;
Morel, C. .
2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, :14-18
[10]   K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors [J].
Roessl, E. ;
Proksa, R. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (15) :4679-4696