Software-based noise reduction in cranial magnetic resonance imaging: Influence on image quality

被引:7
作者
Fuelkell, Philipp [1 ]
Langner, Soenke [2 ]
Friedrich, Nele [3 ]
Kromrey, Marie-Luise [1 ]
Radosa, Christoph G. [4 ]
Platzek, Ivan [4 ]
Mensel, Birger [1 ]
Kuehn, Jens-Peter [1 ,4 ]
机构
[1] Univ Med Greifswald, Dept Diagnost Radiol & Neuroradiol, Greifswald, Germany
[2] Univ Med Rostock, Dept Radiol, Rostock, Germany
[3] Univ Med Greifswald, Dept Lab Med, Greifswald, Germany
[4] Carl Gustav Carus Univ, Clin & Policlin Radiol & Intervent Radiol, Dresden, Germany
关键词
ACOUSTIC-NOISE; HEARING-LOSS; MRI; SCAN;
D O I
10.1371/journal.pone.0206196
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Objectives To investigate acoustic noise reduction, image quality and white matter lesion detection rates of cranial magnetic resonance imaging (MRI) scans acquired with and without sequence-based acoustic noise reduction software. Material and methods Thirty-one patients, including 18 men and 13 women, with a mean age of 58.3 +/- 14.5 years underwent cranial MRI. A fluid-attenuated inversion recovery (FLAIR) sequence was acquired with and without acoustic noise reduction using the Quiet Suite (QS) software (Siemens Healthcare). During data acquisition, peak sound pressure levels were measured with a sound level meter (Testo, Typ 815). In addition, two observers assessed subjective image quality for both sequences using a five-point scale (1 very good-5 inadequate). Signal-to-noise ratio (SNR) was measured for both sequences in the following regions: white matter, gray matter, and cerebrospinal fluid. Furthermore, lesion detection rates in white matter pathologies were evaluated by two observers for both sequences. Acoustic noise, image quality including SNR and white matter lesion detection rates were compared using the Mann-Whitney-U-test. Results Peak sound pressure levels were slightly but significantly reduced using QS, P <= 0.017. Effective sound pressure, measured in Pascal, was decreased by 19.7%. There was no significant difference in subjective image quality between FLAIR sequences acquired without/with QS: observer 1: 2.03/2.07, P = 0.730; observer 2: 1.98/2.10, P = 0.362. In addition, SNR was significantly increased in white matter, P <= 0.001, and gray matter, P = 0.006, using QS. The lesion detection rates did not decline utilizing QS: observer 1: P = 0.944 observer 2: P = 0.952. Conclusions Sequence-based noise reduction software such as QS can significantly reduce peak sound pressure levels, without a loss of subjective image quality and increase SNR at constant lesion detection rates.
引用
收藏
页数:13
相关论文
共 33 条
[1]   Acoustic noise reduction in MRI using Silent Scan: an initial experience [J].
Alibek, Sedat ;
Vogel, Mika ;
Sun, Wei ;
Winkler, David ;
Baker, Christopher A. ;
Burke, Michael ;
Gloger, Hubertus .
DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2014, 20 (04) :360-363
[2]   Reduction of MRI acoustic noise achieved by manipulation of scan parameters e A study using veterinary MR sequences [J].
Baker, Martin A. .
RADIOGRAPHY, 2013, 19 (01) :11-16
[3]   POTENTIAL HEARING-LOSS RESULTING FROM MR IMAGING [J].
BRUMMETT, RE ;
TALBOT, JM ;
CHARUHAS, P .
RADIOLOGY, 1988, 169 (02) :539-540
[4]   A new silent magnetic resonance imaging using a rotating DC gradient [J].
Cho, ZH ;
Chung, ST ;
Chung, JY ;
Park, SH ;
Kim, JS ;
Moon, CH ;
Hong, IK .
MAGNETIC RESONANCE IN MEDICINE, 1998, 39 (02) :317-321
[5]   Quiet PROPELLER MRI Techniques Match the Quality of Conventional PROPELLER Brain Imaging Techniques [J].
Corcuera-Solano, I. ;
Doshi, A. ;
Pawha, P. S. ;
Gui, D. ;
Gaddipati, A. ;
Tanenbaum, L. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2015, 36 (06) :1124-1127
[6]   Reduction of gradient acoustic noise in MRI using SENSE-EPI [J].
de Zwart, JA ;
van Gelderen, P ;
Kellman, P ;
Duyn, JH .
NEUROIMAGE, 2002, 16 (04) :1151-1155
[7]   Active-passive gradient shielding for MRI acoustic noise reduction [J].
Edelstein, WA ;
Kidane, TK ;
Taracila, V ;
Baig, TN ;
Eagan, TP ;
Cheng, YCN ;
Brown, RW ;
Mallick, JA .
MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (05) :1013-1017
[8]   Making MRI Quieter [J].
Edelstein, WA ;
Hedeen, RA ;
Mallozzi, RP ;
El-Hamamsy, SA ;
Ackermann, RA ;
Havens, TJ .
MAGNETIC RESONANCE IMAGING, 2002, 20 (02) :155-163
[9]   Quiet MR sequences in clinical routine: initial experience in abdominal imaging [J].
Fischer, Sebastian ;
Grodzki, David M. ;
Domschke, Markus ;
Albrecht, Moritz ;
Bodelle, Boris ;
Eichler, Katrin ;
Hammerstingl, Renate ;
Vogl, Thomas J. ;
Zangos, Stephan .
RADIOLOGIA MEDICA, 2017, 122 (03) :194-203
[10]   Measuring signal-to-noise ratio in partially parallel imaging MRI [J].
Goerner, Frank L. ;
Clarke, Geoffrey D. .
MEDICAL PHYSICS, 2011, 38 (09) :5049-5057