A Sensitive capacitive immunosensor for direct detection of human heart fatty acid-binding protein (h-FABP)

被引:42
作者
Mihailescu, Carmen-Marinela [1 ]
Stan, Dana [2 ]
Iosub, Rodica [3 ]
Moldovan, Carmen [3 ]
Savin, Mihaela [4 ]
机构
[1] Univ Bucharest, Fac Chem, Dept Phys Chem, Bucharest, Romania
[2] DDS Diagnost, Bucharest 040312, Romania
[3] Natl Inst R&D Microtechnol IMT, Bucharest 077190, Romania
[4] Telemedica, Bucharest, Romania
关键词
Self-assembled monolayer; Cardiac marker; Acute myocardial infarction; Capacitive immunosensor; h-FABP; SELF-ASSEMBLED MONOLAYERS; IMPEDANCE SPECTROSCOPY; GOLD ELECTRODES; BIOSENSORS; ANTIBODY; CHAIN;
D O I
10.1016/j.talanta.2014.08.067
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The fabrication of a capacitive interdigitated immunosensor (CID) based on a mixed self-assembled monolayer (mSAM) film for the direct detection of heart fatty-add binding protein (h-FABP) without any labeling is described. The capacitance changes of mSAMs vs. homogenous ordered self-assembled monolayers (hSAMs) on gold work electrodes/covalently bonded antibodies/buffered medium are utilized for monitoring the specific antibody-antigen interaction. Capacitance measurements in the absence and presence of Faradaic currents were performed. The electrochemical properties of mixed monolayers were compared with those of a pure monolayer of 11-mercaptoundecanoic acid (MUA) self-assembled on gold surfaces. Taking into account the stability of the studied monolayers during the electrochemical experiments with the Faradaic process, the best SAM functionalization method was used for developing a sensitive capacitive immunosensor with a non-Faradaic process for direct immune detection of human h-FABP. Under the optimized conditions, the proposed mixed self-assembled monolayer (mSAM(1)) on gold electrode exhibited good insulating properties such as a capacitive behavior when detecting h-FABP from human serum in the range of 98 pg ml(-1)-100 ng ml(-1), with a detection limit of 0.836 ng ml(-1) comparative with a homogenous self-assembled monolayer (hSAM). (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:37 / 43
页数:7
相关论文
共 28 条
[1]   FORMATION OF MONOLAYERS BY THE COADSORPTION OF THIOLS ON GOLD - VARIATION IN THE LENGTH OF THE ALKYL CHAIN [J].
BAIN, CD ;
WHITESIDES, GM .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1989, 111 (18) :7164-7175
[2]  
Bard A.J., 1980, ELECTROCHEMICAL METH
[3]   An immunological interleukin-6 capacitive biosensor using perturbation with a potentiostatic step [J].
Berggren, C ;
Bjarnason, B ;
Johansson, G .
BIOSENSORS & BIOELECTRONICS, 1998, 13 (10) :1061-1068
[4]  
Berggren C, 2001, ELECTROANAL, V13, P173, DOI 10.1002/1521-4109(200103)13:3<173::AID-ELAN173>3.0.CO
[5]  
2-B
[6]   Capacitance measurements of antibody-antigen interactions in a flow system [J].
Berggren, C ;
Johansson, G .
ANALYTICAL CHEMISTRY, 1997, 69 (18) :3651-3657
[7]   Development of a myoglobin impedimetric immunosensor based on mixed self-assembled monolayer onto gold [J].
Billah, Morsaline ;
Hays, Henry C. W. ;
Millner, Paul A. .
MICROCHIMICA ACTA, 2008, 160 (04) :447-454
[8]   Characterization of alkanethiol-self-assembled monolayers-modified gold electrodes by electrochemical impedance spectroscopy [J].
Campuzano, S ;
Pedrero, M ;
Montemayor, C ;
Fatás, E ;
Pingarrón, JM .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2006, 586 (01) :112-121
[9]  
Chain F., 2000, CLIN CHEM, V46, P718
[10]  
Christenson RH, 1998, CLIN CHEM, V44, P1855