Uncertainty of relative sensitivity factors in glow discharge mass spectrometry

被引:11
作者
Meija, Juris [1 ]
Methven, Brad [1 ]
Sturgeon, Ralph E. [1 ]
机构
[1] Natl Res Council Canada, Measurement Sci & Stand, 1200 Montreal Rd, Ottawa, ON K1A 0R6, Canada
关键词
glow discharge mass spectrometry; calibration model; relative sensitivity factors; traceability; uncertainty evaluation; least squares adjustment; PHYSICAL CONSTANTS; CALIBRATION; PURITY; VALUES;
D O I
10.1088/1681-7575/aa83d6
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The concept of the relative sensitivity factors required for the correction of the measured ion beam ratios in pin-cell glow discharge mass spectrometry is examined in detail. We propose a data-driven model for predicting the relative response factors, which relies on a non-linear least squares adjustment and analyte/matrix interchangeability phenomena. The model provides a self-consistent set of response factors for any analyte/matrix combination of any element that appears as either an analyte or matrix in at least one known response factor.
引用
收藏
页码:796 / 804
页数:9
相关论文
共 50 条
  • [41] Estimation of Uncertainty Budgets in Analysis of Nickel by ICP-MS and High-Resolution Mass Spectrometry with a Glow Discharge
    A. V. Alekseev
    P. V. Yakimovich
    Moscow University Chemistry Bulletin, 2022, 77 : 206 - 213
  • [42] Characterization of an inductively amplified magnetron plasma by glow discharge mass spectrometry
    Molle, C
    Beauvois, A
    Wautelet, M
    Dauchot, JP
    Hecq, M
    VACUUM, 1999, 53 (1-2) : 17 - 20
  • [43] Determination of Metallic Impurities in Carbon Nanotubes by Glow Discharge Mass Spectrometry
    Grinberg, Patricia
    Methven, Bradley A. J.
    Swider, Katarzyna
    Mester, Zoltan
    ACS OMEGA, 2021, 6 (35): : 22717 - 22725
  • [44] Studies on the ion formation in glow discharge mass spectrometry ion sources
    Itoh, S
    Hasegawa, R
    JOURNAL OF THE JAPAN INSTITUTE OF METALS, 1999, 63 (06) : 783 - 789
  • [45] Comparative study on the analysis of cast iron using glow discharge mass spectrometry and spark discharge optical emission spectrometry
    Itoh, S
    Kimura, T
    Yamaguchi, H
    Kobayashi, T
    TETSU TO HAGANE-JOURNAL OF THE IRON AND STEEL INSTITUTE OF JAPAN, 2004, 90 (10): : 846 - 851
  • [46] Analyses of silver and palladium brazing filler metals by glow discharge mass spectrometry
    Itoh, S
    Yamaguchi, H
    Hobo, T
    Kobayashi, T
    JOURNAL OF THE JAPAN INSTITUTE OF METALS, 2000, 64 (09) : 776 - 780
  • [47] ISOTOPE RATIO MEASUREMENTS USING GLOW-DISCHARGE MASS-SPECTROMETRY
    RICIPUTI, LR
    DUCKWORTH, DC
    BARSHICK, CM
    SMITH, DH
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1995, 146 : 55 - 64
  • [48] ANALYSIS OF TITANIUM-ALLOYS BY GLOW-DISCHARGE MASS-SPECTROMETRY
    ITOH, S
    HIROSE, F
    HASEGAWA, S
    HASEGAWA, R
    JOURNAL OF THE JAPAN INSTITUTE OF METALS, 1993, 57 (10) : 1186 - 1191
  • [49] Analysis of trace impurities in palladium metal powders by glow discharge mass spectrometry
    Wayne, DM
    Yoshida, TM
    Vance, DE
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (09) : 861 - 869
  • [50] A novel isotope analysis of oxygen in uranium oxides: comparison of secondary ion mass spectrometry, glow discharge mass spectrometry and thermal ionization mass spectrometry
    Pajo, L
    Tamborini, G
    Rasmussen, G
    Mayer, K
    Koch, L
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2001, 56 (05) : 541 - 549