Uncertainty of relative sensitivity factors in glow discharge mass spectrometry

被引:11
作者
Meija, Juris [1 ]
Methven, Brad [1 ]
Sturgeon, Ralph E. [1 ]
机构
[1] Natl Res Council Canada, Measurement Sci & Stand, 1200 Montreal Rd, Ottawa, ON K1A 0R6, Canada
关键词
glow discharge mass spectrometry; calibration model; relative sensitivity factors; traceability; uncertainty evaluation; least squares adjustment; PHYSICAL CONSTANTS; CALIBRATION; PURITY; VALUES;
D O I
10.1088/1681-7575/aa83d6
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The concept of the relative sensitivity factors required for the correction of the measured ion beam ratios in pin-cell glow discharge mass spectrometry is examined in detail. We propose a data-driven model for predicting the relative response factors, which relies on a non-linear least squares adjustment and analyte/matrix interchangeability phenomena. The model provides a self-consistent set of response factors for any analyte/matrix combination of any element that appears as either an analyte or matrix in at least one known response factor.
引用
收藏
页码:796 / 804
页数:9
相关论文
共 50 条
  • [31] FACTORS INFLUENCING THE QUANTITATIVE-DETERMINATION OF TRACE-ELEMENTS IN SOILS BY GLOW-DISCHARGE MASS-SPECTROMETRY
    TENG, JW
    BARSHICK, CM
    DUCKWORTH, DC
    MORTON, SJ
    SMITH, DH
    KING, FL
    APPLIED SPECTROSCOPY, 1995, 49 (09) : 1361 - 1366
  • [32] Analysis of magnesium alloys by glow-discharge mass spectrometry
    Itoh, S
    Yamaguchi, H
    Hobo, T
    Kobayashi, T
    BUNSEKI KAGAKU, 2004, 53 (06) : 569 - 574
  • [33] Investigations of matrix independent calibration approaches in fast flow glow discharge mass spectrometry
    Gonzalez-Gago, Cristina
    Smid, Petr
    Hofmann, Thomas
    Venzago, Cornel
    Hoffmann, Volker
    Gruner, Wolfgang
    Pfeifer, Jens
    Richter, Silke
    Kipphardt, Heinrich
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2019, 34 (06) : 1109 - 1125
  • [34] Isotope ratio measurements by secondary ion mass spectrometry (SIMS) and glow discharge mass spectrometry (GDMS)
    Betti, M
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2005, 242 (2-3) : 169 - 182
  • [35] Estimation of Uncertainty Budgets in Analysis of Nickel by ICP-MS and High-Resolution Mass Spectrometry with a Glow Discharge
    Alekseev, A. V.
    Yakimovich, P. V.
    MOSCOW UNIVERSITY CHEMISTRY BULLETIN, 2022, 77 (04) : 206 - 213
  • [36] MATRIX EFFECTS ON THE RELATIVE SENSITIVITY FACTORS MEASURED BY LASER MICROPROBE MASS-SPECTROMETRY
    BATE, DJ
    LEAKE, JA
    MATTHEWS, LJ
    WALLACH, ER
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1993, 127 : 85 - 93
  • [37] ANALYSIS OF PLATINUM POWDER BY GLOW-DISCHARGE MASS-SPECTROMETRY
    VANSTRAATEN, M
    SWENTERS, K
    GIJBELS, R
    VERLINDEN, J
    ADRIAENSSENS, E
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (12) : 1389 - 1397
  • [38] Analysis of zirconium-based alloys by glow discharge mass spectrometry
    Itoh, S
    Yamaguchi, H
    Hobo, T
    Kobayashi, T
    JOURNAL OF THE JAPAN INSTITUTE OF METALS, 2001, 65 (01) : 53 - 59
  • [39] Application of argon as the internal standard method in glow discharge mass spectrometry
    Xing, Yu
    Xiaojia, Li
    Haizhou, Wang
    EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2007, 13 (02) : 125 - 133
  • [40] Round robin analysis of aluminium using glow discharge mass spectrometry
    Venzago, C
    Ohanessian-Pierrard, L
    Kasik, M
    Collisi, U
    Baude, S
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (03) : 189 - 193