Early Detection of Pump Thrombosis in Patients With Left Ventricular Assist Device

被引:18
|
作者
Grabska, Joanna [1 ]
Schloeglhofer, Thomas [1 ,2 ,3 ]
Gross, Christoph [1 ,3 ]
Maw, Martin [1 ,3 ]
Dimitrov, Kamen [2 ]
Wiedemann, Dominik [2 ,3 ]
Zimpfer, Daniel [2 ,3 ]
Schima, Heinrich [1 ,2 ,3 ]
Moscato, Francesco [1 ,3 ]
机构
[1] Med Univ Vienna, Ctr Med Phys & Biomed Engn, Vienna, Austria
[2] Med Univ Vienna, Dept Cardiac Surg, Wahringer Gurtel 18, A-1090 Vienna, Austria
[3] Ludwig Boltzmann Cluster Cardiovasc Res, Vienna, Austria
关键词
ventricular assist device; pump thrombus; early detection; log files; anticoagulation quality; INTERMACS ANNUAL-REPORT; DIAGNOSIS; EVENTS;
D O I
10.1097/MAT.0000000000001015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Pump thrombosis (PT) is a serious adverse event in patients receiving left ventricular assist devices (LVAD). The study aims to determine whether pump parameters and clinical data may enable early detection of PT. This retrospective study included 88 patients who received an LVAD between 2012 and 2015 among which those with intra-PT were identified. In a propensity score-matched control group observation, time periods were matched with time before thrombosis. International normalized ratio (INR) time in therapeutic range (TTR) and lactate dehydrogenase (LDH) were analyzed for 60 days preceding PT. Furthermore, pump data (power, flow, and speed) in HeartWare ventricular assist devices (HVAD) patients were analyzed 7 days before PT using a mixed-design analysis of variance to investigate temporal changes in pump data. Pump thrombosis occurred in 15 patients (13 males, age 58 +/- 10 years, 7 HeartMate II and 8 HVAD). International normalized ratio therapeutic range (2.0-3.0) and acetylsalicylic acid daily doses (100-200 mg) were similar for both groups, but patients with PT had lower TTR (36% vs. 65%; p = 0.025). No significant difference in LVAD power between groups was seen at baseline (p = 0.31), and power did not change in the control group over time (p > 0.99). Lactate dehydrogenase increased already 1 week prior PT and power from 4.4 +/- 0.8 W at baseline to 4.9 +/- 0.8 W (p = 0.007) 2 days before readmission and to 6.5 +/- 1.8 W (p = 0.015) at readmission. Pump thrombosis is associated with a lower percentage of INR TTR and elevated LDH before the event. A better monitoring of pump parameters would enable PT detection already up to 2 days in advance.
引用
收藏
页码:348 / 354
页数:7
相关论文
共 50 条
  • [1] Early detection of pump thrombosis in patients with left ventricular assist device
    Grabska, J.
    Schloeglhofer, T.
    Gross, C.
    Dimitrov, K.
    Wiedemann, D.
    Zimpfer, D.
    Schima, H.
    Moscato, F.
    WIENER KLINISCHE WOCHENSCHRIFT, 2017, 129 : S18 - S18
  • [2] Early Detection of Pump Thrombosis in Patients with Left Ventricular Assist Device
    Grabska, J.
    Schloeglhofer, T.
    Gross, C.
    Dimitrov, K.
    Wiedemann, D.
    Zimpfer, D.
    Schima, H.
    Moscato, F.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2017, 36 (04): : S256 - S256
  • [3] Cavitation in left ventricular assist device patients: a potential early sign of pump thrombosis
    Zuin, Marco
    Rigatelli, Gianluca
    Braggion, Gabriele
    Bacich, Daniela
    Thach Nguyen
    HEART FAILURE REVIEWS, 2020, 25 (06) : 965 - 972
  • [4] Cavitation in left ventricular assist device patients: a potential early sign of pump thrombosis
    Marco Zuin
    Gianluca Rigatelli
    Gabriele Braggion
    Daniela Bacich
    Thach Nguyen
    Heart Failure Reviews, 2020, 25 : 965 - 972
  • [5] Extending the perspective on left ventricular assist device pump thrombosis to left ventricular assist device system thrombosis
    Kormos, Robert L.
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2018, 155 (01): : E37 - E38
  • [6] Advanced telemonitoring of Left Ventricular Assist Device patients for the early detection of thrombosis
    Glitza, Jenny I.
    Mueller-von Aschwege, Frerk
    Eichelberg, Marco
    Reiss, Nils
    Schmidt, Thomas
    Feldmann, Christina
    Wendl, Regina
    Schmitto, Jan D.
    Hein, Andreas
    JOURNAL OF NETWORK AND COMPUTER APPLICATIONS, 2018, 118 : 74 - 82
  • [7] TIROFIBAN FOR SUSPECTED PUMP THROMBOSIS IN PATIENTS WITH A LEFT VENTRICULAR ASSIST DEVICE
    Nei, Scott
    Herman, Libby
    Ritchie, Brianne
    Stulak, John
    CRITICAL CARE MEDICINE, 2018, 46 (01) : 57 - 57
  • [8] NATIONAL TRENDS AND OUTCOMES OF PUMP THROMBOSIS IN LEFT VENTRICULAR ASSIST DEVICE PATIENTS
    Ganta, Nitheesha
    Ahmad, Basharat
    Ngwa, Julius
    Opoku-Asare, Isaac
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2019, 73 (09) : 946 - 946
  • [9] A Stepwise Approach to Left Ventricular Assist Device Pump Thrombosis
    Kocabeyoglu, Sinan Sabit
    Kervan, Umit
    Sert, Dogan Emre
    Karahan, Mehmet
    Unal, Ertekin Utku
    Catav, Zeki
    Pac, Mustafa
    HEART LUNG AND CIRCULATION, 2021, 30 (04): : 567 - 576
  • [10] Tirofiban in Suspected Left Ventricular Assist Device Pump Thrombosis
    Nei, Scott D.
    Wieruszewski, Patrick M.
    Orzel, Libby A.
    Ritchie, Brianne M.
    Stulak, John M.
    ASAIO JOURNAL, 2020, 66 (08) : 886 - 889