Automating the Timed Up and Go Test Using a Depth Camera

被引:18
作者
Dubois, Amandine [1 ]
Bihl, Titus [2 ]
Bresciani, Jean-Pierre [1 ]
机构
[1] Univ Fribourg, Dept Med, CH-1700 Fribourg, Switzerland
[2] Cantonal Hosp, CH-1700 Fribourg, Switzerland
关键词
timed up and go; automated clinical test; objective assessment; elderly people; depth camera; fall prevention; FALLS; RISK;
D O I
10.3390/s18010014
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Fall prevention is a human, economic and social issue. The Timed Up and Go (TUG) test is widely used to identify individuals with a high fall risk. However, this test has been criticized because its "diagnostic" is too dependent on the conditions in which it is performed and on the healthcare professionals running it. We used the Microsoft Kinect ambient sensor to automate this test in order to reduce the subjectivity of outcome measures and to provide additional information about patient performance. Each phase of the TUG test was automatically identified from the depth images of the Kinect. Our algorithms accurately measured and assessed the elements usually measured by healthcare professionals. Specifically, average TUG test durations provided by our system differed by only 0.001 s from those measured by clinicians. In addition, our system automatically extracted several additional parameters that allowed us to accurately discriminate low and high fall risk individuals. These additional parameters notably related to the gait and turn pattern, the sitting position and the duration of each phase. Coupling our algorithms to the Kinect ambient sensor can therefore reliably be used to automate the TUG test and perform a more objective, robust and detailed assessment of fall risk.
引用
收藏
页数:13
相关论文
共 32 条
[1]   Is the Timed Up and Go test a useful predictor of risk of falls in community dwelling older adults: a systematic review and meta- analysis [J].
Barry, Emma ;
Galvin, Rose ;
Keogh, Claire ;
Horgan, Frances ;
Fahey, Tom .
BMC GERIATRICS, 2014, 14
[2]   Convergent Validity of a Wearable Sensor System for Measuring Sub-Task Performance during the Timed Up-and-Go Test [J].
Beyea, James ;
McGibbon, Chris A. ;
Sexton, Andrew ;
Noble, Jeremy ;
O'Connell, Colleen .
SENSORS, 2017, 17 (04)
[3]   Estimation of the concordance correlation coefficient for repeated measures using SAS and R [J].
Carrasco, Josep L. ;
Phillips, Brenda R. ;
Puig-Martinez, Josep ;
King, Tonya S. ;
Chinchilli, Vernon M. .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2013, 109 (03) :293-304
[4]  
Carstensen B., Int J Biostat, DOI DOI 10.2202/1557-4679.1107
[5]   A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques [J].
Critchley, LAH ;
Critchley, JAJH .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) :85-91
[6]  
Demiris George, 2009, Technol Health Care, V17, P41, DOI 10.3233/THC-2009-0530
[7]   A clinical test of stepping and change of direction to identify multiple falling older adults [J].
Dite, W ;
Temple, VA .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2002, 83 (11) :1566-1571
[8]  
Dubois A, 2014, P 36 ANN INT C IEEE
[9]  
Dubois A, 2013, IEEE ENG MED BIO, P4666, DOI 10.1109/EMBC.2013.6610588
[10]   Performing gait analysis within the timed up & go assessment test: comparison of aTUG to a marker-based tracking system [J].
Frenken, Thomas ;
Lohmann, Okko ;
Frenken, Melina ;
Steen, Enno-Edzard ;
Hein, Andreas .