Current Clinical Practice in Patients With Cardiac Implantable Electronic Devices (CIED) Undergoing Radiotherapy (RT)

被引:3
作者
Azraai, Meor [1 ]
Souza, Daniel [1 ]
Nadurata, Voltaire [1 ]
机构
[1] Bendigo Hlth, Dept Cardiol, 100 Barnard St, Bendigo, Vic, Australia
关键词
Cardiac implantable electronic devices; Pacemaker; Implantable cardioverter defibrillator; Cardiac resynchronisation therapy; Arrhythmia; Radiotherapy; Cancer; PHOTON-BEAM IRRADIATION; RADIATION-THERAPY; CARDIOVERTER-DEFIBRILLATORS; LATEST GENERATION; PACEMAKER; MANAGEMENT; ENERGY; CANCER; MALFUNCTIONS;
D O I
10.1016/j.hlc.2021.10.020
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Patients with cardiac implantable electronic devices (CIED) undergoing radiotherapy (RT) becoming more common due to ageing of the population. With newer CIEDs implementing the complementary metal -oxide semiconductor (CMOS) technology which allows the miniaturisation of CIED, it is also more sus-ceptible to RT. Effects of RT on CIED ranges from device interference, device operational/memory errors to permanent damage. These malfunctions can cause life threatening clinical effects. Cumulative dose is not the only component of RT that causes CIED malfunction, as neutron use and dose rate effect also affects CIEDs. The management of this patient cohort in clinical practice is inconsistent due to lack of a consistent guideline from manufacturers and physician specialty societies. Our review will focus on the current clinical practice and the recent updated guidelines of managing patients with CIED undergoing RT. We aim to simplify the evidence and provide a simple and easy to use guide based on the recent guidelines.
引用
收藏
页码:327 / 340
页数:14
相关论文
共 59 条
[1]  
[Anonymous], 2011, RAD THER BIOTRONIK C
[2]  
[Anonymous], 2012, WALTER MILLERS TXB R
[3]   Cardiac resynchronization therapy: implant rates, temporal trends and relationships with heart failure epidemiology [J].
Boriani, Giuseppe ;
Berti, Elena ;
Belotti, Laura Maria Beatrice ;
Biffi, Mauro ;
Carboni, Angelo ;
Bandini, Alberto ;
Casali, Edoardo ;
Tomasi, Corrado ;
Toselli, Tiziano ;
Baraldi, Paolo ;
Bottoni, Nicola ;
Barbato, Gaetano ;
Sassone, Biagio .
JOURNAL OF CARDIOVASCULAR MEDICINE, 2014, 15 (02) :147-154
[4]  
Boston Scientific,, 2013, 0021675 BOST SCIENT
[5]   Single event upsets in implantable cardioverter defibrillators [J].
Bradley, PD ;
Normand, E .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (06) :2929-2940
[6]   Trends in the incidence and prevalence of cardiac pacemaker insertions in an ageing population [J].
Bradshaw, Pamela J. ;
Stobie, Paul ;
Knuiman, Matthew W. ;
Briffa, Thomas G. ;
Hobbs, Michael S. T. .
OPEN HEART, 2014, 1 (01)
[7]   PACEMAKER FAILURE ASSOCIATED WITH THERAPEUTIC RADIATION [J].
BROOKS, C ;
MUTTER, M .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1988, 6 (06) :591-593
[8]   Out-of-field doses and neutron dose equivalents for electron beams from modern Varian and Elekta linear accelerators [J].
Cardenas, Carlos E. ;
Nitsch, Paige L. ;
Kudchadker, Rajat J. ;
Howell, Rebecca M. ;
Kry, Stephen F. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (04) :442-455
[9]   Monitoring Patients With Implanted Cardiac Rhythm Devices Receiving Radiation Therapy [J].
Chavez, Marilyn I. .
ONCOLOGY NURSING FORUM, 2009, 36 (06) :629-632
[10]   High incidence of implantable cardioverter defibrillator malfunctions during radiation therapy: neutrons as a probable cause of soft errors [J].
Elders, Jan ;
Kunze-Busch, Martina ;
Jan Smeenk, Robert ;
Smeets, Joep L. R. M. .
EUROPACE, 2013, 15 (01) :60-65