Phase-encode reordering to minimize errors caused by motion

被引:19
作者
Macgowan, CK
Wood, ML
机构
[1] UNIV TORONTO,DEPT MED BIOPHYS,TORONTO,ON,CANADA
[2] UNIV TORONTO,DEPT MED IMAGING,TORONTO,ON,CANADA
[3] ST MICHAELS HOSP,TORONTO,ON M5B 1W8,CANADA
关键词
motion artifacts; phase-encode reordering;
D O I
10.1002/mrm.1910350318
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new method for suppressing the effects of motion in MR images by reordering the acquisition of k space has been developed. Existing reordering methods suffer from image blurring, The method presented here applies specifically to translation along the phase-encoding direction, in which case it reduces both ghosting and blurring. The method is conceptually similar to a linear phase shift of k space, except that it has the advantage of not corrupting stationary structures inadvertently. The method is intended for anatomic sites in which substantial translational motion occurs along the phase-encoding direction, such as the cranial-caudal motion of the liver and kidneys, The reordering method is motivated from an analysis of factors affecting the severity of motion artifacts. The theory behind the reordering method is described and validated experimentally by imaging a moving phantom.
引用
收藏
页码:391 / 398
页数:8
相关论文
共 13 条
  • [1] RESPIRATORY EFFECTS IN TWO-DIMENSIONAL FOURIER-TRANSFORM MR IMAGING
    AXEL, L
    SUMMERS, RM
    KRESSEL, HY
    CHARLES, C
    [J]. RADIOLOGY, 1986, 160 (03) : 795 - 801
  • [2] RESPIRATORY ORDERED PHASE ENCODING (ROPE) - A METHOD FOR REDUCING RESPIRATORY MOTION ARTIFACTS IN MR IMAGING
    BAILES, DR
    GILDERDALE, DJ
    BYDDER, GM
    COLLINS, AG
    FIRMIN, DN
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1985, 9 (04) : 835 - 838
  • [3] Bracewell R., 1984, FOURIER TRANSFORM IT
  • [4] ADAPTIVE TECHNIQUE FOR HIGH-DEFINITION MR IMAGING OF MOVING STRUCTURES
    EHMAN, RL
    FELMLEE, JP
    [J]. RADIOLOGY, 1989, 173 (01) : 255 - 263
  • [5] HAACKE E M, 1986, Magnetic Resonance Imaging, V4, P359, DOI 10.1016/0730-725X(86)91046-5
  • [6] SELECTION OF A CONVOLUTION FUNCTION FOR FOURIER INVERSION USING GRIDDING
    JACKSON, JI
    MEYER, CH
    NISHIMURA, DG
    MACOVSKI, A
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (03) : 473 - 478
  • [7] COMPENSATION FOR EFFECTS OF LINEAR MOTION IN MR IMAGING
    KORIN, HW
    FARZANEH, F
    WRIGHT, RC
    RIEDERER, SJ
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1989, 12 (01) : 99 - 113
  • [8] RESPIRATORY KINEMATICS OF THE UPPER ABDOMINAL ORGANS - A QUANTITATIVE STUDY
    KORIN, HW
    EHMAN, RL
    RIEDERER, SJ
    FELMLEE, JP
    GRIMM, RC
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (01) : 172 - 178
  • [9] ALTERED PHASE-ENCODING ORDER FOR REDUCED SENSITIVITY TO MOTION IN 3-DIMENSIONAL MR IMAGING
    KORIN, HW
    RIEDERER, SJ
    BAMPTON, AEH
    EHMAN, RL
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (06): : 687 - 693
  • [10] NMR FOURIER ZEUGMATOGRAPHY
    KUMAR, A
    WELTI, D
    ERNST, RR
    [J]. JOURNAL OF MAGNETIC RESONANCE, 1975, 18 (01) : 69 - 83