Percutaneous Antegrade Ureteral Stent Placement During Pediatric Robot-Assisted Laparoscopic Pyeloplasty

被引:19
作者
Noh, Paul H. [1 ]
DeFoor, W. Robert [1 ]
Reddy, Pramod P. [1 ]
机构
[1] Cincinnati Childrens Hosp, Med Ctr, Div Pediat Urol, Cincinnati, OH 45229 USA
关键词
DISMEMBERED PYELOPLASTY; INSERTION;
D O I
10.1089/end.2011.0168
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background and Purpose: Robot-assisted laparoscopic pyeloplasty has become more widely used. Intraoperative placement and confirmation of ureteral stent position can be cumbersome with the robotic arms in place. We present a technique of percutaneous antegrade stent placement that is reliable with minimal morbidity. Patients and Methods: A retrospective cohort study was performed. Patient demographics, radiographic imaging, intraoperative details, and surgical outcomes were abstracted from the medical record. A 14-gauge angiocatheter was placed through the abdominal wall. A ureteral stent was guided over a wire down the dismembered ureter. Stent position was confirmed by retrograde reflux of methylene blue. A urethral catheter was left in place for 12 to 36 hours. Results: Twenty-nine patients (15 male, 14 female) were identified. Average age was 10 years. Average follow-up was 14 months. Fifteen left-and 14 right-sided procedures were performed. Two patients needed retrograde stent placement. Mean time to correctly position the stent was less than 5 minutes. Postoperatively, one patient had a urine leak managed by an indwelling urethral catheter and did not need percutaneous drainage. All stents were removed approximately 4 to 6 weeks postoperatively. One patient had retrograde migration of the stent managed by ureteroscopy at the time of stent retrieval. Conclusions: Antegrade ureteral stent placement through a percutaneous angiocatheter, during robot-assisted laparoscopic pyeloplasty, is a rapid and effective technique. Intraoperative confirmation of stent position can be obtained, using methylene blue bladder distention, without repositioning the patient or undocking the surgical robot.
引用
收藏
页码:1847 / 1851
页数:5
相关论文
共 21 条
[1]   Laparoscopic cholangiogram-guide device applied for intracorporeal antegrade ureteric stenting during laparoscopic pyeloplasty [J].
Andreoni, C ;
Paiva, C ;
Sabino, L ;
Gattás, N ;
Ortiz, V ;
Srougi, M .
BJU INTERNATIONAL, 2005, 96 (07) :1139-1141
[2]   Antegrade versus retrograde stenting in laparoscopic pyeloplasty [J].
Arumainayagam, Nimalan ;
Minervini, Andrea ;
Davenport, Kim ;
Kumar, Vivekanandan ;
Masieri, Lorenzo ;
Serni, Sergio ;
Carini, Marco ;
Timoney, Anthony G. ;
Keeley, Francis X., Jr. .
JOURNAL OF ENDOUROLOGY, 2008, 22 (04) :671-674
[3]   Is retrograde stenting more reliable than antegrade stenting for pyeloplasty in infants and children? [J].
Chandrasekharam, VVSS .
UROLOGY, 2005, 66 (06) :1301-1304
[4]   Endoscopic Bridge Operating-Guide Device Applied for Intracorporeal Antegrade Ureteric Stenting During Laparoscopic Pyeloplasty [J].
Derouiche, Amine ;
El Atat, Rabii ;
Ben Slama, Mohamed Riadh ;
Chebil, Mohamed .
JOURNAL OF ENDOUROLOGY, 2009, 23 (11) :1871-1874
[5]  
Eassa W, 2010, J PEDIAT UROL
[6]   Antegrade Ureteral Stenting During Laparoscopic Dismembered Pyeloplasty: Intraoperative Findings and Long-Term Outcome [J].
El-Feel, Ahmed Sobhy ;
Abdel-Hakim, Mahmoud Amr ;
Abouel-Fettouh, Hazem Ibrahim ;
Abdel-Hakim, Amr Mahmoud .
JOURNAL OF ENDOUROLOGY, 2010, 24 (04) :551-555
[7]  
Elmalik K, 2008, J Pediatr Urol, V4, P275, DOI 10.1016/j.jpurol.2008.01.205
[8]   Novel technique of retrograde ureteral stenting during laparoscopic pyeloplasty [J].
Gaitonde, Krishnanath ;
Roesel, Gwen ;
Donovan, James .
JOURNAL OF ENDOUROLOGY, 2008, 22 (06) :1199-1202
[9]  
Ginger VAT, 2009, UROLOGY, V73, P377, DOI 10.1016/j.urology.2008.08.507
[10]   Retrograde Percutaneous Access for Kidney Internal Splint Stent Catheter Placement in Pediatric Laparoscopic Pyeloplasty: Avoiding Stent Removal in the Operating Room [J].
Hadley, David A. ;
Wicher, Christopher ;
Wallis, M. Chad .
JOURNAL OF ENDOUROLOGY, 2009, 23 (12) :1991-1994