Stripping voltammetry of mercury(II) with a chemically modified carbon paste electrode containing silica gel functionalized with 2,5-dimercapto-1,3,4-thiadiazole

被引:15
作者
Dias, NL [1 ]
do Carmo, DR [1 ]
机构
[1] Univ Estadual Paulista, UNESP, Dept Quim & Fis, BR-15385000 Ilha Solteira, SP, Brazil
关键词
silica; carbon paste electrodes; voltammetry; mercury preconcentration;
D O I
10.1002/elan.200403252
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The accumulation voltammetry of mercury(II) was investigated at a carbon paste electrode chemically modified with silica gel functionalized with 2,5-dimercapto-1,3,4-thiadiazole (DTTPSG-CPE). The repetitive cyclic voltammogram of mercury(II) solution in the potential range -0.2 to +0.8 V (vs. Ag/AgCl), (0.02 mol L-1 KNO3; nu=20 mV s(-1)) show two peaks one at about 0.0 V and other at 0.31 V. However, the cathodic wave peak, around 0.0 V, is irregular and changes its form in each cycle. This peak at about 0.0 V is the reduction current for mercury(II) accumulated in the DTTPSG-CPE. The anodic wave peak at 0.31 V is well-defined and does not change during the cycles. The resultant material was characterized by cyclic and differential pulse anodic stripping voltammetry performed with the electrode in differents supporting electrolytes. The mercury response was evaluated with respect to pH, electrode composition, preconcentration time, mercury concentration, "cleaning" solution, possible interferences and other variables. The precision for six determinations (n=6) of 0.05 and 0.20 mg (L)-(1) Hg(II) was 2.8 and 2.2% (relative standard deviation), respectively. The method was satisfactory and used to determine the concentration of mercury(II) in natural waters contaminated by this metal.
引用
收藏
页码:1540 / 1546
页数:7
相关论文
共 42 条
[1]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF PENICILLAMINE IN WHOLE-BLOOD, PLASMA, AND URINE [J].
BERGSTROM, RF ;
KAY, DR ;
WAGNER, JG .
JOURNAL OF CHROMATOGRAPHY, 1981, 222 (03) :445-452
[2]   THE PRICE OF GOLD - MERCURY EXPOSURE IN THE AMAZONIAN RAIN-FOREST [J].
BRANCHES, FJP ;
ERICKSON, TB ;
AKS, SE ;
HRYHORCZUK, DO .
JOURNAL OF TOXICOLOGY-CLINICAL TOXICOLOGY, 1993, 31 (02) :295-306
[3]   CAPILLARY COLUMN GAS-CHROMATOGRAPHY FOR MERCURY SPECIATION [J].
BULSKA, E ;
BAXTER, DC ;
FRECH, W .
ANALYTICA CHIMICA ACTA, 1991, 249 (02) :545-554
[4]  
Carmo Devaney R. do, 2004, Port. Electrochim. Acta, V22, P71
[5]   Mediated biosensors [J].
Chaubey, A ;
Malhotra, BD .
BIOSENSORS & BIOELECTRONICS, 2002, 17 (6-7) :441-456
[6]  
Dias Filho N. L., 2004, ENCY SURFACE COLLOID, P1
[7]   Adsorption and structure of copper(II) complexes on a silica gel surface chemically modified with 2-aminothiazole [J].
Dias, NL .
POLYHEDRON, 1999, 18 (17) :2241-2247
[8]   Adsorption of copper(II) and cobalt(II) complexes on a silica gel surface chemically modified with 3-amino-1,2,4-triazole [J].
Dias, NL .
COLLOIDS AND SURFACES A-PHYSICOCHEMICAL AND ENGINEERING ASPECTS, 1998, 144 (1-3) :219-227
[9]   Adsorption of Cu(II) and Co(II) complexes on a silica gel surface chemically modified with 2-mercaptoimidazole [J].
Dias, NL .
MIKROCHIMICA ACTA, 1999, 130 (04) :233-240
[10]  
Dias NL, 1997, SEPAR SCI TECHNOL, V32, P2535