Accuracy testing of a new optical device for noninvasive estimation of systolic and diastolic blood pressure compared to intra-arterial measurements

被引:25
作者
Pellaton, Cyril [1 ]
Vybornova, Anna [2 ]
Fallet, Sibylle [2 ]
Marques, Luisa [1 ]
Grossenbacher, Olivier [2 ]
De Marco, Bastien [2 ]
Chapuis, Valentin [2 ]
Bertschi, Mattia [2 ]
Alpert, Bruce S. [3 ]
Sola, Josep [2 ]
机构
[1] Hop Neuchatelois Pourtales, Neuchatel, Switzerland
[2] Aktiia SA, Neuchatel, Switzerland
[3] Univ Tennessee, Memphis, TN USA
关键词
blood pressure; cuffless blood pressure; optical blood pressure monitoring algorithms; optical technology;
D O I
10.1097/MBP.0000000000000421
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective The objective of this study was to compare the systolic (S) and diastolic (D) blood pressure (BP) estimations from a new optical device at the wrist with invasive measurements performed on patients scheduled for radial arterial catheterization in the ICU. Optical signals were automatically processed by a library of algorithms from Aktiia SA (OBPM - optical blood pressure monitoring algorithms). Methods A total of 31 participants from both sexes, aged 32-87 years, were enrolled in the study (NCT03837769). The measurement protocol consisted of the simultaneous recording of reflective photoplethysmographic signals (PPG) from the cuffless optical device and the reference BP values recorded by a contralateral radial arterial catheter. From the 31 participants, 23 subjects whose reference data quality requirements were adequate were retained for further analysis. The PPG signals from these patients were then automatically processed by the Aktiia OBPM library of algorithms, which generated uncalibrated estimates of SBP and DBP. After the automatic assessment of optical signal quality, 326 pairs of uncalibrated SBP and DBP determinations from 16 patients were available for analysis. These values were finally transformed into calibrated estimations (in mmHg) using arterial catheter SBP and DBP values, respectively. Results For SBP, a mean difference (+/- SD) of 0.0 +/- 7.1 mmHg between the arterial catheter and the optical device values was found, with 95% limits of agreement in the Bland-Altman method of -11.9 to + 12.2 mmHg (correlation of r = 0.87, P < 0.001). For DBP, a mean difference (+/- SD) of 0.0 +/- 2.9 mmHg between arterial catheter and the optical device values was found, with 95% limits of agreement in the Bland-Altman method of -4.8 to + 5.5 mmHg (correlation of r = 0.98, P < 0.001). Conclusion SBP and DBP values obtained by radial artery catheterization and those obtained from optical measurements at the wrist were compared. The new optical technique appears to be capable of replacing more traditional methods of BP estimation.
引用
收藏
页码:105 / 109
页数:5
相关论文
共 7 条
[1]  
Ghamri YA, 2018, P EUR
[2]  
Lima AP, 2002, CRIT CARE MED, V30, P1210
[3]   Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and Practice [J].
Mukkamala, Ramakrishna ;
Hahn, Jin-Oh ;
Inan, Omer T. ;
Mestha, Lalit K. ;
Kim, Chang-Sei ;
Toreyin, Hakan ;
Kyal, Survi .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2015, 62 (08) :1879-1901
[4]   Validation of an integrated pedal desk and electronic behavior tracking platform [J].
Schuna J.M., Jr. ;
Tudor-Locke C. ;
Proença M. ;
Barreira T.V. ;
Hsia D.S. ;
Pitta F. ;
Vatsavai P. ;
Guidry R.D., Jr. ;
Magnusen M.R. ;
Cowley A.D. ;
Martin C.K. .
BMC Research Notes, 9 (1)
[5]  
Sola J, 2019, P IEEE EMBC JUL 23 2
[6]   Measuring Pressure Introducing oBPM, the optical revolution for blood pressure monitoring [J].
Sola, Josep ;
Bertschi, Mattia ;
Krauss, Jens .
IEEE PULSE, 2018, 9 (05) :31-33
[7]   A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement [J].
Stergiou, George S. ;
Alpert, Bruce ;
Mieke, Stephan ;
Asmar, Roland ;
Atkins, Neil ;
Eckert, Siegfried ;
Frick, Gerhard ;
Friedman, Bruce ;
Grassl, Thomas ;
Ichikawa, Tsutomu ;
Ioannidis, John P. ;
Lacy, Peter ;
McManus, Richard ;
Murray, Alan ;
Myers, Martin ;
Palatini, Paolo ;
Parati, Gianfranco ;
Quinn, David ;
Sarkis, Josh ;
Shennan, Andrew ;
Usuda, Takashi ;
Wang, Jiguang ;
Wu, Colin O. ;
O'Brien, Eoin .
JOURNAL OF HYPERTENSION, 2018, 36 (03) :472-478