X-ray crystal spectrometer for opacity measurements in the 8-18 Å spectral range at the LULI laser facility

被引:7
作者
Reverdin, C. [1 ]
Thais, F. [2 ]
Loisel, G. [2 ]
Bougeard, M. [2 ]
机构
[1] CEA DAM DIF, F-91297 Arpajon, France
[2] CEA DSM IRAMIS, F-91191 Gif Sur Yvette, France
关键词
SPECTROSCOPY;
D O I
10.1063/1.3491285
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An x-ray crystal spectrometer was built in order to measure opacities in the 8-18 angstrom spectral range with an average spectral resolution of <lambda/delta lambda >similar to 400. It has been successfully used at the LULI-2000 laser facility (See C. Sauteret, rapport LULI 2001, 88 (2002) at Ecole Polytechnique (France) to measure in the same experimental conditions the 2p-3d transitions of several elements with the neighboring atomic number Z: Fe, Ni, Cu, and Ge [G. Loisel et al., High Energy Density Phys. 5, 173 (2009)]. Hence, a spectrometer with a wide spectral range is needed. This spectrometer features two lines of sight. In this example, one line of sight looks through the sample and the other one is looking directly at the backlighter emission. Both are outfitted with a spherical condensing mirror. A TlAP crystal is used for spectral dispersion. Detection is made with an image plate Fuji BAS TR2025, which is sensitive to x rays. We present some experimental results showing the performances of this spectrometer. (C) 2010 American Institute of Physics. [doi:10.1063/1.3491285]
引用
收藏
页数:5
相关论文
共 15 条
[1]   Extreme ultraviolet-vacuum ultraviolet spectrum detection using image plates [J].
Ben-Kish, A ;
Fisher, A ;
Cheifetz, E ;
Schwob, JL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2000, 71 (07) :2651-2654
[2]   DMX: An absolutely calibrated time-resolved broadband soft x-ray spectrometer designed for MJ class laser-produced plasmas (invited) [J].
Bourgade, JL ;
Villette, B ;
Bocher, JL ;
Boutin, JY ;
Chiche, S ;
Dague, N ;
Gontier, D ;
Jadaud, JP ;
Savale, B ;
Wrobel, R ;
Turner, RE .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (01) :1173-1182
[3]   A FOCUSING X-RAY CRYSTAL SPECTROGRAPH [J].
HALL, TA .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1984, 17 (02) :110-112
[4]  
HEETER RF, 2008, REV SCI INSTRUM, V79, pNI366
[5]   X-RAY INTERACTIONS - PHOTOABSORPTION, SCATTERING, TRANSMISSION, AND REFLECTION AT E=50-30,000 EV, Z=1-92 [J].
HENKE, BL ;
GULLIKSON, EM ;
DAVIS, JC .
ATOMIC DATA AND NUCLEAR DATA TABLES, 1993, 54 (02) :181-342
[6]   Comparison of film detectors, charged-coupled devices, and imaging plates in x-ray spectroscopy of hot dense plasma [J].
Howe, J ;
Chambers, DM ;
Courtois, C ;
Förster, E ;
Gregory, CD ;
Hall, IM ;
Renner, O ;
Uschmann, I ;
Woolsey, NC .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (03)
[7]   Absorption spectroscopy of mid and neighboring Z plasmas: Iron, nickel, copper and germanium [J].
Loisel, G. ;
Arnault, P. ;
Bastiani-Ceccotti, S. ;
Blenski, T. ;
Caillaud, T. ;
Fariaut, J. ;
Foelsner, W. ;
Gilleron, F. ;
Pain, J-C ;
Poirier, M. ;
Reverdin, C. ;
Silvert, V. ;
Thais, F. ;
Turck-Chieze, S. ;
Villette, B. .
HIGH ENERGY DENSITY PHYSICS, 2009, 5 (03) :173-181
[8]   Conical crystal spectrograph for high brightness x-ray Kα spectroscopy in subpicosecond laser-solid interaction [J].
Martinolli, E ;
Koenig, M ;
Boudenne, JM ;
Perelli, E ;
Batani, D ;
Hall, TA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (06) :2024-2028
[9]   Evaluation of the sensitivity and fading characteristics of an image plate system for x-ray diagnostics [J].
Meadowcroft, A. L. ;
Bentley, C. D. ;
Stott, E. N. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (11)
[10]   Extreme luminosity imaging conical spectrograph [J].
Pikuz, S. A. ;
Shelkovenko, T. A. ;
Mitchell, M. D. ;
Chandler, K. M. ;
Douglass, J. D. ;
McBride, R. D. ;
Jackson, D. P. ;
Hammer, D. A. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (10)