Effect of sintering on the marginal and internal fit of CAD/CAM-fabricated zirconia frameworks

被引:104
作者
Kunii, Jun [1 ]
Hotta, Yasuhiro [1 ]
Tamaki, Yukimichi [1 ]
Ozawa, Atushl [1 ]
Kobayashi, Yukitaka [1 ]
Fujishima, Akihiro [1 ]
Miyazaki, Takashi [1 ]
Fujiwara, Toshihisa [1 ]
机构
[1] Showa Univ, Sch Dent, Dept Oral Biomat & Technol, Shinagawa Ku, Tokyo 1428555, Japan
关键词
zirconia ceramics; CAD/CAM system; framework;
D O I
10.4012/dmj.26.820
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The aim of this study was to investigate the effect of post-machining sintering on marginal and internal fit of CAD/CAM-fabricated zirconia frameworks. Single crown copings (A: abutment), three-unit bridge frameworks (APA, P: pontic), four-unit bridge frameworks (APPA), and five-unit bridge frameworks (A(1)P(1)A(2)P(2)A(3)) were fabricated with raw-stage zirconia blanks using a commercial CAD/CAM system (KATANA (R), Noritake Dental Supply Co. Ltd., Aichi, Japan). Crown copings and frameworks were cemented to their respective master abutment models, and thickness of the cement layer was measured at specific measuring points. Marginal and internal fit of both APA and APPA were within clinical acceptance. However, the marginal gap and thickness of the cement layer on the axial surface of the pontic side of APA and APPA were slightly higher than those of the non-pontic side. As for the marginal gap of A(1)P(1)A(2)P(2)A(3) framework, it was superior to those of APA and APPA because the center abutment supported the framework to prevent distortion.
引用
收藏
页码:820 / 826
页数:7
相关论文
共 22 条
[1]   Marginal and internal fit of all-ceramic CAD/CAM crown-copings on chamfer preparations [J].
Bindl, A ;
Mörmann, WH .
JOURNAL OF ORAL REHABILITATION, 2005, 32 (06) :441-447
[2]  
Blatz MB, 2002, QUINTESSENCE INT, V33, P415
[3]   Porcelain-fused-to-metal vs. nonmetal crowns [J].
Christensen, GJ .
JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 1999, 130 (03) :409-411
[4]   CERAMIC VS PORCELAIN-FUSED-TO-METAL CROWNS - GIVE YOUR PATIENTS A CHOICE [J].
CHRISTENSEN, GJ .
JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 1994, 125 (03) :311-&
[5]  
Coli P, 2004, INT J PROSTHODONT, V17, P59
[6]  
FELTON DA, 1991, J PROSTHET DENT, V65, P375
[7]   Relative translucency of six all-ceramic systems. Part I: Core materials [J].
Heffernan, MJ ;
Aquilino, SA ;
Diaz-Arnold, AM ;
Haselton, DR ;
Stanford, CM ;
Vargas, MA .
JOURNAL OF PROSTHETIC DENTISTRY, 2002, 88 (01) :4-9
[8]   Relative translucency of six all-ceramic systems. Part II: Core and veneer materials [J].
Heffernan, MJ ;
Aquilino, SA ;
Diaz-Arnold, AM ;
Haselton, DR ;
Stanford, CM ;
Vargas, MA .
JOURNAL OF PROSTHETIC DENTISTRY, 2002, 88 (01) :10-15
[9]  
HERTLEIN G, 2005, MARGINAL FIT ZIRCONI
[10]   Durability of tungsten carbide burs for the fabrication of titanium crowns using dental CAD/CAM [J].
Hotta, Y ;
Miyazaki, T ;
Fujiwara, T ;
Tomita, S ;
Shinya, A ;
Sugai, Y ;
Ogura, H .
DENTAL MATERIALS JOURNAL, 2004, 23 (02) :190-196