Evaluation of a two-dimensional diode array for patient-specific quality assurance of HyperArc

被引:8
|
作者
Popple, Richard A. [1 ]
Sullivan, Rodney J. [1 ]
Yuan, Yuan [1 ]
Wu, Xingen [1 ]
Covington, Elizabeth L. [1 ]
机构
[1] Univ Alabama Birmingham, Dept Radiat Oncol, 176F RM 2240H,619 19TH St S,1720 2nd Ave South, Birmingham, AL 35249 USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2021年 / 22卷 / 12期
关键词
dosimetry; quality assurance; radiosurgery; ASTRO PRACTICE GUIDELINE; PLAN QUALITY; AMERICAN SOCIETY; SINGLE-ISOCENTER; FILM DOSIMETRY; RADIOLOGY ACR; RADIOSURGERY; DELIVERY; COLLEGE;
D O I
10.1002/acm2.13438
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To evaluate a two-dimensional diode array for patient-specific quality assurance of VMAT stereotactic radiosurgery (SRS) plans. Methods The diode array, an SRS MapCHECK (SRSMC), was composed of a 77 mm x77 mm face-centered array having a spacing of 2.47 mm. Sixty SRS plans were selected from our clinical database, 30 for treatment of a single target and 30 for multiple targets. The target sizes ranged from 2.4 mm to 44.7 mm equivalent diameter (median 8.7 mm). The plans were delivered to the diode array. For multiple target plans, two measurements were obtained at two locations, one corresponding to the largest target and the other to the smallest target. Gamma using a 3%/1 mm criteria and the dose to the center diode were compared with radiochromic film (RCF). Dose to selected regions of the detector electronics was calculated. Results The mean difference between the center diode and RCF was -1.2%. For a threshold of at least 95% of detectors/pixels having gamma < 1 for a 3%/1 mm criteria, SRSMC and RCF gave consistent results for 79 of the 90 measurements. For plans with an arc having a patient support angle of 90 degrees or 270 degrees, the median dose to the electronics was 0.65% of the prescription dose. Conclusions SRSMC is an efficient tool for accurate patient-specific quality assurance of VMAT single and multiple target radiosurgery, yielding similar clinical decisions as radiochromic film.
引用
收藏
页码:203 / 210
页数:8
相关论文
共 50 条
  • [1] Patient-Specific Quality Assurance in Pencil Beam Scanning by 2-Dimensional Array
    Rawiwan, Nuttida
    Chatchumnan, Nichakan
    Vimolnoch, Mananchaya
    Kingkaew, Sakda
    Oonsiri, Sornjarod
    INTERNATIONAL JOURNAL OF PARTICLE THERAPY, 2023, 10 (02) : 105 - 110
  • [2] Evaluation and Validation of IBA I'MatriXX Array for Patient-Specific Quality Assurance of TomoTherapy®
    Sresty, N. V. N. Madhusudhana
    Raju, A. Krishnam
    Reddy, B. Nagarjuna
    Sahithya, V. C.
    Mohmd, Yakub
    Kumar, G. Deleep
    Kumar, T. Anil
    Rushdi, Tasneem
    Aparna, S.
    Prasad, G. Durga
    Bajwa, Harjot Kaur
    JOURNAL OF MEDICAL PHYSICS, 2019, 44 (03) : 222 - 227
  • [3] Evaluation of Patient-Specific Quality Assurance Based On Different Methods
    Yang, T.
    Qu, B.
    Dai, X.
    Xie, C.
    Xiaoshen, W.
    Cong, X.
    Wang, J.
    Xu, S.
    MEDICAL PHYSICS, 2017, 44 (06) : 2896 - 2896
  • [4] Tornotherapy patient-specific quality assurance
    Chaudhari, S.
    Goddu, S.
    Mutic, S.
    Grigsby, J.
    Low, D.
    MEDICAL PHYSICS, 2007, 34 (06) : 2498 - 2499
  • [5] Evaluation of a Phantom for Patient-specific Quality Assurance in the MR Linac
    Baumgartl, M.
    Safarkhanlo, Y.
    Ehrbar, S.
    Wilke, L.
    Krayenbuhl, J.
    Chamberlain, M.
    Zamburlini, M.
    Guckenberger, M.
    Andratschke, N.
    Tanadini-Lang, S.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2021, 197 (SUPPL 1) : S33 - S34
  • [6] Evaluation of Patient-Specific Quality Assurance Based On Different Methods
    Lin, X.
    Sun, T.
    Liu, X.
    MEDICAL PHYSICS, 2018, 45 (06) : E211 - E211
  • [7] Dosimetric characterization of a new two-dimensional diode detector array used for stereotactic radiosurgery quality assurance
    Yasui, K.
    Saito, Y.
    Ogawa, S.
    Hayashi, N.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2021, 19 (02): : 281 - 289
  • [8] IMAT Patient Specific Quality Assurance Using ArcCHECK Diode Array Detector
    Thiyagarajan, Rajesh
    Sinha, S.
    Yadav, G.
    Ashokkumar, S.
    Raman, Kothanda
    Mishra, M.
    Nambiraj, N. Arunai
    MEDICAL PHYSICS, 2012, 39 (06) : 3733 - 3733
  • [9] Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance
    Narayanasamy, Ganesh
    Zalman, Travis
    Ha, Chul S.
    Papanikolaou, Niko
    Stathakis, Sotirios
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03): : 329 - 338
  • [10] Stereotactic body radiation therapy patient specific quality assurance using a two-dimensional array at extended source to surface distance
    Stathakis, Sotirios
    Mavroidis, Panayiotis
    Narayanasamy, Ganesh
    Markovic, Miljenko
    Myers, Pamela
    Papanikolaou, Niko
    JOURNAL OF BUON, 2015, 20 (04): : 1154 - 1163