Failure mode and effect analysis-based quality assurance for dynamic MLC tracking systems

被引:65
作者
Sawant, Amit [1 ]
Dieterich, Sonja [1 ]
Svatos, Michelle [2 ]
Keall, Paul [1 ]
机构
[1] Stanford Univ, Stanford, CA 94394 USA
[2] Varian Med Syst, Palo Alto, CA 94304 USA
关键词
MLC tracking; FMEA; quality assurance; commissioning; COLLIMATOR TARGET TRACKING; DMLC MOTION-TRACKING; RADIATION-THERAPY; MULTILEAF-COLLIMATOR; INTRAFRACTION MOTION; MANAGEMENT; RADIOTHERAPY; ALGORITHM; ACCURACY; DELIVERY;
D O I
10.1118/1.3517837
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To develop and implement a failure mode and effect analysis (FMEA)-based commissioning and quality assurance framework for dynamic multileaf collimator (DMLC) tumor tracking systems. Methods: A systematic failure mode and effect analysis was performed for a prototype real-time tumor tracking system that uses implanted electromagnetic transponders for tumor position monitoring and a DMLC for real-time beam adaptation. A detailed process tree of DMLC tracking delivery was created and potential tracking-specific failure modes were identified. For each failure mode, a risk probability number (RPN) was calculated from the product of the probability of occurrence, the severity of effect, and the detectibility of the failure. Based on the insights obtained from the FMEA, commissioning and QA procedures were developed to check (i) the accuracy of coordinate system transformation, (ii) system latency, (iii) spatial and dosimetric delivery accuracy, (iv) delivery efficiency, and (v) accuracy and consistency of system response to error conditions. The frequency of testing for each failure mode was determined from the RPN value. Results: Failures modes with RPN >= 125 were recommended to be tested monthly. Failure modes with RPN <125 were assigned to be tested during comprehensive evaluations, e.g., during commissioning, annual quality assurance, and after major software/hardware upgrades. System latency was determined to be similar to 193 ms. The system showed consistent and accurate response to erroneous conditions. Tracking accuracy was within 3%-3 mm gamma (100% pass rate) for sinusoidal as well as a wide variety of patient-derived respiratory motions. The total time taken for monthly QA was similar to 35 min, while that taken for comprehensive testing was similar to 3.5 h. Conclusions: FMEA proved to be a powerful and flexible tool to develop and implement a quality management (QM) framework for DMLC tracking. The authors conclude that the use of FMEA-based QM ensures efficient allocation of clinical resources because the most critical failure modes receive the most attention. It is expected that the set of guidelines proposed here will serve as a living document that is updated with the accumulation of progressively more intrainstitutional and interinstitutional experience with DMLC tracking. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3517837]
引用
收藏
页码:6466 / 6479
页数:14
相关论文
共 33 条
[11]   ELECTROMAGNETIC-GUIDED DYNAMIC MULTILEAF COLLIMATOR TRACKING ENABLES MOTION MANAGEMENT FOR INTENSITY-MODULATED ARC THERAPY [J].
Keall, Paul J. ;
Sawant, Amit ;
Cho, Byungchul ;
Ruan, Dan ;
Wu, Junqing ;
Poulsen, Per ;
Petersen, Jay ;
Newell, Laurence J. ;
Cattell, Herbert ;
Korreman, Stine .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 79 (01) :312-320
[12]   An Alternative Evaluation of FMEA: Fuzzy ART Algorithm [J].
Keskin, Guelsen Aydin ;
Ozkan, Coskun .
QUALITY AND RELIABILITY ENGINEERING INTERNATIONAL, 2009, 25 (06) :647-661
[13]   Task Group 142 report: Quality assurance of medical accelerators [J].
Klein, Eric E. ;
Hanley, Joseph ;
Bayouth, John ;
Yin, Fang-Fang ;
Simon, William ;
Dresser, Sean ;
Serago, Christopher ;
Aguirre, Francisco ;
Ma, Lijun ;
Arjomandy, Bijan ;
Liu, Chihray ;
Sandin, Carlos ;
Holmes, Todd .
MEDICAL PHYSICS, 2009, 36 (09) :4197-4212
[14]   COMPREHENSIVE QA FOR RADIATION ONCOLOGY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK-GROUP-40 [J].
KUTCHER, GJ ;
COIA, L ;
GILLIN, M ;
HANSON, WF ;
LEIBEL, S ;
MORTON, RJ ;
PALTA, JR ;
PURDY, JA ;
REINSTEIN, LE ;
SVENSSON, GK ;
WELLER, M ;
WINGFIELD, L .
MEDICAL PHYSICS, 1994, 21 (04) :581-618
[15]   RapidArc - Commissioning and QA procedures [J].
Ling, C. ;
Zhang, P. ;
Archambault, Y. ;
Bocanek, J. ;
Tang, G. ;
LoSasso, T. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01) :S594-S594
[16]   IMRT delivery performance with a varian multileaf collimator [J].
Losasso, Thomas .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (01) :S85-S88
[17]   Use of the 4D phantom to test real-time targeted radiation therapy device accuracy [J].
Malinowski, K. ;
Lechleiter, K. ;
Hubenschmidt, J. ;
Low, D. ;
Parikh, P. .
MEDICAL PHYSICS, 2007, 34 (06) :2611-2611
[18]  
OELFKE U, 2009, MED PHYS, V37, P753
[19]   Modified failure mode and effects analysis using approximate reasoning [J].
Pillay, A ;
Wang, J .
RELIABILITY ENGINEERING & SYSTEM SAFETY, 2003, 79 (01) :69-85
[20]   Tools for developing a quality management program: Proactive tools (process mapping, value stream mapping, fault tree analysis, and failure mode and effects analysis) [J].
Rath, Frank .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (01) :S187-S190