Intraoperative Intracochlear Electrocochleography and Residual Hearing Preservation Outcomes When Using Two Types of Slim Electrode Arrays in Cochlear Implantation

被引:32
作者
Ramos-Macias, Angel [1 ]
O'Leary, Stephen [2 ]
Ramos-deMiguel, Angel [1 ]
Bester, Christo [2 ]
Falcon-Gonzalez, Juan C. [1 ]
机构
[1] Las Palmas Univ, Dept Otolaryngol, Las Palmas Univ Hosp, Hearing & Balance BioEngineer Lab, Avda Maritma Sur S-N, Las Palmas Gran Canaria 35016, Spain
[2] Univ Melbourne, Dept Otolaryngol, Melbourne Univ Hosp, Melbourne, Vic, Australia
关键词
Cochlear implant; Electrocochleography; Residual hearing; ROUND WINDOW ELECTROCOCHLEOGRAPHY; SPEECH; TRAUMA;
D O I
10.1097/MAO.0000000000002212
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objectives: To report residual hearing preservation outcomes in patients with low frequency hearing, after cochlear implant (CI) electrode insertion with two types of electrode arrays: one straight and other perimodiolar, when using intraoperative intracochlear electrocochleography (ECochG) during (CI) electrode insertion. Study Design: Prospective, randomized study. Setting: Tertiary referral otology center. Patients: Fifteen patients ranging from 33 to 54 years old (mean 51.19). They had been diagnosed with a bilateral, profound sensorineural hearing loss and treated with a unilateral cochlear implant: eight of them with the CI532 and seven of them with the CI522 (Cochlear Ltd, Sydney, Australia). Intervention: Pure-tone audiometry was performed preoperatively and at 1 and 6 months postoperatively. Interoperatively, intracochlear ECochG was performed using the apical-most electrode. The amplitude of the first harmonic was plotted and monitored in real time by the audiologist-surgeon team during their CI electrode insertion. The different ECoch patterns of the insertion track were recorded and analyzed. Results: In 12 cases ECochG responses were successfully recorded. In three cases no ECochG responses could be recorded with no residual hearing observed postoperatively in two of them. With respect to the first harmonic amplitude changes, we found: four cases with an overall increase in amplitude measured from the beginning of insertion until completion, all of them showed residual hearing (<15 dB HL) at 6 months postoperation. Three cases with an increasing amplitude at the beginning of insertion, with a decrease in amplitude as insertion progressed to completion, in two cases dropping of residual hearing (15-30 dB HL) were observed after 6 months postoperation and, in one case, complete residual hearing was observed at 6 months postoperatively. And finally five cases presented amplitudes at the start of insertion with modifications of amplitude during the insertion dynamic, with increasing and descending in amplitude range during the whole insertion, two of them showed residual hearing at 6 months postoperation and three cases a drop of residual hearing (15-30 dB HL) was observed after 6 months postoperation. No statistical differences between CI532 and CI522 electrodes were found. Data of the ECochG responses are also presented (p value >= 0.05). Conclusion: ECochG is a useful tool to evaluate the residual hearing in CI patients with straight and perimodiolar cochlear implant. More studies are needed to fully understand the relationship between ECochG and the presence of residual hearing, cochlear trauma, and functional outcomes.
引用
收藏
页码:S29 / S37
页数:9
相关论文
共 25 条
  • [1] Round window electrocochleography before and after cochlear implant electrode insertion
    Adunka, Oliver F.
    Giardina, Christopher K.
    Formeister, Eric J.
    Choudhury, Baishakhi
    Buchman, Craig A.
    Fitzpatrick, Douglas C.
    [J]. LARYNGOSCOPE, 2016, 126 (05) : 1193 - 1200
  • [2] Characterizing Electrocochleography in Cochlear Implant Recipients with Residual Low-Frequency Hearing
    Bester, Christopher W.
    Campbell, Luke
    Dragovic, Adrian
    Collins, Aaron
    O'Leary, Stephen J.
    [J]. FRONTIERS IN NEUROSCIENCE, 2017, 11
  • [3] Calloway NH, 2014, OTOL NEUROTOL, V35, P1451, DOI 10.1097/MAO.0000000000000451
  • [4] Intraoperative Real-time Cochlear Response Telemetry Predicts Hearing Preservation in Cochlear Implantation
    Campbell, Luke
    Kaicer, Arielle
    Sly, David
    Iseli, Claire
    Wei, Benjamin
    Briggs, Robert
    O'Leary, Stephen
    [J]. OTOLOGY & NEUROTOLOGY, 2016, 37 (04) : 332 - 338
  • [5] Cochlear Response Telemetry: Intracochlear Electrocochleography via Cochlear Implant Neural Response Telemetry Pilot Study Results
    Campbell, Luke
    Kaicer, Arielle
    Briggs, Robert
    O'Leary, Stephen
    [J]. OTOLOGY & NEUROTOLOGY, 2015, 36 (03) : 399 - 405
  • [6] Implications of Minimizing Trauma During Conventional Cochlear Implantation
    Carlson, Matthew L.
    Driscoll, Colin L. W.
    Gifford, Rene H.
    Service, Geoffrey J.
    Tombers, Nicole M.
    Hughes-Borst, Becky J.
    Neff, Brian A.
    Beatty, Charles W.
    [J]. OTOLOGY & NEUROTOLOGY, 2011, 32 (06) : 962 - 968
  • [7] Chakravorti S, 2017, J MED IMAGING, V4, DOI 10.1117/1.JMI.4.4.045002
  • [8] Assessment of Cochlear Trauma During Cochlear Implantation Using Electrocochleography and Cone Beam Computed Tomography
    Dalbert, Adrian
    Huber, Alexander
    Veraguth, Dorothe
    Roosli, Christof
    Pfiffner, Flurin
    [J]. OTOLOGY & NEUROTOLOGY, 2016, 37 (05) : 446 - 453
  • [9] Extra- and Intracochlear Electrocochleography in Cochlear Implant Recipients
    Dalbert, Adrian
    Pfiffner, Flurin
    Roeoesli, Christof
    Thoele, Konrad
    Sim, Jae Hoon
    Gerig, Rahel
    Huber, Alexander M.
    [J]. AUDIOLOGY AND NEURO-OTOLOGY, 2015, 20 (05) : 339 - 348
  • [10] Correlation of Electrophysiological Properties and Hearing Preservation in Cochlear Implant Patients
    Dalbert, Adrian
    Sim, Jae Hoon
    Gerig, Rahel
    Pfiffner, Flurin
    Roosli, Christof
    Huber, Alexander
    [J]. OTOLOGY & NEUROTOLOGY, 2015, 36 (07) : 1172 - 1180