A robotic C-arm cone beam CT system for image-guided proton therapy: design and performance

被引:34
作者
Hua, Chiaho [1 ]
Yao, Weiguang [1 ]
Kidani, Takao [2 ]
Tomida, Kazuo [2 ]
Ozawa, Saori [2 ]
Nishimura, Takenori [2 ]
Fujisawa, Tatsuya [2 ]
Shinagawa, Ryousuke [2 ]
Merchant, Thomas E. [1 ]
机构
[1] St Jude Childrens Res Hosp, Dept Radiat Oncol, 332 N Lauderdale St, Memphis, TN 38105 USA
[2] Hitachi Ltd, Proton Therapy Syst Design Dept, Hitachi, Ibaraki, Japan
关键词
COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; PARTICLE THERAPY; DEFORMABLE REGISTRATION; CONFORMAL RADIOTHERAPY; LUNG-CANCER; HEAD; CBCT;
D O I
10.1259/bjr.20170266
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: A ceiling-mounted robotic C-arm cone beam CT (CBCT) system was developed for use with a 190 degrees proton gantry system and a 6-degree-of-freedom robotic patient positioner. We report on the mechanical design, system accuracy, image quality, image guidance accuracy, imaging dose, workflow, safety and collision-avoidance. Methods: The robotic CBCT system couples a rotating C-ring to the C-arm concentrically with a kV X-ray tube and a flat-panel imager mounted to the C-ring. CBCT images are acquired with flex correction and maximally 360 degrees rotation for a 53 cm field of view. The system was designed for clinical use with three imaging locations. Anthropomorphic phantoms were imaged to evaluate the image guidance accuracy. Results: The position accuracy and repeatability of the robotic C-arm was high (< 0.5 mm), as measured with a high-accuracy laser tracker. The isocentric accuracy of the C-ring rotation was within 0.7 mm. The coincidence of CBCT imaging and radiation isocentre was better than 1 mm. The average image guidance accuracy was within 1 mm and 1 degrees for the anthropomorphic phantoms tested. Daily volumetric imaging for proton patient positioning was specified for routine clinical practice. Conclusion: Our novel gantry-independent robotic CBCT system provides high-accuracy volumetric image guidance for proton therapy.
引用
收藏
页数:9
相关论文
共 27 条
[1]  
[Anonymous], J APPL CLIN MED PHYS
[2]   Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179 [J].
Bissonnette, Jean-Pierre ;
Balter, Peter A. ;
Dong, Lei ;
Langen, Katja M. ;
Lovelock, D. Michael ;
Miften, Moyed ;
Moseley, Douglas J. ;
Pouliot, Jean ;
Sonke, Jan-Jakob ;
Yoo, Sua .
MEDICAL PHYSICS, 2012, 39 (04) :1946-1963
[3]  
Chan Mf, 2011, Biomed Imaging Interv J, V7, pe11, DOI 10.2349/biij.7.2.e11
[4]   A novel dose-based positioning method for CT image-guided proton therapy [J].
Cheung, Joey P. ;
Park, Peter C. ;
Court, Laurence E. ;
Zhu, X. Ronald ;
Kudchadker, Rajat J. ;
Frank, Steven J. ;
Dong, Lei .
MEDICAL PHYSICS, 2013, 40 (05)
[5]   Physics Controversies in Proton Therapy [J].
Engelsman, Martijn ;
Schwarz, Marco ;
Dong, Lei .
SEMINARS IN RADIATION ONCOLOGY, 2013, 23 (02) :88-96
[6]   Image guided particle therapy in CNAO room 2: Implementation and clinical validation [J].
Fattori, G. ;
Riboldi, M. ;
Pella, A. ;
Peroni, M. ;
Cerveri, P. ;
Desplanques, M. ;
Fontana, G. ;
Tagaste, B. ;
Valvo, F. ;
Orecchia, R. ;
Baroni, G. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (01) :9-15
[7]   C-arm cone-beam computed tomography in interventional oncology: technical aspects and clinical applications [J].
Floridi, Chiara ;
Radaelli, Alessandro ;
Abi-Jaoudeh, Nadine ;
Grass, Micheal ;
De Lin, Ming ;
Chiaradia, Melanie ;
Geschwind, Jean-Francois ;
Kobeiter, Hishman ;
Squillaci, Ettore ;
Maleux, Geert ;
Giovagnoni, Andrea ;
Brunese, Luca ;
Wood, Bradford ;
Carrafiello, Gianpaolo ;
Rotondo, Antonio .
RADIOLOGIA MEDICA, 2014, 119 (07) :521-532
[8]   EFFECTS OF INTERFRACTIONAL, MOTION AND ANATOMIC CHANGES ON PROTON THERAPY DOSE DISTRIBUTION IN LUNG CANCER [J].
Hui, Zhouguang ;
Zhang, Xiaodong ;
Starkschall, George ;
Li, Yupeng ;
Mohan, Radhe ;
Komaki, Ritsuko ;
Cox, James D. ;
Chang, Joe Y. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (05) :1385-1395
[9]   Flat-panel cone-beam computed tomography for image-guided radiation therapy [J].
Jaffray, DA ;
Siewerdsen, JH ;
Wong, JW ;
Martinez, AA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05) :1337-1349
[10]  
Kaissl W, 2015, 54 ANN C PART THER C