Determination of the copper (II) by a chemically modified screen-printing carbon electrode

被引:0
作者
Tu, YF [1 ]
Xu, J
机构
[1] Suzhou Univ, Dept Chem & Chem Engn, Suzhou 215006, Peoples R China
[2] Yangzhou Univ, Coll Irrigat & Architectural Works, Yangzhou 225009, Peoples R China
关键词
screen-printing; chemically modified electrode; anodic stripping voltammetry; copper; tap water;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The technique for developing a screen-printing disposable carbon electrode and modifying its surface with electropolymer of catechol-formaldehyde was described. This modified electrode could be used to determine the Cu2+ with high selectivity. The linear range of the determination was from 2.5 mug/L to 50 mug/L. The linear regression coefficient and the RSD were 0.9992 and 2.7%, respectively. The content of Cu2+ in local tap water was determined by this modified electrode for 6.27 mug/L.
引用
收藏
页码:109 / 111
页数:3
相关论文
共 4 条
  • [1] JING LT, 1992, CHEM MODIFIED ELECTR, P9
  • [2] WANG LZ, 1994, CHINESE J ANAL CHEM, V22, P1185
  • [3] CHEMICALLY MODIFIED, SCREEN-PRINTED CARBON ELECTRODES
    WRING, SA
    HART, JP
    [J]. ANALYST, 1992, 117 (08) : 1281 - 1286
  • [4] XU JR, 1995, ELECTROCHEMISTRY, V1, P209