MRI-based transfer function determination for the assessment of implant safety

被引:26
|
作者
Tokaya, J. P. [1 ]
Raaijmakers, A. J. E. [1 ,2 ]
Luijten, P. R. [3 ]
Bakker, J. F. [4 ]
van den Berg, C. A. T. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiotherapy, Heidelberglaan 100,POB 85500, NL-3508 GA Utrecht, Netherlands
[2] Eindhoven Univ Technol, Biomed Image Anal, Eindhoven, Netherlands
[3] Univ Med Ctr Utrecht, Dept Radiol, Utrecht, Netherlands
[4] Medtron Eindhoven Design Ctr, Eindhoven, Netherlands
关键词
active implantable medical device (AIMD); safety; transfer function; EM simulations; RF heating; RF SAFETY; DEVICES; LEADS; WIRES;
D O I
10.1002/mrm.26613
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeWe introduce a new MR-based method to determine the transfer function (TF) for radiofrequency (RF) safety assessment of active implantable medical devices. Transfer functions are implant-specific measures that relate the incident tangential electric field on an (elongated) implant to a scattered electric field at its tip. The proposed method allows for TF determination with a high spatial resolution in relatively fast measurements without requiring dedicated bench setups from MRI images. Theory and MethodsThe principle of reciprocity is used in conjunction with the potential to measure currents with MRI to determine TF. Low-flip angle 3D dual gradient echo MRI data are acquired with an implant as transceive antenna, which requires minimal hardware adaptations. The implant-specific TF is determined from the acquired MRI data, with two different postprocessing methods for comparison. ResultsTFs of linear and helical implants can be determined accurately (with a Pearson correlation coefficient R0.7 between measurements and simulations, and a difference in field at the tip Etip19%) from relatively quick (t<20minutes) MRI acquisitions with (several) millimeter spatial resolution. ConclusionTransfer function determination with MRI for RF safety assessment of implantable medical devices is possible. The proposed MR-based method allows for TF determination in more realistic exposure scenarios and solid media. Magn Reson Med 78:2449-2459, 2017. (c) 2017 International Society for Magnetic Resonance in Medicine
引用
收藏
页码:2449 / 2459
页数:11
相关论文
共 50 条
  • [21] Two-phase survey on the frequency of use and safety of MRI for hearing implant recipients
    van de Heyning, Paul
    Mertens, Griet
    Topsakal, Vedat
    de Brito, Ruben
    Wimmer, Wilhelm
    Caversaccio, Marco D.
    Dazert, Stefan
    Volkenstein, Stefan
    Zernotti, Mario
    Parnes, Lorne S.
    Staecker, Hinrich
    Bruce, Iain A.
    Rajan, Gunesh
    Atlas, Marcus
    Friedland, Peter
    Skarzynski, Piotr H.
    Sugarova, Serafima
    Kuzovkov, Vladislav
    Hagr, Abdulrahman
    Mlynski, Robert
    Schmutzhard, Joachim
    Usami, Shin-Ichi
    Lassaletta, Luis
    Gavilan, Javier
    Godey, Benoit
    Raine, Christopher H.
    Hagen, Rudolf
    Sprinzl, Georg M.
    Brown, Kevin
    Baumgartner, Wolf-Dieter
    Karltorp, Eva
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2021, 278 (11) : 4225 - 4233
  • [22] Safety assessment of spine MRI in deep brain stimulation patients
    Boutet, Alexandre
    Elias, Gavin J. B.
    Gramer, Robert
    Neurdorfer, Clemens
    Germann, Jurgen
    Naheed, Asma
    Bennett, Nicole
    Li, Bryan
    Gwun, Dave
    Chow, Clement T.
    Maciel, Ricardo
    Valencia, Alejandro
    Fasano, Alfonso
    Munhoz, Renato P.
    Foltz, Warren
    Mikulis, David
    Hancu, Ileana
    Kalia, Suneil K.
    Hodaie, Mojgan
    Kucharczyk, Walter
    Lozano, Andres M.
    JOURNAL OF NEUROSURGERY-SPINE, 2020, 32 (06) : 973 - 983
  • [23] Modeling and Characterization of the Implant Intra-Body Communication Based on Capacitive Coupling Using a Transfer Function Method
    Zhang, Kai
    Hao, Qun
    Song, Yong
    Wang, Jingwen
    Huang, Ruobing
    Liu, Yue
    SENSORS, 2014, 14 (01): : 1740 - 1756
  • [24] Structural effects of intra-articular TGF-β1 in moderate to advanced knee osteoarthritis: MRI-based assessment in a randomized controlled trial
    Guermazi, A.
    Kalsi, G.
    Niu, J.
    Crema, M. D.
    Copeland, R. O.
    Orlando, A.
    Noh, M. J.
    Roemer, F. W.
    BMC MUSCULOSKELETAL DISORDERS, 2017, 18
  • [25] Acoustic Noise Transfer Function in Clinical MRI: A Multicenter Analysis
    Hamaguchi, Takashi
    Miyati, Tosiaki
    Ohno, Naoki
    Hirano, Masaya
    Hayashi, Norio
    Gabata, Toshifumi
    Matsui, Osamu
    Matsushita, Tatsuhiko
    Yamamoto, Tomoyuki
    Fujiwara, Yasuhiro
    Kimura, Hirohiko
    Takeda, Hiroyasu
    Takehara, Yasuo
    ACADEMIC RADIOLOGY, 2011, 18 (01) : 101 - 106
  • [26] A framework for developing generic implant safety procedures for scanning patients with medical implants and devices in MRI
    Ashmore, Jonathan P.
    Prescott, Sarah J.
    Mclean, John
    Wilson, Daniel J.
    Charles-Edwards, Geoff
    Wright, Peter
    Grainger, David
    Barker, Gareth J.
    Lipton, Alexandra J.
    Watt, Rachel
    Gopalan, Deepa
    Radon, Mark R.
    BRITISH JOURNAL OF RADIOLOGY, 2025, : 336 - 344
  • [27] Assessment of flaw detectability using transfer function
    Koshti, Ajay M.
    NONDESTRUCTIVE CHARACTERIZATION AND MONITORING OF ADVANCED MATERIALS, AEROSPACE, CIVIL INFRASTRUCTURE, AND TRANSPORTATION XV, 2021, 11592
  • [28] Determination of transfer function for a superconducting gravimeter at Wuhan station
    Chen X.
    Yan R.
    Niu X.
    Li H.
    Zhang M.
    Sun H.
    Huazhong Keji Daxue Xuebao (Ziran Kexue Ban)/Journal of Huazhong University of Science and Technology (Natural Science Edition), 2022, 50 (09): : 35 - 40
  • [29] Determination of the electronics transfer function for current transient measurements
    Scharf, Christian
    Klanner, Robert
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 779 : 1 - 5
  • [30] Preclinical Safety Assessment of Gadopiclenol: A High-Relaxivity Macrocyclic Gadolinium-Based MRI Contrast Agent
    Gendron, Celia
    Bourrinet, Philippe
    Dencausse, Anne
    Fretellier, Nathalie
    INVESTIGATIVE RADIOLOGY, 2024, 59 (02) : 108 - 123