MRI-based transfer function determination for the assessment of implant safety

被引:26
|
作者
Tokaya, J. P. [1 ]
Raaijmakers, A. J. E. [1 ,2 ]
Luijten, P. R. [3 ]
Bakker, J. F. [4 ]
van den Berg, C. A. T. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiotherapy, Heidelberglaan 100,POB 85500, NL-3508 GA Utrecht, Netherlands
[2] Eindhoven Univ Technol, Biomed Image Anal, Eindhoven, Netherlands
[3] Univ Med Ctr Utrecht, Dept Radiol, Utrecht, Netherlands
[4] Medtron Eindhoven Design Ctr, Eindhoven, Netherlands
关键词
active implantable medical device (AIMD); safety; transfer function; EM simulations; RF heating; RF SAFETY; DEVICES; LEADS; WIRES;
D O I
10.1002/mrm.26613
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeWe introduce a new MR-based method to determine the transfer function (TF) for radiofrequency (RF) safety assessment of active implantable medical devices. Transfer functions are implant-specific measures that relate the incident tangential electric field on an (elongated) implant to a scattered electric field at its tip. The proposed method allows for TF determination with a high spatial resolution in relatively fast measurements without requiring dedicated bench setups from MRI images. Theory and MethodsThe principle of reciprocity is used in conjunction with the potential to measure currents with MRI to determine TF. Low-flip angle 3D dual gradient echo MRI data are acquired with an implant as transceive antenna, which requires minimal hardware adaptations. The implant-specific TF is determined from the acquired MRI data, with two different postprocessing methods for comparison. ResultsTFs of linear and helical implants can be determined accurately (with a Pearson correlation coefficient R0.7 between measurements and simulations, and a difference in field at the tip Etip19%) from relatively quick (t<20minutes) MRI acquisitions with (several) millimeter spatial resolution. ConclusionTransfer function determination with MRI for RF safety assessment of implantable medical devices is possible. The proposed MR-based method allows for TF determination in more realistic exposure scenarios and solid media. Magn Reson Med 78:2449-2459, 2017. (c) 2017 International Society for Magnetic Resonance in Medicine
引用
收藏
页码:2449 / 2459
页数:11
相关论文
共 50 条
  • [1] MRI-based, wireless determination of the transfer function of a linear implant: Introduction of the transfer matrix
    Tokaya, Janot P.
    Raaijmakers, Alexander J. E.
    Luijten, Peter R.
    van den Berg, Cornelis A. T.
    MAGNETIC RESONANCE IN MEDICINE, 2018, 80 (06) : 2771 - 2784
  • [2] MRI-based transfer function determination through the transfer matrix by jointly fitting the incident and scattered B1+field
    Tokaya, Janot P.
    Raaijmakers, Alexander J. E.
    Luijten, Peter R.
    Sbrizzi, Alessandro
    van den Berg, Cornelis A. T.
    MAGNETIC RESONANCE IN MEDICINE, 2020, 83 (03) : 1081 - 1095
  • [3] Simplified Transfer Function Assessment of Implantable Leads for MRI Safety Evaluations
    Feng, Shi
    Wang, Qingyan
    Wen, Chenxi
    Zheng, Jianfeng
    Kainz, Wolfgang
    Chen, Ji
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2019, 61 (05) : 1432 - 1437
  • [4] A single setup approach for the MRI-based measurement and validation of the transfer function of elongated medical implants
    Stijnman, Peter R. S.
    Erturk, M. Arcan
    van den Berg, Cornelis A. T.
    Raaijmakers, Alexander J. E.
    MAGNETIC RESONANCE IN MEDICINE, 2021, 86 (05) : 2751 - 2765
  • [5] Coupled transfer function model for the evaluation of implanted cables safety in MRI
    Kabil, Julie
    Felblinger, Jacques
    Vuissoz, Pierre-Andre
    Missoffe, Alexia
    MAGNETIC RESONANCE IN MEDICINE, 2020, 84 (02) : 991 - 999
  • [6] Towards an integrated radiofrequency safety concept for implant carriers in MRI based on sensor-equipped implants and parallel transmission
    Petzold, Johannes
    Schmitter, Sebastian
    Silemek, Berk
    Winter, Lukas
    Speck, Oliver
    Ittermann, Bernd
    Seifert, Frank
    NMR IN BIOMEDICINE, 2023, 36 (07)
  • [7] A machine learning-based approach for RF transfer function modeling of active implantable medical electrodes at 3T MRI
    Yao, Aiping
    Ma, Mingjuan
    Shi, Hexuan
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (17)
  • [8] Measurement, simulation and uncertainty assessment of implant heating during MRI
    Neufeld, E.
    Kuehn, S.
    Szekely, G.
    Kuster, N.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (13) : 4151 - 4169
  • [9] A Transfer Function Measurement Setup With an Electro-Optic Sensor for MR Safety Assessment in Cascaded Media
    Lottner, Thomas
    Reiss, Simon
    Bitzer, Andreas
    Bock, Michael
    Oezen, Ali Caglar
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2021, 63 (03) : 662 - 672
  • [10] Neurosurgical Implant Safety in 7 T MRI: A Scoping Review
    Shaffer, Annabelle
    Weisbaum, David
    Naik, Anant
    Anderson, Aaron
    Wszalek, Tracey
    Cohen, Mark
    Sutton, Brad
    Webb, Andrew
    Damon, Bruce
    Arnold, Paul M.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2023, 57 (03) : 661 - 669