Optical function of intraocular lenses in different opacification patterns: metrology analysis of 67 explants

被引:11
作者
Labuz, Grzegorz [1 ]
Yildirim, Timur M. [1 ]
Khoramnia, Ramin [1 ]
Son, Hyeck-Soo [1 ]
Auffarth, Gerd U. [1 ]
机构
[1] Univ Hosp Heidelberg, Dept Ophthalmol, David J Apple Ctr Vis Res, Neuenheimer Feld 400, D-69120 Heidelberg, Germany
关键词
LATE POSTOPERATIVE OPACIFICATION; LIGHT-SCATTERING; VISUAL-ACUITY; LOCALIZED OPACIFICATION; OCULAR STRAYLIGHT; AIR INJECTION; CALCIFICATION; QUALITY; TRANSMITTANCE; PERFORMANCE;
D O I
10.1097/j.jcrs.0000000000000553
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To determine how the type of opacification would have compromised the function of intraocular lenses (IOLs) prior to explantation using metrology. Setting: David J Apple Laboratory, Heidelberg, Germany. Design: Laboratory investigation. Methods: Of 67 explanted IOLs, 28 with homogeneous calcification, 21 with localized calcification, and 18 subluxated IOLs without calcification that served as controls were identified. The modulation transfer function (MTF) cutoff (0.43 at 100 lp/mm) was used to define decreased optical quality, following the manufacturers' guidelines for IOL testing. Light scattering was evaluated in vitro using a clinical device. Results: Only 1 control IOL demonstrated a decreased MTF compared with 4 and 15 in the homogeneous and localized groups, respectively. The MTF-derived metrics did not differ between the homogeneous and control las (P = .99), but both showed better performance than those with localized opacification (P < .001). The median straylight parameter in the homogeneous pattern was 181.8 deg(2)/sr (108.5 to 244.1 deg(2)/sr), but in the localized group, it was less [69.8 (17.7 to 250.8) deg(2)/sr] (P = .02). Both opacification patterns yielded a significant straylight increase compared with the controls (P < .001). Conclusions: The study demonstrated that optical quality differed between the 2 types of opacification. The centrally localized pattern showed a significant MTF reduction, indicating a larger potential to affect patient visual acuity. Although localized calcification demonstrated lower straylight than that found in the homogeneous form, both conditions may cause severe glare phenomena.
引用
收藏
页码:1210 / 1217
页数:8
相关论文
共 49 条
[11]   AN IMPROVED MATHEMATICAL-DESCRIPTION OF THE FOVEAL VISUAL POINT SPREAD FUNCTION WITH PARAMETERS FOR AGE, PUPIL SIZE AND PIGMENTATION [J].
IJSPEERT, JK ;
VANDENBERG, TJTP ;
SPEKREIJSE, H .
VISION RESEARCH, 1993, 33 (01) :15-20
[12]   Microscopic Characteristics of Late Intraocular Lens Opacifications [J].
Kanclerz, Piotr ;
Yildirim, Timur Mert ;
Khoramnia, Ramin .
ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2021, 145 (06) :759-767
[13]  
Khoramnia R, 2012, OPHTHALMOLOGE, V109, P483, DOI 10.1007/s00347-011-2487-6
[14]   Trifocality Achieved Through Polypseudophakia: Optical Quality and Light Loss Compared With a Single Trifocal Intraocular Lens [J].
Labuz, Grzegorz ;
Auffarth, Gerd U. ;
Knorz, Michael C. ;
Son, Hyeck-Soo ;
Yildirim, Timur M. ;
Khoramnia, Ramin .
JOURNAL OF REFRACTIVE SURGERY, 2020, 36 (09) :570-577
[15]   The Effect of a Spectral Filter on Visual Quality in Patients with an Extended-Depth-Of-Focus Intraocular Lens [J].
Labuz, Grzegorz ;
Auffarth, Gerd U. ;
Ozen, Aydin ;
Van den Berg, Thomas J. T. P. ;
Yildirim, Timur M. ;
Son, Hyeck-Soo ;
Khoramnia, Ramin .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2019, 208 :56-63
[16]   Longitudinal Chromatic Aberration and Polychromatic Image Quality Metrics of Intraocular Lenses [J].
Labuz, Grzegorz ;
Papadatou, Eleni ;
Khoramnia, Ramin ;
Auffarth, Gerd U. .
JOURNAL OF REFRACTIVE SURGERY, 2018, 34 (12) :832-+
[17]   Assessment of straylight and the modulation transfer function of intraocular lenses with centrally localized opacification associated with the intraocular injection of gas [J].
Labuz, Grzegorz ;
Yildirim, Timur M. ;
van den Berg, Thomas J. T. P. ;
Khoramnia, Ramin ;
Auffarth, Gerd U. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2018, 44 (05) :615-622
[18]   Light scattering levels from intraocular lenses extracted from donor eyes [J].
Labuz, Grzegorz ;
Reus, Nicolaas J. ;
van den Berg, Thomas J. T. P. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2017, 43 (09) :1207-1212
[19]   Validation of a spectral light scattering method to differentiate large from small particles in intraocular lenses [J].
Labuz, Grzegorz ;
Papadatou, Eleni ;
Vargas-Martin, Fernando ;
Lopez-Gil, Norberto ;
Reus, Nicolaas J. ;
van den Berg, Thomas J. T. P. .
BIOMEDICAL OPTICS EXPRESS, 2017, 8 (03) :1889-1894
[20]   Method for in vitro assessment of straylight from intraocular lenses [J].
Labuz, Grzegorz ;
Vargas-Martin, Fernando ;
van den Berg, Thomas J. T. P. ;
Lopez-Gil, Norberto .
BIOMEDICAL OPTICS EXPRESS, 2015, 6 (11) :4457-4464