Optical function of intraocular lenses in different opacification patterns: metrology analysis of 67 explants

被引:11
作者
Labuz, Grzegorz [1 ]
Yildirim, Timur M. [1 ]
Khoramnia, Ramin [1 ]
Son, Hyeck-Soo [1 ]
Auffarth, Gerd U. [1 ]
机构
[1] Univ Hosp Heidelberg, Dept Ophthalmol, David J Apple Ctr Vis Res, Neuenheimer Feld 400, D-69120 Heidelberg, Germany
关键词
LATE POSTOPERATIVE OPACIFICATION; LIGHT-SCATTERING; VISUAL-ACUITY; LOCALIZED OPACIFICATION; OCULAR STRAYLIGHT; AIR INJECTION; CALCIFICATION; QUALITY; TRANSMITTANCE; PERFORMANCE;
D O I
10.1097/j.jcrs.0000000000000553
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To determine how the type of opacification would have compromised the function of intraocular lenses (IOLs) prior to explantation using metrology. Setting: David J Apple Laboratory, Heidelberg, Germany. Design: Laboratory investigation. Methods: Of 67 explanted IOLs, 28 with homogeneous calcification, 21 with localized calcification, and 18 subluxated IOLs without calcification that served as controls were identified. The modulation transfer function (MTF) cutoff (0.43 at 100 lp/mm) was used to define decreased optical quality, following the manufacturers' guidelines for IOL testing. Light scattering was evaluated in vitro using a clinical device. Results: Only 1 control IOL demonstrated a decreased MTF compared with 4 and 15 in the homogeneous and localized groups, respectively. The MTF-derived metrics did not differ between the homogeneous and control las (P = .99), but both showed better performance than those with localized opacification (P < .001). The median straylight parameter in the homogeneous pattern was 181.8 deg(2)/sr (108.5 to 244.1 deg(2)/sr), but in the localized group, it was less [69.8 (17.7 to 250.8) deg(2)/sr] (P = .02). Both opacification patterns yielded a significant straylight increase compared with the controls (P < .001). Conclusions: The study demonstrated that optical quality differed between the 2 types of opacification. The centrally localized pattern showed a significant MTF reduction, indicating a larger potential to affect patient visual acuity. Although localized calcification demonstrated lower straylight than that found in the homogeneous form, both conditions may cause severe glare phenomena.
引用
收藏
页码:1210 / 1217
页数:8
相关论文
共 49 条
[1]  
[Anonymous], 2014, ISO119792
[2]   Optical quality of intraocular lenses after several years in the human eye [J].
Artigas, J. M. ;
Felipe, A. ;
Diez-Ajenjo, A. ;
Garcia-Domene, M. C. ;
Capilla, P. ;
Desco, M. C. .
JOURNAL OF MODERN OPTICS, 2013, 60 (05) :399-405
[3]   Light scattering and light transmittance in a series of calcified single-piece hydrophilic acrylic intraocular lenses of the same design [J].
Barra, Daniel ;
Werner, Liliana ;
Pacini Costa, Joao Luiz ;
Morris, Caleb ;
Ribeiro, Tiago ;
Ventura, Bruna Vieira ;
Dornelles, Fabio .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (01) :121-128
[4]   Late postoperative opacification of a hydrophilic-hydrophobic acrylic intraocular lens [J].
Bompastor-Ramos, Paula ;
Povoa, Joao ;
Lobo, Conceicao ;
Rodriguez, Alejandra E. ;
Alio, Jorge L. ;
Werner, Liliana ;
Murta, Joaquim N. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (09) :1324-1331
[5]   Calcification in Hydrophilic Intraocular Lenses Associated With Injection of Intraocular Gas [J].
Dhital, Anish ;
Spalton, David J. ;
Goyal, Saurabh ;
Werner, Liliana .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2012, 153 (06) :1154-1160
[6]   Proposed pathogenesis for the delayed postoperative opacification of the hydroview hydrogel intraocular lens [J].
Dorey, MW ;
Brownstein, S ;
Hill, VE ;
Mathew, B ;
Botton, G ;
Kertes, PJ ;
El-Defrawy, S .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2003, 135 (05) :591-598
[7]   On the Chromatic Dispersion of Hydrophobic and Hydrophilic Intraocular Lenses [J].
Eppig, Timo ;
Rawer, Antonia ;
Hoffmann, Peter ;
Langenbucher, Achim ;
Schroeder, Simon .
OPTOMETRY AND VISION SCIENCE, 2020, 97 (04) :305-313
[8]   Hydrophilic intraocular lens opacification after posterior lamellar keratoplasty - a material analysis with special reference to optical quality assessment [J].
Giers, Bert C. ;
Tandogan, Tamer ;
Auffarth, Gerd U. ;
Choi, Chul Y. ;
Auerbach, Florian N. ;
Sel, Saadettin ;
Mayer, Christian ;
Khoramnia, Ramin .
BMC OPHTHALMOLOGY, 2017, 17
[9]   Serial intraocular lens opacifications of different designs from the same manufacturer: Clinical and light microscopic results of 71 explant cases [J].
Gurabardhi, Maceda ;
Haeberle, Heike ;
Aurich, Henning ;
Werner, Liliana ;
Duy-Thoai Pham .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2018, 44 (11) :1326-1332
[10]   The relation between visual acuity and illumination. [J].
Hecht, S .