An efficient treatment planning approach to reduce the critical organ dose in volumetric modulated arc therapy for synchronous bilateral breast cancer patients

被引:3
|
作者
Lin, Y. F. [1 ]
Shueng, P. W. [1 ,2 ]
Lin, H. H. [3 ,4 ,5 ]
Tien, H. J. [1 ]
Lai, L. H. [6 ]
机构
[1] Far Eastern Mem Hosp, Dept Radiat Oncol, New Taipei, Taiwan
[2] Natl Yang Ming Univ, Sch Med, Dept Med, Taipei, Taiwan
[3] Chang Gung Univ, Inst Radiol Res, Med Phys Res Ctr, Chang Gung Mem Hosp, Taoyuan, Taiwan
[4] Chang Gung Mem Hosp, Dept Radiat Oncol, Taoyuan, Taiwan
[5] Chang Gung Univ, Inst Radiol Res, Particle Phys & Beam Delivery Core Lab, Chang Gung Mem Hosp, Taoyuan, Taiwan
[6] China Med Univ, Dept Biomed Imaging & Radiol Sci, 91 Hsueh Shih Rd, Taichung 40402, Taiwan
关键词
Synchronous bilateral breast cancer; Volumetric modulated arc therapy; Partial arc; Pseudo planning structure; RADIOTHERAPY; RADIATION; RISK; OPTIMIZATION; IRRADIATION; DISEASE;
D O I
10.1016/j.radphyschem.2020.108957
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Treatment planning for patients with synchronous bilateral breast cancer (SBBC) is challenging due to the technical complexity of bilateral treatment. To minimize both high-dose and low-dose volumes of the heart and lungs being irradiated, we proposed an efficient and simple strategy for volumetric modulated arc therapy (VMAT) treatment planning in this study. Five SBBC patients were enrolled in this study. We utilized two pseudo block planning structures to minimize the high-dose volume of the whole lungs and heart and confine the low-dose areas in normal organs for VMAT planning. The prescription delivered was 45 Gy in 25 fractions to the whole breast. The VMAT treatment plan consisted of two partial arcs. In terms of the whole lungs, the average V-5Gy, V-10Gy and V-20Gy in the VMAT-B plan (with two pseudo planning block structures) for the five patients were 34.8 +/- 7.2%, 23.4 +/- 5.0% and 14.8 +/- 3.0%, respectively. For the VMAT-NB plan (without two pseudo planning block structures), the average V-5Gy, V-10Gy and V-20Gy were 38.6 +/- 6.8%, 27 +/- 5.9% and 18.4 +/- 4.0%, respectively. For the heart, the average V-5Gy, V-10Gy, V-20Gy and V-30Gy in the VMAT-B plan were 19.6 +/- 3.1%, 7.0 +/- 2.5%, 2.2 +/- 1.2% and 0.7 +/- 0.7%, respectively, while the mean values in the VMAT-NB plan were 26.2 +/- 6.3%, 11.3 +/- 3.9%, 3.9 +/- 1.8% and 1.5 +/- 1.1%, respectively. We designed an approach to reduce both the high-dose and low-dose volumes by utilizing pseudo planning structures in the VMAT treatment plan. The dosimetric results of this study showed that the pseudo block structures have the potential to reduce the risks of radiation pneumonitis and heart disease, while preserving the same tumor control rate.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Treatment planning evaluation for bilateral breast cancer using volumetric modulated arc therapy (VMAT)
    Abdullah, Siham Sabah
    Alabedi, Haydar H.
    Taha, Jenan Hussein
    Abdullah, Khalid
    REVISTA DE SENOLOGIA Y PATOLOGIA MAMARIA, 2023, 36 (01):
  • [2] Efficacy, safety, and feasibility of volumetric modulated arc therapy for synchronous bilateral breast cancer management
    Quesada, Stanislas
    Fenoglietto, Pascal
    Gourgou, Sophie
    Lemanski, Claire
    Draghici, Roxana
    Ailleres, Norbert
    Prunaretty, Jessica
    Azria, David
    Bourgier, Celine
    FRONTIERS IN ONCOLOGY, 2022, 12
  • [3] Treatment planning comparison of volumetric modulated arc therapy with the trilogy and the Halcyon for bilateral breast cancer
    Sun, Tao
    Lin, Xiutong
    Zhang, Guifang
    Qiu, Qingtao
    Li, Chengqiang
    Yin, Yong
    RADIATION ONCOLOGY, 2021, 16 (01)
  • [4] Treatment planning comparison of volumetric modulated arc therapy with the trilogy and the Halcyon for bilateral breast cancer
    Tao Sun
    Xiutong Lin
    Guifang Zhang
    Qingtao Qiu
    Chengqiang Li
    Yong Yin
    Radiation Oncology, 16
  • [5] Critical appraisal of the role of volumetric modulated arc therapy in the radiation therapy management of breast cancer
    Cozzi, Luca
    Lohr, Frank
    Fogliata, Antonella
    Franceschini, Davide
    De Rose, Fiorenza
    Filippi, A. R.
    Guidi, Gabriele
    Vanoni, Valentina
    Scorsetti, Marta
    RADIATION ONCOLOGY, 2017, 12
  • [6] Patient specific quality assurance of volumetric modulated arc therapy of synchronous bilateral breast cancer
    Zamo, Francis C. Djoumessi
    Njeh, Christopher F.
    Colliaux, Anthony
    Blot-Lafond, Valerie
    Moyo, M. Ndontchueng
    MEDICAL DOSIMETRY, 2024, 49 (03) : 177 - 184
  • [7] Volumetric Modulated Arc Therapy Dose Distribution Prediction for Breast Cancer Patients: CNN Approach
    Krawczyk, Zuzanna
    Szmurlo, Robert
    Zawadzki, Pawel
    Kot, Estera
    Starzynski, Jacek
    Zawadzka, Anna
    2021 INTERNATIONAL JOINT CONFERENCE ON NEURAL NETWORKS (IJCNN), 2021,
  • [8] Physical and Radiobiological Dosimetric Comparison of Volumetric Arc Treatment Plans with or without Flattening Filter in Synchronous Bilateral Breast Cancer
    Patil, Deepali
    Zope, Mukesh
    Madhawi, Richa
    Raj, Shraddha
    Kishor, Kunal
    Devi, Seema
    Singh, Rajesh
    Singh, Devraj
    TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2024, 39 (04): : 411 - 422
  • [9] A fast optimization approach for treatment planning of volumetric modulated arc therapy
    Yan, Hui
    Dai, Jian-Rong
    Li, Ye-Xiong
    RADIATION ONCOLOGY, 2018, 13
  • [10] Hybrid volumetric modulated arc therapy for hypofractionated radiotherapy of breast cancer: a treatment planning study
    Venjakob, Alexander
    Oertel, Michael
    Hering, Dominik Alexander
    Moustakis, Christos
    Haverkamp, Uwe
    Eich, Hans Theodor
    STRAHLENTHERAPIE UND ONKOLOGIE, 2021, 197 (04) : 296 - 307