A Monte Carlo study of dose rate distribution around the specially asymmetric CSM3-a 137Cs source

被引:8
作者
Pérez-Calatayud, J
Lliso, F
Ballester, F
Serrano, MA
Lluch, JL
Limami, Y
Puchades, V
Casal, E
机构
[1] La Fe Univ Hosp, Dept Radiat Oncol, Phys Sect, E-46009 Valencia, Spain
[2] Univ Valencia, Dept Atom Mol & Nucl Phys, E-46100 Burjassot, Spain
[3] Univ Valencia, CSIC, IFIC, Inst Fis Corpuscular, E-46100 Burjassot, Spain
[4] CND, E-46009 Valencia, Spain
关键词
D O I
10.1088/0031-9155/46/7/401
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The CSM3 Cs-137 type stainless-steel encapsulated source is widely used in manually afterloaded low close rate brachytherapy. A specially asymmetric source. CSM3-a, has been designed by CIS Bio International (France) substituting the eyelet side seed with an inactive material in the CSM3 source. This modification has been done in order to allow a uniform dose level over the upper vaginal surface when this 'linear' source is inserted at the top of the dome vaginal applicators. In this study the Monte Carlo GEANT3 simulation code. incorporating the source geometry in detail, was used to investigate the dosimetric characteristics of this special CSM3-a Cs-137 brachytherapy source. The absolute dose rate distribution in water around this source was calculated and is presented in the form of an along-away table. Comparison of Sievert integral type calculations with Monte Carlo results are discussed.
引用
收藏
页码:N169 / N174
页数:6
相关论文
共 13 条
[1]  
AAPM, 1987, 21 AAPM
[2]   A Monte Carlo investigation of the dosimetric characteristics of the CSM11 137Cs source from CIS [J].
Ballester, F ;
Lluch, JL ;
Limami, Y ;
Serrano, MA ;
Casal, E ;
Pérez-Calatayud, J ;
Lliso, F .
MEDICAL PHYSICS, 2000, 27 (09) :2182-2189
[3]   Monte Carlo calculation of dose rate distributions around Ir-192 wires [J].
Ballester, F ;
Hernandez, C ;
PerezCalatayud, J ;
Lliso, F .
MEDICAL PHYSICS, 1997, 24 (08) :1221-1228
[4]  
BRUN R, 1987, DDEE841 CERN
[5]   Monte Carlo calculations of dose rate distributions around the Amersham CDCS-M-type 137Cs source [J].
Casal, E ;
Ballester, F ;
Lluch, JL ;
Pérez-Calatayud, J ;
Lliso, F .
MEDICAL PHYSICS, 2000, 27 (01) :132-140
[7]   MINICOLPOSTATS, DOME CYLINDERS, OTHER ADDITIONS AND IMPROVEMENTS OF THE FLETCHER-SUIT AFTERLOADABLE SYSTEM - INDICATIONS AND LIMITATIONS OF THEIR USE [J].
DELCLOS, L ;
FLETCHER, GH ;
MOORE, EB ;
SAMPIERE, VA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1980, 6 (09) :1195-1206
[8]   Code of practice for brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 56 [J].
Nath, R ;
Anderson, LL ;
Meli, JA ;
Olch, AJ ;
Stitt, JA ;
Williamson, JF .
MEDICAL PHYSICS, 1997, 24 (10) :1557-1598
[9]   DOSIMETRY OF INTERSTITIAL BRACHYTHERAPY SOURCES - RECOMMENDATIONS OF THE AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO 43 [J].
NATH, R ;
ANDERSON, LL ;
LUXTON, G ;
WEAVER, KA ;
WILLIAMSON, JF ;
MEIGOONI, AS .
MEDICAL PHYSICS, 1995, 22 (02) :209-234
[10]   Monte Carlo calculation of dose rate distributions around 0.5 and 0.6 mm in diameter 192Ir wires [J].
Pérez-Calatayud, J ;
Lliso, F ;
Carmona, V ;
Ballester, F ;
Hernández, C .
MEDICAL PHYSICS, 1999, 26 (03) :395-401