End-to-end tests using alanine dosimetry in scanned proton beams

被引:20
作者
Carlino, A. [1 ,3 ]
Gouldstone, C. [2 ]
Kragl, G. [1 ]
Traneus, E. [4 ]
Marrale, M. [3 ]
Vatnitsky, S. [1 ]
Stock, M. [1 ]
Palmans, H. [1 ,2 ]
机构
[1] EBG MedAustron GmbH, Marie Curie Str 5, A-2700 Wiener Neustadt, Austria
[2] Natl Phys Lab, Hampton Rd, Teddington TW11 0LW, Middx, England
[3] Univ Palermo, Dept Phys & Chem, Viale Sci,Edificio 18, I-90128 Palermo, Italy
[4] RaySearch Labs AB, Sveavagen 44,POB 3297, Stockholm, Sweden
关键词
protons; alanine; audit; dosimetry; anthropomorphic phantoms; end to end test; HEAVY CHARGED-PARTICLES; CLINICAL-TRIALS; AUDIT PROGRAM; THERAPY BEAMS; ION-BEAMS; DETECTORS; PHANTOM; SYSTEM; FILMS; INSTITUTIONS;
D O I
10.1088/1361-6560/aaac23
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes end-to-end test procedures as the last fundamental step of medical commissioning before starting clinical operation of the MedAustron synchrotron-based pencil beam scanning (PBS) therapy facility with protons. One in-house homogeneous phantom and two anthropomorphic heterogeneous (head and pelvis) phantoms were used for end-to-end tests at MedAustron. The phantoms were equipped with alanine detectors, radiochromic films and ionization chambers. The correction for the 'quenching' effect of alanine pellets was implemented in the Monte Carlo platform of the evaluation version of RayStation TPS. During the end-to-end tests, the phantoms were transferred through the workflow like real patients to simulate the entire clinical workflow: immobilization, imaging, treatment planning and dose delivery. Different clinical scenarios of increasing complexity were simulated: delivery of a single beam, two oblique beams without and with range shifter. In addition to the dose comparison in the plastic phantoms the dose obtained from alanine pellet readings was compared with the dose determined with the Farmer ionization chamber in water. A consistent systematic deviation of about 2% was found between alanine dosimetry and the ionization chamber dosimetry in water and plastic materials. Acceptable agreement of planned and delivered doses was observed together with consistent and reproducible results of the end-to-end testing performed with different dosimetric techniques (alanine detectors, ionization chambers and EBT3 radiochromic films). The results confirmed the adequate implementation and integration of the new PBS technology at MedAustron. This work demonstrates that alanine pellets are suitable detectors for end-to-end tests in proton beam therapy and the developed procedures with customized anthropomorphic phantoms can be used to support implementation of PBS technology in clinical practice.
引用
收藏
页数:17
相关论文
共 46 条
[1]   Dosimetry auditing procedure with alanine dosimeters for light ion beam therapy [J].
Ableitinger, Alexander ;
Vatnitsky, Stanislav ;
Herrmann, Rochus ;
Bassler, Niels ;
Palmans, Hugo ;
Sharpe, Peter ;
Ecker, Swantje ;
Chaudhri, Naved ;
Jaekel, Oliver ;
Georg, Dietmar .
RADIOTHERAPY AND ONCOLOGY, 2013, 108 (01) :99-106
[2]   Optically Stimulated Light Dosimetry: Commissioning of An Optically Stimulated Luminescence (OSL) System for Remote Dosimetry Audits, the Radiological Physics Center Experience [J].
Aguirre, J. ;
Alvarez, P. ;
Followill, D. ;
Ibbott, G. ;
Amador, C. ;
Tailor, A. .
MEDICAL PHYSICS, 2009, 36 (06)
[3]   Experimental verification of IMPT treatment plans in an anthropomorphic phantom in the presence of delivery uncertainties [J].
Albertini, F. ;
Casiraghi, M. ;
Lorentini, S. ;
Rombi, B. ;
Lomax, A. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (14) :4415-4431
[4]  
[Anonymous], 2008, GUID EXPR UNC MEAS
[5]  
[Anonymous], 1993, 49 ICRU
[6]   Dosimetry intercomparison of four proton therapy institutions in Germany employing spot scanning [J].
Baeumer, Christian ;
Ackermann, Benjamin ;
Hillbrand, Martin ;
Kaiser, Franz -Joachim ;
Koska, Benjamin ;
Latzel, Harald ;
Luehr, Armin ;
Menkel, Stefan ;
Timmermann, Beate .
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2017, 27 (02) :80-85
[7]   The antiproton depth-dose curve measured with alanine detectors [J].
Bassler, Niels ;
Hansen, Johnny W. ;
Palmans, Hugo ;
Holzscheiter, Michael H. ;
Kovacevic, Sandra .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2008, 266 (06) :929-936
[8]   A national dosimetric audit of IMRT [J].
Budgell, Geoff ;
Berresford, Joe ;
Trainer, Michael ;
Bradshaw, Ellie ;
Sharpe, Peter ;
Williams, Peter .
RADIOTHERAPY AND ONCOLOGY, 2011, 99 (02) :246-252
[9]   A multi-institutional dosimetry audit of rotational intensity-modulated radiotherapy [J].
Clark, Catharine H. ;
Hussein, Mohammad ;
Tsang, Yatman ;
Thomas, Russell ;
Wilkinson, Dean ;
Bass, Graham ;
Snaith, Julia ;
Gouldstone, Clare ;
Bolton, Steve ;
Nutbrown, Rebecca ;
Venables, Karen ;
Nisbet, Andrew .
RADIOTHERAPY AND ONCOLOGY, 2014, 113 (02) :272-278
[10]   Dosimetry audit for a multi-centre IMRT head and neck trial [J].
Clark, Catharine H. ;
Hansen, Vibeke Nordmark ;
Chantler, Hannah ;
Edwards, Craig ;
James, Hayley V. ;
Webster, Gareth ;
Miles, Elizabeth A. ;
Urbano, M. Teresa Guerrero ;
Bhide, Shree A. ;
Bidmead, A. Margaret ;
Nutting, Christoper M. .
RADIOTHERAPY AND ONCOLOGY, 2009, 93 (01) :102-108