Assessment of the validation of blood pressure monitors: a statistical reappraisal

被引:34
作者
Friedman, Bruce A. [1 ]
Alpert, Bruce S. [2 ]
Osborn, David [3 ]
Prisant, L. Michael [4 ]
Quinn, David E. [5 ]
Seller, John [5 ]
机构
[1] GE Healthcare, Tampa, FL USA
[2] Univ Tennessee, Ctr Hlth Sci, Memphis, TN 38163 USA
[3] Philips Healthcare, Andover, MA USA
[4] Med Coll Georgia, Augusta, GA 30912 USA
[5] WelchAllyn, Skaneateles Falls, NY USA
关键词
Association for the Advancement of Medical Instrumentation; blood pressure measurement; British Hypertension Society; European Society of Hypertension; international protocol;
D O I
10.1097/MBP.0b013e3283071a64
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Protocols to determine the accuracy of noninvasive blood pressure monitors have been published by the Association for the Advancement of Medical Instrumentation, the British Hypertension Society, and the European Society of Hypertension (ESH). We compared the statistical basis for each of these protocols. Although the Association for the Advancement of Medical Instrumentation and British Hypertension Society protocols require a minimum of 85 participants, the ESH protocol requires only 33 participants. The reduced sample size results in a reduction in statistical power from 98 to 70%, which brings into question the applicability of the ESH (International) protocol.
引用
收藏
页码:187 / 191
页数:5
相关论文
共 12 条
[1]  
[Anonymous], CRC HDB TABLES PROBA
[2]  
*ASS ADV MED INSTR, 2002, ANSI AAMI
[3]  
*ASS ADV MED INSTR, 1987, ANSI AAMI
[4]  
*ASS ADV MED INSTR, 1993, ANSI AAMI
[5]   The 'International Protocol': more insight or more arithmetic? [J].
Braam, RL ;
Thien, T .
BLOOD PRESSURE MONITORING, 2002, 7 (05) :289-290
[6]   Limitations of current validation protocols for home blood pressure monitors for individual patients [J].
Gerin, W ;
Schwartz, AR ;
Schwartz, JE ;
Pickering, TG ;
Davidson, KW ;
Bress, J ;
O'Brien, E ;
Atkins, N .
BLOOD PRESSURE MONITORING, 2002, 7 (06) :313-318
[7]   Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults [J].
O'Brien, E ;
Pickering, T ;
Asmar, R ;
Myers, M ;
Parati, G ;
Staessen, J ;
Mengden, T ;
Imai, Y ;
Waeber, B ;
Palatini, P ;
Atkins, N ;
Gerin, W .
BLOOD PRESSURE MONITORING, 2002, 7 (01) :3-17
[8]   THE BRITISH-HYPERTENSION-SOCIETY PROTOCOL FOR THE EVALUATION OF AUTOMATED AND SEMIAUTOMATED BLOOD-PRESSURE MEASURING DEVICES WITH SPECIAL REFERENCE TO AMBULATORY SYSTEMS [J].
OBRIEN, E ;
PETRIE, J ;
LITTLER, W ;
DESWIET, M ;
PADFIELD, PL ;
OMALLEY, K ;
JAMIESON, M ;
ALTMAN, D ;
BLAND, M ;
ATKINS, N .
JOURNAL OF HYPERTENSION, 1990, 8 (07) :607-619
[9]  
OBRIEN E, 1994, J HYPERTENS, V12, P1089
[10]   THE BRITISH-HYPERTENSION-SOCIETY PROTOCOL FOR THE EVALUATION OF AUTOMATED AND SEMIAUTOMATED BLOOD-PRESSURE MEASURING DEVICES WITH SPECIAL REFERENCE TO AMBULATORY SYSTEMS [J].
OBRIEN, E ;
PETRIE, J ;
LITTLER, W ;
DESWIET, M ;
PADFIELD, PL ;
OMALLEY, K ;
JAMIESON, M ;
ALTMAN, D ;
BLAND, M ;
ATKINS, N .
JOURNAL OF HYPERTENSION, 1990, 8 (07) :607-619