Contactless radar-based breathing monitoring of premature infants in the neonatal intensive care unit

被引:40
作者
Beltrao, Gabriel [1 ]
Stutz, Regine [2 ]
Hornberger, Franziska [2 ]
Martins, Wallace A. [1 ]
Tatarinov, Dimitri [3 ]
Alaee-Kerahroodi, Mohammad [1 ]
Lindner, Ulrike [2 ]
Stock, Lilly [2 ]
Kaiser, Elisabeth [2 ]
Goedicke-Fritz, Sybelle [2 ]
Schroeder, Udo [3 ]
Shankar, Bhavani M. R. [1 ]
Zemlin, Michael [2 ]
机构
[1] Univ Luxembourg, SnT Interdisciplinary Ctr Secur Reliabil & Trust, Luxembourg, Luxembourg
[2] Saarland Univ, Dept Gen Pediat & Neonatol, Med Sch, Homburg, Germany
[3] IEE SA, Bissen, Luxembourg
关键词
HEART-RATE; HEALTH-CARE; PRETERM; TIME; PATTERN; MODEL; PAIN;
D O I
10.1038/s41598-022-08836-3
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Vital sign monitoring systems are essential in the care of hospitalized neonates. Due to the immaturity of their organs and immune system, premature infants require continuous monitoring of their vital parameters and sensors need to be directly attached to their fragile skin. Besides mobility restrictions and stress, these sensors often cause skin irritation and may lead to pressure necrosis. In this work, we show that a contactless radar-based approach is viable for breathing monitoring in the Neonatal intensive care unit (NICU). For the first time, different scenarios common to the NICU daily routine are investigated, and the challenges of monitoring in a real clinical setup are addressed through different contributions in the signal processing framework. Rather than just discarding measurements under strong interference, we present a novel random body movement mitigation technique based on the time-frequency decomposition of the recovered signal. In addition, we propose a simple and accurate frequency estimator which explores the harmonic structure of the breathing signal. As a result, the proposed radar-based solution is able to provide reliable breathing frequency estimation, which is close to the reference cabled device values most of the time. Our findings shed light on the strengths and limitations of this technology and lay the foundation for future studies toward a completely contactless solution for vital signs monitoring.
引用
收藏
页数:15
相关论文
共 79 条
[1]   Neonatal non-contact respiratory monitoring based on real-time infrared thermography [J].
Abbas, Abbas K. ;
Heimann, Konrad ;
Jergus, Katrin ;
Orlikowsky, Thorsten ;
Leonhardt, Steffen .
BIOMEDICAL ENGINEERING ONLINE, 2011, 10
[2]   Smart Homes that Monitor Breathing and Heart Rate [J].
Adib, Fadel ;
Mao, Hongzi ;
Kabelac, Zachary ;
Katabi, Dina ;
Miller, Robert C. .
CHI 2015: PROCEEDINGS OF THE 33RD ANNUAL CHI CONFERENCE ON HUMAN FACTORS IN COMPUTING SYSTEMS, 2015, :837-846
[3]   Monitoring of Cardiorespiratory Signal: Principles of Remote Measurements and Review of Methods [J].
Al-Naji, Ali ;
Gibson, Kim ;
Lee, Sang-Heon ;
Chahl, Javaan .
IEEE ACCESS, 2017, 5 :15776-15790
[4]   An integrated mathematical model of the human cardiopulmonary system: model development [J].
Albanese, Antonio ;
Cheng, Limei ;
Ursino, Mauro ;
Chbat, Nicolas W. .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2016, 310 (07) :H899-H921
[5]   Remote Monitoring of Human Vital Signs Using mm-Wave FMCW Radar [J].
Alizadeh, Mostafa ;
Shaker, George ;
Martins De Almeida, Joao Carlos ;
Morita, Plinio Pelegrini ;
Safavi-Naeini, Safeddin .
IEEE ACCESS, 2019, 7 :54958-54968
[6]   Photoplethysmography and its application in clinical physiological measurement [J].
Allen, John .
PHYSIOLOGICAL MEASUREMENT, 2007, 28 (03) :R1-R39
[7]   Clinical importance of pain and stress in preterm neonates [J].
Anand, KJS .
BIOLOGY OF THE NEONATE, 1998, 73 (01) :1-9
[8]  
[Anonymous], 2005, Spectral Analysis of Signals
[9]   Cobedding of Twin Premature Infants: Calming Effects on Pain Responses [J].
Badiee, Zohreh ;
Nassiri, Zohreh ;
Armanian, Amirmohammad .
PEDIATRICS AND NEONATOLOGY, 2014, 55 (04) :262-268
[10]   A Real-time Heart Rate Analysis for a Remote Millimeter Wave I-Q Sensor [J].
Bakhtiari, Sasan ;
Liao, Shaolin ;
Elmer, Thomas, II ;
Gopalsami, Nachappa 'Sami' ;
Raptis, A. C. .