Comparison of the influence of cataract and pupil size on retinal nerve fibre layer thickness measurements with time-domain and spectral-domain optical coherence tomography

被引:33
作者
Cheng, Clarissa S. M. [1 ]
Natividad, Maricel G. [1 ]
Earnest, Arul [2 ]
Yong, Vernon [1 ]
Lim, Boon A. [1 ]
Wong, Hon T. [1 ]
Yip, Leonard W. [1 ]
机构
[1] Tan Tock Seng Hosp, Dept Ophthalmol, Singapore 308433, Singapore
[2] Tan Tock Seng Hosp, Clin Res Unit, Singapore 308433, Singapore
关键词
cataract; glaucoma; optical coherence tomography; pupil; retinal nerve fibre layer; CLASSIFICATION-SYSTEM-III; SIGNAL STRENGTH; STRATUS OCT; GLAUCOMA; VARIABILITY; PERFORMANCE; PROGRESSION; EXTRACTION;
D O I
10.1111/j.1442-9071.2010.02460.x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
P>Background: To investigate and compare the effect of cataract and pupil size on retinal nerve fibre layer (RNFL) thickness measurements using spectral-domain optical coherence tomography (Cirrus OCT) and time-domain OCT (Stratus OCT). Design: Prospective, hospital-based study. Participants: Twenty-five eyes from 25 normal subjects undergoing cataract surgery. Methods: Three retinal nerve fibre layer (RNFL) thickness measurements were taken before and after dilation, preoperatively and postoperatively, using Cirrus 200 x 200 Optic Disc Scan and Stratus Fast RNFL Scan. Main Outcome Measures: Linear regression, intraclass correlation coefficient (ICC) and coefficient of variation analysis. Results: Cataract removal caused significant increase in RNFL measurements in both modalities (Cirrus P < 0.02; Stratus P < 0.001). There was no significant difference in the increase in measurements between the two machines. Pupil dilation had variable and non-statistically significant effect in both (P > 0.05). ICC showed excellent reproducibility with Cirrus OCT after mydriasis, preoperatively (ICC = 0.78-0.90) and postoperatively (ICC = 0.90-0.97), but poor reproducibility before mydriasis (P < 0.75). Stratus OCT achieved excellent reproducibility after cataract removal both before (ICC = 0.86-0.96) and after mydriasis (ICC = 0.92-0.95), but poor reproducibility before cataract surgery (P < 0.75). Conclusions: Cataracts, not pupil size, cause significant underestimation of RNFL measurements in both Cirrus and Stratus OCT. The extent of influence exerted does not appear different between the two instruments. Reproducibility of each machine appears to be affected differently. Mydriasis is required to achieve excellent reproducibility with Cirrus OCT, and media clarity is required with Stratus OCT.
引用
收藏
页码:215 / 221
页数:7
相关论文
共 24 条
[1]   Sensitivity and Specificity of Time-Domain versus Spectral-Domain Optical Coherence Tomography in Diagnosing Early to Moderate Glaucoma [J].
Chang, Robert T. ;
Knight, O'Rese J. ;
Feuer, William J. ;
Budenz, Donald L. .
OPHTHALMOLOGY, 2009, 116 (12) :2294-2299
[2]   Relationship between retinal nerve fiber layer measurement and signal strength in optical coherence tomography [J].
Cheung, Carol Yim Lui ;
Leung, Christopher Kai Shun ;
Lin, Dusheung ;
Pang, Chi-Pui ;
Lam, Dennis Shun Chiu .
OPHTHALMOLOGY, 2008, 115 (08) :1347-1351
[3]   THE LENS OPACITIES CLASSIFICATION SYSTEM-III [J].
CHYLACK, LT ;
WOLFE, JK ;
SINGER, DM ;
LESKE, MC ;
BULLIMORE, MA ;
BAILEY, IL ;
FRIEND, J ;
MCCARTHY, D ;
WU, SY .
ARCHIVES OF OPHTHALMOLOGY, 1993, 111 (06) :831-836
[4]   Glaucoma [J].
Coleman, AL .
LANCET, 1999, 354 (9192) :1803-1810
[5]   Clinical application of the lens opacities classification system III in the performance of phacoemulsification [J].
Davison, JA ;
Chylack, LT .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (01) :138-145
[6]   Effect of cataract extraction and intraocular lens implantation on nerve fibre layer thickness measurements by scanning laser polarimeter (GDx) in glaucoma patients [J].
Gazzard, G ;
Foster, PJ ;
Devereux, JG ;
Oen, F ;
Chew, PTK ;
Khaw, PT ;
Seah, SKL .
EYE, 2004, 18 (02) :163-168
[7]   Reduction of intraocular pressure and glaucoma progression - Results from the early manifest glaucoma trial [J].
Heijl, A ;
Leske, MC ;
Bengtsson, B ;
Hyman, L ;
Bengtsson, B ;
Hussein, M .
ARCHIVES OF OPHTHALMOLOGY, 2002, 120 (10) :1268-1279
[8]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[9]   Comparison of Cirrus OCT and Stratus OCT on the Ability to Detect Localized Retinal Nerve Fiber Layer Defects in Preperimetric Glaucoma [J].
Jeoung, Jin Wook ;
Park, Ki Ho .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (02) :938-945
[10]   Spectral-Domain Optical Coherence Tomography: A Comparison of Modern High-Resolution Retinal Imaging Systems [J].
Kiernan, Daniel F. ;
Mieler, William F. ;
Hariprasad, Seenu M. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2010, 149 (01) :18-31