Ceramic femoral heads prevent runaway wear for highly crosslinked polyethylene acetabular cups by third-body bone cement particles

被引:14
作者
Wang, A [1 ]
Schmidig, G [1 ]
机构
[1] Stryker Howmedica Osteon, Mahwah, NJ USA
关键词
crosslinked UHMWPE; third-body wear; runaway wear; ceramic head; bone cement; hip simulator;
D O I
10.1016/S0043-1648(03)00289-8
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Third-body wear due to entrapment of loose bone cement particles has been identified as one of the major factors that are responsible for elevated wear debris generation in total hip replacement. Previous studies have shown that the wear rate of non-crosslinked UHMWPE acetabular cups against metallic femoral heads accelerates significantly with increasing the concentration of cement particles in lubricant whereas the use of ceramic heads prevents such a runaway wear. The present study looks at the effect of third-body bone cement particles on the wear rate of highly crosslinked UHMWPE acetabular cups against both metal and ceramic femoral heads in a hip joint simulator. It is hypothesized that ceramic heads would prevent runaway wear for highly crosslinked polyethylene cups against a heavy concentration of polymethylmethacrylate (PMMA) bone cement particles in lubricant. This hypothesis was indeed verified by the present hip simulator study. (C) 2003 Published by Elsevier B.V.
引用
收藏
页码:1057 / 1063
页数:7
相关论文
共 20 条
[1]  
CAPELLO WN, 2000, CLIN ORTHOP RELAT R, V370, P183
[2]   Characterization of long-term femoral-head-penetration rates - Association with and prediction of osteolysis [J].
Dowd, JE ;
Sychterz, CJ ;
Young, AM ;
Engh, CA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (08) :1102-1107
[3]   A comparative study of the performance of metallic and ceramic femoral head components in total replacement hip joints [J].
Dowson, D .
WEAR, 1995, 190 (02) :171-183
[4]   A literature review of the association between wear rate and osteolysis in total hip arthroplasty [J].
Dumbleton, JH ;
Manley, MT ;
Edidin, AA .
JOURNAL OF ARTHROPLASTY, 2002, 17 (05) :649-661
[5]  
ESSNER A, 1998, T SOC BIOMATER, V21, P4
[6]  
Hall R M, 1996, Proc Inst Mech Eng H, V210, P197, DOI 10.1243/PIME_PROC_1996_210_413_02
[7]   Wear of polyethylene acetabular components in total hip arthroplasty - An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations [J].
Jasty, M ;
Goetz, DD ;
Bragdon, CR ;
Lee, KR ;
Hanson, AE ;
Elder, JR ;
Harris, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (03) :349-358
[8]  
LANGER G, 1991, ULTRA HIGH MOL WEIGH, P154
[9]  
MARS H, 1998, T ORTHOPAED RES SOC, V23, P100
[10]   Wear of gamma-crosslinked polyethylene acetabular cups against roughened femoral balls [J].
McKellop, H ;
Shen, FW ;
DiMaio, W ;
Lancaster, JG .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (369) :73-82