Gold-tip electrodes -: A new "deep lesion" technology for catheter ablation?: In vitro comparison of a gold alloy versus platinum-iridium tip electrode ablation catheter

被引:26
作者
Lewalter, T
Bitzen, A
Wurtz, S
Blum, R
Schlodder, K
Yang, A
Lickfett, L
Schwab, JO
Schrickel, JW
Tiemann, K
Linhart, M
Zima, E
Merkely, B
Lüderitz, B
机构
[1] Univ Bonn, Dept Med Cardiol, D-53105 Bonn, Germany
[2] BIOTRONIK Berlin, Berlin, Germany
[3] Semmelweis Univ Clin, Dept Cardiovasc Surg, Expt Cardiovasc Res Lab, Budapest, Hungary
关键词
radiofrequency catheter ablation; electrodes; tachycardia; arrhythmias;
D O I
10.1111/j.1540-8167.2005.40832.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Radiofrequency (RF) catheter ablation is widely used to induce focal myocardial necrosis using the effect of resistive heating through high-frequency current delivery. It is current standard to limit the target tissue-electrode interface temperature to a maximum of 60-70 degrees C to avoid char formation. Gold (Au) exhibits a thermal conductivity of nearly four times greater than platinum (Pt-Ir) (3.17 W/cm Kelvin vs 0.716 W/cm Kelvin), it was therefore hypothesized that RF ablation using a gold electrode would create broader and deeper lesions as a result of a better heat conduction from the tissue-electrode interface and additional cooling of the gold electrode by "heat loss" to the intracardiac blood. Both mechanisms would allow applying more RF power to the tissue before the electrode-tissue interface temperature limit is reached. To test this hypothesis, we performed in vitro isolated liver and pig heart investigations comparing lesion depths of a new Au-alloy-tip electrode to standard Pt-Ir electrode material. Mean lesion depth in liver tissue for Pt-Ir was 4.33 +/- 0.45 mm (n = 60) whereas Au electrode was able to achieve significantly deeper lesions (5.86 +/- 0.37 mm [n = 60; P < 0.001]). The mean power delivered using Pt-Ir was 6.95 +/- 2.41 W whereas Au tip electrode delivered 9.64 +/- 3.78 W indicating a statistically significant difference (P < 0.05). In vitro pig heart tissue Au ablation (n = 20) increased significantly the lesion depth (Au: 4.85 +/- 1.01 mm, Pt-Ir: 2.96 +/- 0.81 mm, n = 20; P < 0.001). Au tip electrode again applied significantly more power (P < 0.001). Gold-tip electrode catheters were able to induce deeper lesions using RF ablation in vitro as compared to Pt-Ir tip electrode material. In liver and in pig heart tissue, the increase in lesion depth was associated with a significant increase in the average power applied with the gold electrode at the same level of electrode-tissue temperature as compared to platinum material.
引用
收藏
页码:770 / 772
页数:3
相关论文
共 7 条
[1]  
[Anonymous], 1986, J APPL CARDIOL
[2]   THE EFFECT OF ELECTRODE DESIGN ON THE EFFICIENCY OF DELIVERY OF RADIOFREQUENCY ENERGY TO CARDIAC TISSUE INVITRO [J].
BLOUIN, LT ;
MARCUS, FI .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (01) :136-143
[3]   TISSUE HEATING DURING RADIOFREQUENCY CATHETER ABLATION - A THERMODYNAMIC MODEL AND OBSERVATIONS IN ISOLATED PERFUSED AND SUPERFUSED CANINE RIGHT VENTRICULAR FREE WALL [J].
HAINES, DE ;
WATSON, DD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (06) :962-976
[4]   Drug therapy - Radio-frequency ablation as treatment for cardiac arrhythmias [J].
Morady, F .
NEW ENGLAND JOURNAL OF MEDICINE, 1999, 340 (07) :534-544
[5]   Comparison of gold versus platinum electrodes on myocardial lesion size using radiofrequency energy [J].
Simmons, WN ;
Mackey, S ;
He, DS ;
Marcus, FI .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (04) :398-402
[6]   Is 8-mm more effective than 4-mm tip electrode catheter for ablation of typical atrial flutter? [J].
Tsai, CF ;
Tai, CT ;
Yu, WC ;
Chen, YJ ;
Hsieh, MH ;
Chiang, CE ;
Ding, YA ;
Chang, MS ;
Chen, SA .
CIRCULATION, 1999, 100 (07) :768-771
[7]   CONTROL OF RADIOFREQUENCY LESION SIZE BY POWER REGULATION [J].
WITTKAMPF, FHM ;
HAUER, RNW ;
DEMEDINA, EOR .
CIRCULATION, 1989, 80 (04) :962-968