Dose Reduction in Coronary Artery Calcium Scoring Using Mono-Energetic Images from Reduced Tube Voltage Dual-Source Photon-Counting CT Data: A Dynamic Phantom Study

被引:29
|
作者
van der Werf, Niels R. [1 ]
van Gent, Margo [2 ]
Booij, Ronald [1 ]
Bos, Daniel [1 ]
van der Lugt, Aad [1 ]
Budde, Ricardo P. J. [1 ]
Greuter, Marcel J. W. [2 ,3 ]
van Straten, Marcel [1 ]
机构
[1] Erasmus MC, Dept Radiol & Nucl Med, NL-3015 GD Rotterdam, Netherlands
[2] Univ Groningen, Univ Med Ctr Groningen, Dept Radiol, NL-9713 GZ Groningen, Netherlands
[3] Univ Twente, Dept Robot & Mechatron, NL-7522 NB Enschede, Netherlands
关键词
X-ray computed tomography; calcium; coronary vessels; imaging phantoms; photon counting detector; radiation dose; image quality; MULTISLICE COMPUTED-TOMOGRAPHY; HEART-RATE; IN-VITRO; QUANTIFICATION; AGATSTON; DISEASE; SCORES; REPRODUCIBILITY; CALCIFICATION; PERFORMANCE;
D O I
10.3390/diagnostics11122192
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In order to assess coronary artery calcium (CAC) quantification reproducibility for photon-counting computed tomography (PCCT) at reduced tube potential, an anthropomorphic thorax phantom with low-, medium-, and high-density CAC inserts was scanned with PCCT (NAEOTOM Alpha, Siemens Healthineers) at two heart rates: 0 and 60-75 beats per minute (bpm). Five imaging protocols were used: 120 kVp standard dose (IQ level 16, reference), 90 kVp at standard (IQ level 16), 75% and 45% dose and tin-filtered 100 kVp at standard dose (IQ level 16). Each scan was repeated five times. Images were reconstructed using monoE reconstruction at 70 keV. For each heart rate, CAC values, quantified as Agatston scores, were compared with the reference, whereby deviations >10% were deemed clinically relevant. Reference protocol radiation dose (as volumetric CT dose index) was 4.06 mGy. Radiation dose was reduced by 27%, 44%, 67%, and 46% for the 90 kVp standard dose, 90 kVp 75% dose, 90 kVp 45% dose, and Sn100 standard dose protocol, respectively. For the low-density CAC, all reduced tube current protocols resulted in clinically relevant differences with the reference. For the medium- and high-density CAC, the implemented 90 kVp protocols and heart rates revealed no clinically relevant differences in Agatston score based on 95% confidence intervals. In conclusion, PCCT allows for reproducible Agatston scores at a reduced tube voltage of 90 kVp with radiation dose reductions up to 67% for medium- and high-density CAC.
引用
收藏
页数:9
相关论文
共 27 条
  • [21] Automated tube voltage selection for radiation dose and contrast medium reduction at coronary CT angiography using 3rd generation dual-source CT
    Mangold, Stefanie
    Wichmann, Julian L.
    Schoepf, U. Joseph
    Poole, Zachary B.
    Canstein, Christian
    Varga-Szemes, Akos
    Caruso, Damiano
    Bamberg, Fabian
    Nikolaou, Konstantin
    De Cecco, Carlo N.
    EUROPEAN RADIOLOGY, 2016, 26 (10) : 3608 - 3616
  • [22] Automated tube voltage selection for radiation dose and contrast medium reduction at coronary CT angiography using 3rd generation dual-source CT
    Stefanie Mangold
    Julian L. Wichmann
    U. Joseph Schoepf
    Zachary B. Poole
    Christian Canstein
    Akos Varga-Szemes
    Damiano Caruso
    Fabian Bamberg
    Konstantin Nikolaou
    Carlo N. De Cecco
    European Radiology, 2016, 26 : 3608 - 3616
  • [23] Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom study
    Felix G. Meinel
    Christian Canstein
    U. Joseph Schoepf
    Martin Sedlmaier
    Bernhard Schmidt
    Brett S. Harris
    Thomas G. Flohr
    Carlo N. De Cecco
    European Radiology, 2014, 24 : 1643 - 1650
  • [24] Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom study
    Meinel, Felix G.
    Canstein, Christian
    Schoepf, U. Joseph
    Sedlmaier, Martin
    Schmidt, Bernhard
    Harris, Brett S.
    Flohr, Thomas G.
    De Cecco, Carlo N.
    EUROPEAN RADIOLOGY, 2014, 24 (07) : 1643 - 1650
  • [25] Dose reduction in spiral CT coronary angiography with dual-source equipment. Part I. A phantom study applying different prospective tube current modulation algorithms
    Martini, C.
    Palumbo, A.
    Maffei, E.
    Rossi, A.
    Rengo, M.
    Malago, R.
    Dijkshoorn, M.
    Weustink, A.
    Mollet, N.
    Krestin, G.
    Cademartiri, F.
    RADIOLOGIA MEDICA, 2009, 114 (07): : 1037 - 1052
  • [26] Validation of Dual-Source Single-Tube Reconstruction as a Method to Obtain Half-Dose Images to Evaluate Radiation Dose and Noise Reduction: Phantom and Human Assessment Using CT Colonography and Sinogram-Affirmed Iterative Reconstruction (SAFIRE)
    Fletcher, Joel G.
    Grant, Katharine L. R.
    Fidler, Jeff L.
    Shiung, Maria
    Yu, Lifeng
    Wang, Jia
    Schmidt, Bernhard
    Allmendinger, Thomas
    McCollough, Cynthia H.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2012, 36 (05) : 560 - 569
  • [27] Dose reduction in spiral CT coronary angiography with dual-source equipment. Part I. A phantom study applying different prospective tube current modulation algorithms [Riduzione della dose in angiografia coronarica con TC spirale con apparecchiatura a doppia sorgente. Parte I. Studio su fantoccio della dose totale applicando differenti algoritmi di modulazione prospettica della corrente del tubo]
    Martini C.
    Palumbo A.
    Maffei E.
    Rossi A.
    Rengo M.
    Malagò R.
    Dijkshoorn M.
    Weustink A.
    Mollet N.
    Krestin G.
    Cademartiri F.
    La radiologia medica, 2009, 114 (7) : 1037 - 1052