ScatterNet: A convolutional neural network for cone-beam CT intensity correction

被引:134
作者
Hansen, David C. [1 ]
Landry, Guillaume [2 ]
Kamp, Florian [3 ]
Li, Minglun [3 ]
Belka, Claus [3 ,4 ]
Parodi, Katia [2 ]
Kurz, Christopher [2 ,3 ]
机构
[1] Aarhus Univ Hosp, Dept Med Phys, DK-8200 Aarhus, Denmark
[2] Ludwig Maximilians Univ Munchen LMU Munich, Fac Phys, Dept Med Phys, D-85748 Garching, Germany
[3] Ludwig Maximilians Univ Munchen, Univ Hosp, Dept Radiat Oncol, D-81377 Munich, Germany
[4] German Canc Consortium DKTK, Munich, Germany
关键词
adaptive radiotherapy; CBCT imaging; deep learning; DEFORMABLE IMAGE REGISTRATION; FLAT-PANEL IMAGER; COMPUTED-TOMOGRAPHY; GENERAL FRAMEWORK; DOSE CALCULATION;
D O I
10.1002/mp.13175
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Methods To demonstrate a proof-of-concept for fast cone-beam CT (CBCT) intensity correction in projection space by the use of deep learning. The CBCT scans and corresponding projections were acquired from 30 prostate cancer patients. Reference shading correction was performed using a validated method (CBCTcor), which estimates scatter and other low-frequency deviations in the measured CBCT projections on the basis of a prior CT image obtained from warping the planning CT to the CBCT. A convolutional neural network (ScatterNet) was designed, consisting of an attenuation conversion stage followed by a shading correction stage using a UNet-like architecture. The combined network was trained in 2D, utilizing pairs of measured and corrected projections of the reference method, in order to perform shading correction in projection space before reconstruction. The number of patients used for training, testing, and evaluation was 15, 7, and 8, respectively. The reconstructed CBCTScatterNet was compared to CBCTcor in terms of mean and absolute errors (ME and MAE) for the eight evaluation patients (not included in the network training). Volumetric modulated arc photon therapy (VMAT) and intensity-modulated proton therapy (IMPT) plans were generated on CBCTcor. Dose was recalculated on CBCTScatterNet to evaluate its dosimetric accuracy. Single-field uniform dose proton plans were utilized for proton range comparison of CBCTScatterNet and CBCTcor. Results Conclusions The CBCTScatterNet showed no cupping artifacts and a considerably smaller MAE and ME with respect to CBCTcor than the uncorrected CBCT (on average 144 Hounsfield units (HU) vs 46 HU for MAE and 138 HU vs -3 HU for ME). The pass-rates using a 2% dose-difference criterion at 50% dose cut-off, were close to 100% for the VMAT plans of all patients when comparing CBCTScatterNet to CBCTcor. For IMPT plans pass-rates were clearly lower, ranging from 15% to 81%. Proton range differences of up to 5 mm occurred. Using a deep convolutional neural network for CBCT intensity correction was shown to be feasible in the pelvic region for the first time. Dose calculation accuracy on CBCTScatterNet was high for VMAT, but unsatisfactory for IMPT. With respect to the reference technique (CBCTcor), the neural network enabled a considerable increase in speed for intensity correction and might eventually allow for on-the-fly shading correction during CBCT acquisition.
引用
收藏
页码:4916 / 4926
页数:11
相关论文
共 37 条
[1]  
Bertram M, 2008, P SPIE
[2]   Multi-criterial patient positioning based on dose recalculation on scatter-corrected CBCT images [J].
Hofmaier, Jan ;
Haehnle, Jonas ;
Kurz, Christopher ;
Landry, Guillaume ;
Maihoefer, Cornelius ;
Schuettrumpf, Lars ;
Suess, Philipp ;
Teichert, Katrin ;
Soehn, Matthias ;
Spahr, Nadine ;
Brachmann, Christoph ;
Weiler, Florian ;
Thieke, Christian ;
Kuefer, Karl-Heinz ;
Belka, Claus ;
Parodi, Katia ;
Kamp, Florian .
RADIOTHERAPY AND ONCOLOGY, 2017, 125 (03) :464-469
[3]   A robotic C-arm cone beam CT system for image-guided proton therapy: design and performance [J].
Hua, Chiaho ;
Yao, Weiguang ;
Kidani, Takao ;
Tomida, Kazuo ;
Ozawa, Saori ;
Nishimura, Takenori ;
Fujisawa, Tatsuya ;
Shinagawa, Ryousuke ;
Merchant, Thomas E. .
BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1079)
[4]  
Kingma D. P., P 3 INT C LEARN REPR
[5]   Investigating deformable image registration and scatter correction for CBCT-based dose calculation in adaptive IMPT [J].
Kurz, Christopher ;
Kamp, Florian ;
Park, Yang-Kyun ;
Zoellner, Christoph ;
Rit, Simon ;
Hansen, David ;
Podesta, Mark ;
Sharp, Gregory C. ;
Li, Minglun ;
Reiner, Michael ;
Hofmaier, Jan ;
Neppl, Sebastian ;
Thieke, Christian ;
Nijhuis, Reinoud ;
Ganswindt, Ute ;
Belka, Claus ;
Winey, Brian A. ;
Parodi, Katia ;
Landry, Guillaume .
MEDICAL PHYSICS, 2016, 43 (10) :5635-5646
[6]  
Li DP, 2015, PROC CVPR IEEE, P213, DOI 10.1109/CVPR.2015.7298617
[7]  
Maier J, 2018, P SPIE
[8]   Variance reduction techniques for fast Monte Carlo CBCT scatter correction calculations [J].
Mainegra-Hing, Ernesto ;
Kawrakow, Iwan .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (16) :4495-4507
[9]   USE OF NORMAL TISSUE COMPLICATION PROBABILITY MODELS IN THE CLINIC [J].
Marks, Lawrence B. ;
Yorke, Ellen D. ;
Jackson, Andrew ;
Ten Haken, Randall K. ;
Constine, Louis S. ;
Eisbruch, Avraham ;
Bentzen, Soren M. ;
Nam, Jiho ;
Deasy, Joseph O. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (03) :S10-S19
[10]   A fast and pragmatic approach for scatter correction in flat-detector CT using elliptic modeling and iterative optimization [J].
Meyer, Michael ;
Kalender, Willi A. ;
Kyriakou, Yiannis .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (01) :99-120