Comparison of measured and calculated emission probabilities for K and L X-rays following radioactive disintegration processes

被引:5
作者
Schönfeld, E [1 ]
Schötzig, U [1 ]
机构
[1] Phys Tech Bundesanstalt, Lab 6 11, D-38116 Braunschweig, Germany
关键词
KX-rays; L X-rays; Ce-139; Hg-203; Am-241; X-ray emission probabilities;
D O I
10.1016/S0969-8043(00)00321-3
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The program EMISSION has been developed at the Physikalisch-Technische Bundesanstalt (PTB). It allows emission probabilities to be calculated for K and L X-rays following radioactive disintegration processes. This program was already applied to the calculation of total K X-ray emission probabilities for six radionuclides. The results are found to be in reasonable agreement with measured values. In order to check the validity of this program also for the individual components of K and L X-rays, calculated emission probabilities of K X-ray components (Ce-139, Hg-203) and L X-ray components (Ce-139, Hg-203, Am-241) were compared with remeasured (Ce-139, Hg-203) and already available (Am-241) emission probabilities, measured earlier at the PTB. In most cases there is an agreement within the standard uncertainties of the measured and calculated values, usually within 1-5%. Larger deviations, not covered by three times the standard uncertainty have been found for some weak lines. (C) 2001 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:785 / 791
页数:7
相关论文
共 23 条
[1]   X-RAY WAVELENGTHS [J].
BEARDEN, JA .
REVIEWS OF MODERN PHYSICS, 1967, 39 (01) :78-&
[2]   DETERMINATION OF PU-239 AND AM-241 LX-RAY INTENSITIES USING A SIMPLEX-METHOD FOR FITTING PEAKS [J].
BLAND, CJ ;
MOREL, J ;
ETCHEVERRY, E ;
LEPY, MC .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 312 (1-2) :323-333
[3]  
CHECHEV VP, 1999, UNPUB DATA EVALUATIO
[5]   LIMITATIONS OF PULSER METHOD FOR PILE-UP CORRECTIONS IN GE(LI)-SPECTROMETRY [J].
DEBERTIN, K ;
SCHOTZIG, U .
NUCLEAR INSTRUMENTS & METHODS, 1977, 140 (02) :337-340
[6]   COINCIDENCE SUMMING CORRECTIONS IN GE(LI)-SPECTROMETRY AT LOW SOURCE-TO-DETECTOR DISTANCES [J].
DEBERTIN, K ;
SCHOTZIG, U .
NUCLEAR INSTRUMENTS & METHODS, 1979, 158 (2-3) :471-477
[7]   STANDARD SOURCES FOR THE EFFICIENCY CALIBRATION OF PHOTON DETECTORS IN THE ENERGY-RANGE FROM 5 TO 40 KEV [J].
DEBERTIN, K ;
FURNARI, JC ;
WALZ, KF .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1985, 36 (12) :977-980
[8]  
GEIGER JS, 1965, P NUCL PHYS, V68, P352
[9]  
HANSEN HH, 1981, 171 FIZ KARLSR
[10]  
HANSEN HH, 1985, 172 FIZ KARLSR