Exact and efficient cone-beam reconstruction algorithm for a short-scan circle combined with various lines

被引:7
作者
Dennerlein, F [1 ]
Katsevich, A [1 ]
Lauritsch, G [1 ]
Hornegger, J [1 ]
机构
[1] Siemens Med Solut, D-91301 Frochheim, Germany
来源
MEDICAL IMAGING 2005: IMAGE PROCESSING, PT 1-3 | 2005年 / 5747卷
关键词
image reconstruction; computed tomography; cone-beam; theoretically exact; filtered backprojection algorithm; C-arm imaging; circle and line; data completeness; geometric calibration;
D O I
10.1117/12.595186
中图分类号
TB8 [摄影技术];
学科分类号
0804 ;
摘要
X-ray 3D rotational angiography based on C-arm systems has become a versatile and established tomographic imaging modality for high contrast objects in interventional environment. Improvements in data acquisition, e.g. by use of flat panel detectors, will enable C-arm systems to resolve even low-contrast details. However, further progress will be limited by the incompleteness of data acquisition on the conventional short-scan circular source trajectories. Cone artifacts, which result from that incompleteness, significantly degrade image quality by severe smearing and shading. To assure data completeness a combination of a partial circle with one or several line segments is investigated. A new and efficient reconstruction algorithm is deduced from a general inversion formula based on 3D Radon theory. The method is theoretically exact, possesses shift-invariant filtered backprojection (FBP) structure, and solves the long object problem. The algorithm is flexible in dealing with various circle and line configurations. The reconstruction method requires nothing more than the theoretically minimum length of scan trajectory. It consists of a conventional short-scan circle and a line segment approximately twice as long as the height of the region-of-interest. Geometrical deviations from the ideal source trajectory are considered in the implementation in order to handle data of real C-arm systems. Reconstruction results show excellent image quality free of cone artifacts. The proposed scan trajectory and reconstruction algorithm assure excellent image quality and allow low-contrast tomographic imaging with C-arm based cone-beam systems. The method can be implemented without any hardware modifications on systems commercially available today.
引用
收藏
页码:388 / 399
页数:12
相关论文
共 50 条
  • [1] A cone-beam FBP reconstruction algorithm for short-scan and super-short-scan source trajectories
    Zhuang, TL
    Nett, B
    Tang, XY
    Chen, GH
    DEVELOPMENTS IN X-RAY TOMOGRAPHY IV, 2004, 5535 : 566 - 576
  • [2] A factorization approach for cone-beam reconstruction on a circular short-scan
    Dennerlein, Frank
    Noo, Frederic
    Schoendube, Harald
    Lauritsch, Guenter
    Hornegger, Joachim
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2008, 27 (07) : 887 - 896
  • [3] The fan-beam short-scan FBP algorithm is not exact
    Zeng, G. L.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (08) : N131 - N139
  • [4] A new super-short-scan algorithm for fan-beam and cone-beam reconstruction
    Li, LA
    Chen, ZQ
    Kang, KJ
    Zhao, ZR
    IMAGE RECONSTRUCTION FROM INCOMPLETE DATA III, 2004, 5562 : 80 - 87
  • [5] Exact image reconstruction in a helical cone-beam scan with a variable pitch
    Zou, Y
    Pan, XC
    Xia, D
    Wang, G
    Jiang, TZ
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 4200 - 4203
  • [6] New super-short-scan algorithms for fan-beam and cone-beam reconstruction
    Kudo, H
    Noo, F
    Defrise, M
    Clackdoyle, R
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 902 - 906
  • [7] Cone-beam reconstruction of flat specimens in a super-short scan
    Wang, Xianchao
    Tang, Ziyue
    Zhu, Zhenbo
    Jiang, Chengshun
    INSIGHT, 2015, 57 (10) : 571 - 575
  • [8] An exact reconstruction algorithm for triple-source helical cone-beam CT
    Zhao, Jun
    Jiang, Ming
    Zhuang, Tiange
    Wang, Ge
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2006, 14 (03) : 191 - 206
  • [9] Feldkamp-type VOI reconstruction from super-short-scan cone-beam data
    Yu, HY
    Wang, G
    MEDICAL PHYSICS, 2004, 31 (06) : 1357 - 1362
  • [10] Comparing Short Scan CT Reconstruction Algorithms Regarding Cone-Beam Artifact Performance
    Maass, Clemens
    Dennerlein, Frank
    Noo, Frederic
    Kachelriess, Marc