High-sensitivity mid-infrared heterodyne spectrometer with a tunable diode laser as a local oscillator

被引:20
作者
Schmulling, F [1 ]
Klumb, B [1 ]
Harter, M [1 ]
Schieder, R [1 ]
Vowinkel, B [1 ]
Winnewisser, G [1 ]
机构
[1] Univ Cologne, Inst Phys 1, D-50937 Koln, Germany
来源
APPLIED OPTICS | 1998年 / 37卷 / 24期
关键词
D O I
10.1364/AO.37.005771
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A new mid-IR heterodyne spectrometer, which is intended to be applied for atmospheric and astrophysical studies, is presented. The spectrometer uses a frequency-stabilized tunable diode laser as a local oscillator. Owing to the low output power of available single-mode diode lasers, a newly developed confocal-ring resonator, the diplexer, is used to superimpose the source signal efficiently with that of the local oscillator. Additionally, the diplexer serves as an optical filter that establishes controlled optical feedback between the laser diode and the detector, which allows stable laser operation with linewidths of the order of 1 MHz. The heterodyne signal from the HgCdTe detector is analyzed by means of a 1.4-GHz acousto-optical spectrometer. With this setup we find system temperatures as low as 4400 K (double sideband), that is, approximately a factor of 6 of the quantum Limit. (C) 1998 Optical Society of America.
引用
收藏
页码:5771 / 5776
页数:6
相关论文
共 23 条
[1]   SENSITIVITY LIMITS OF AN INFRARED HETERODYNE SPECTROMETER FOR ASTROPHYSICAL APPLICATIONS [J].
ABBAS, MM ;
MUMMA, MJ ;
KOSTIUK, T ;
BUHL, D .
APPLIED OPTICS, 1976, 15 (02) :427-436
[3]   MEASUREMENTS OF FOREIGN GAS-PRESSURE SHIFT AND BROADENING EFFECTS IN THE (1-0) BAND OF CO WITH N2 AND AR [J].
ANSELM, N ;
YAMADA, KMT ;
SCHIEDER, R ;
WINNEWISSER, G .
JOURNAL OF MOLECULAR SPECTROSCOPY, 1993, 161 (01) :284-296
[4]   SIGNAL-TO-NOISE RATIO AND OTHER CHARACTERISTICS OF HETERODYNE RADIATION RECEIVERS [J].
BLANEY, TG .
SPACE SCIENCE REVIEWS, 1975, 17 (05) :691-702
[5]   Atmospheric laser heterodyne detection [J].
Delahaigue, A ;
Courtois, D ;
Thiebeaux, C ;
Kalite, S ;
Parvitte, B .
INFRARED PHYSICS & TECHNOLOGY, 1996, 37 (01) :7-12
[6]   LASER HETERODYNE SPECTROMETER USING A LIQUID-NITROGEN COOLED TUNABLE DIODE-LASER FOR REMOTE MEASUREMENTS OF ATMOSPHERIC O-3 AND N2O [J].
FUKUNISHI, H ;
OKANO, S ;
TAGUCHI, M ;
OHNUMA, T .
APPLIED OPTICS, 1990, 29 (18) :2722-2728
[7]   TUNABLE DIODE-LASER HETERODYNE SPECTROMETER FOR REMOTE OBSERVATIONS NEAR 8-MU-M [J].
GLENAR, D ;
KOSTIUK, T ;
JENNINGS, DE ;
BUHL, D ;
MUMMA, MJ .
APPLIED OPTICS, 1982, 21 (02) :253-259
[8]  
HACHENBERG O, 1982, TECHNISCHE GRUNDLAGE
[9]  
HARRIS AI, 1991, HOUCHES SERIES, P7
[10]   OPTICAL FEEDBACK EFFECTS ON THE PERFORMANCE OF PB1-XSNXSE SEMICONDUCTOR-LASERS [J].
HARWARD, CN ;
HOELL, JM .
APPLIED OPTICS, 1979, 18 (23) :3978-3983