Towards automatic calibration of Fourier-Domain OCT for robot-assisted vitreoretinal surgery

被引:39
作者
Liu, Xuan [1 ]
Balicki, Marcin [2 ]
Taylor, Russell H. [2 ]
Kang, Jin U. [1 ]
机构
[1] Johns Hopkins Univ, Dept Elect & Comp Engn, Baltimore, MD 21211 USA
[2] Johns Hopkins Univ, Dept Comp Sci, Baltimore, MD 21211 USA
基金
美国国家卫生研究院;
关键词
OPTICAL COHERENCE TOMOGRAPHY; TO-NOISE RATIO; SWEPT-SOURCE; HIGH-SPEED; ULTRAHIGH-RESOLUTION;
D O I
10.1364/OE.18.024331
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We present a new real-time automatic spectral calibration (ASC) method for Fourier domain optical coherence tomography (FD OCT) that can be automatically performed by the system. The ASC method proposed can be performed during OCT scanning operation and does not require an external calibrating light source or a commercial optical spectrum analyzer. Spectral data used for calibration can be interferograms obtained from an arbitrary sample which may have complicated internal structures, such as ones found in biological tissue. Moreover, our ASC method incorporates known robot motion to calibrate physical pixel spacing of the A-scan in static or dynamic environments. Experimental results show that our ASC method can provide high-performance calibration for FD OCT in terms of axial resolution and ranging accuracy without increasing hardware complexity. (C) 2010 Optical Society of America
引用
收藏
页码:24331 / 24343
页数:13
相关论文
共 26 条
[1]   Real-time and high-performance calibration method for high-speed swept-source optical coherence tomography [J].
Azimi, Ehsan ;
Liu, Bin ;
Brezinski, Mark E. .
JOURNAL OF BIOMEDICAL OPTICS, 2010, 15 (01)
[2]  
Balicki M, 2009, LECT NOTES COMPUT SC, V5761, P108, DOI 10.1007/978-3-642-04268-3_14
[3]  
Bouma B. E., 2002, Handbook of Optical Coherence Tomography, Patent No. 26406747
[4]   Ultrahigh-resolution high-speed retinal imaging using spectral-domain optical coherence tomography [J].
Cense, B ;
Nassif, NA ;
Chen, T ;
Pierce, M ;
Yun, SH ;
Park, BH ;
Bouma, BE ;
Tearney, GJ ;
de Boer, JF .
OPTICS EXPRESS, 2004, 12 (11) :2435-2447
[5]   Densely folded spectral images of a CCD spectrometer working in the full 200-1000 nm wavelength range with high resolution [J].
Chen, YR ;
Sun, B ;
Han, T ;
Kong, YF ;
Xu, CH ;
Zhou, P ;
Li, XF ;
Wang, SY ;
Zheng, YX ;
Chen, LY .
OPTICS EXPRESS, 2005, 13 (25) :10049-10054
[6]   Sensitivity advantage of swept source and Fourier domain optical coherence tomography [J].
Choma, MA ;
Sarunic, MV ;
Yang, CH ;
Izatt, JA .
OPTICS EXPRESS, 2003, 11 (18) :2183-2189
[7]   Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography [J].
de Boer, JF ;
Cense, B ;
Park, BH ;
Pierce, MC ;
Tearney, GJ ;
Bouma, BE .
OPTICS LETTERS, 2003, 28 (21) :2067-2069
[8]   A new spectral calibration method for Fourier domain optical coherence tomography [J].
Ding, Chao ;
Bu, Peng ;
Wang, Xiangzhao ;
Sasaki, Osami .
OPTIK, 2010, 121 (11) :965-970
[9]  
Fercher AF, 2003, REP PROG PHYS, V66, P239, DOI 10.2184/lsj.31.635
[10]  
Han J., 2009, P CLEO C