Performance of three biometry devices in patients with different grades of age-related cataract

被引:39
作者
Mylonas, Georgios [1 ]
Sacu, Stefan [1 ]
Buehl, Wolf [1 ]
Ritter, Markus [1 ]
Georgopoulos, Michael [1 ]
Schmidt-Erfurth, Ursula [1 ]
机构
[1] Med Univ Vienna, Dept Ophthalmol, A-1090 Vienna, Austria
关键词
biometry; cataract; LenStar; IOL-Master; Visante-OCT; PARTIAL COHERENCE INTERFEROMETRY; IOLMASTER; ULTRASOUND; PACHYMETRY; ACCURACY; ANALYZER;
D O I
10.1111/j.1755-3768.2010.02042.x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: The new Lenstar biometry device was compared in a typical clinical setting to the IOL-Master and Visante-OCT. Methods: Fifty-one eyes of 51 patients with age-related cataract were examined with Lenstar LS900 (Haag Streit AG) biometer, IOL-Master V.5 (Carl Zeiss Meditec AG) and Visante-OCT (Carl Zeiss Meditec AG) before cataract surgery. Central corneal thickness (CCT), anterior chamber depth (ACD), keratometry readings of flattest and steepest meridian (K), corneal radius (R) and axial length (AL) values were correlated. Cataracts were graded according to the Lens Opacities Classification System III (LOCS) regarding nuclear colour (NC), nuclear opalescence (NO), cortical (C) and posterior subcapsular (P) cataract. Results: Mean values and standard deviations for AL, K and R was 23.66 +/- 1.23 mm and 23.67 +/- 1.26 mm, 43.24 +/- 1.69 dpt and 43.16 +/- 1.71 dpt, 7.68 +/- 0.29 mm and 7.70 +/- 0.28 mm with the IOL-Master and with the LS900, respectively (r = 0.99 and p = 0.76, r = 0.99 and p = 0.029, r = 0.89 and p = 0.14, respectively). Visante-OCT demonstrated highest values of three devices regarding to ACD followed by Lenstar LS900 and IOLMaster. Axial length measurements were unfeasible in 10% of the cases (five patients) and this significantly correlated with the presence of posterior subcapsular cataract of LOCS III grade 4.0 or higher. Conclusions: IOL-Master, Lenstar LS900 and AC-OCT proved to be excellent non-contact measurement methods in eyes with age-related cataract. Nevertheless, ultrasound biometry is still required for cases with dense posterior subcapsular cataract.
引用
收藏
页码:e237 / e241
页数:5
相关论文
共 14 条
[1]   A new optical low coherence reflectometry device for ocular biometry in cataract patients [J].
Buckhurst, P. J. ;
Wolffsohn, J. S. ;
Shah, S. ;
Naroo, S. A. ;
Davies, L. N. ;
Berrow, E. J. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2009, 93 (07) :949-953
[2]   THE LENS OPACITIES CLASSIFICATION SYSTEM-III [J].
CHYLACK, LT ;
WOLFE, JK ;
SINGER, DM ;
LESKE, MC ;
BULLIMORE, MA ;
BAILEY, IL ;
FRIEND, J ;
MCCARTHY, D ;
WU, SY .
ARCHIVES OF OPHTHALMOLOGY, 1993, 111 (06) :831-836
[3]   Evaluation of the Lenstar LS 900 non-contact biometer [J].
Cruysberg, L. P. J. ;
Doors, M. ;
Verbakel, F. ;
Berendschot, T. T. J. M. ;
De Brabander, J. ;
Nuijts, R. M. M. A. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2010, 94 (01) :106-110
[4]   The impact of cataract severity on measurement acquisition with the IOLMaster [J].
Freeman, G ;
Pesudovs, K .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2005, 83 (04) :439-442
[5]   Evaluation of a new IOLMaster algorithm to measure axial length [J].
Hill, Warren ;
Angeles, Raymund ;
Otani, Todd .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (06) :920-924
[6]   Accuracy of a new partial coherence interferometry analyser for biometric measurements [J].
Holzer, M. P. ;
Mamusa, M. ;
Auffarth, G. U. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2009, 93 (06) :807-810
[7]   Comparison of anterior chamber depth measurements using the IOLMaster, scanning peripheral anterior chamber depth analyser, and anterior segment optical coherence tomography [J].
Lavanya, Raghavan ;
Teo, Livia ;
Friedman, David S. ;
Aung, Han T. ;
Baskaran, Mani ;
Gao, Hong ;
Alfred, Tamuno ;
Seah, Steve K. ;
Kashiwagi, Kenji ;
Foster, Paul J. ;
Aung, Tin .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2007, 91 (08) :1023-1026
[8]   The repeatability of corneal thickness measures [J].
Marsich, MM ;
Bullimore, MA .
CORNEA, 2000, 19 (06) :792-795
[9]   Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness [J].
Miglior, S ;
Albe, E ;
Guareschi, M ;
Mandelli, G ;
Gomarasca, S ;
Orzalesi, N .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2004, 88 (02) :174-177
[10]   Central corneal thickness measurements with partial coherence interferometry, ultrasound, and the Orbscan system [J].
Rainer, G ;
Findl, O ;
Petternel, V ;
Kiss, B ;
Drexler, W ;
Skorpik, C ;
Georgopoulos, M ;
Schmetterer, L .
OPHTHALMOLOGY, 2004, 111 (05) :875-879