Radiation dosimetry with fiber Bragg gratings

被引:3
作者
Avino, S. [1 ]
D'Avino, V. [2 ]
Giorgini, A. [1 ]
Pacelli, R. [3 ]
Liuzzi, R. [2 ]
Cella, L. [2 ]
Gagliardi, G. [1 ,2 ]
De Natale, P. [1 ]
机构
[1] CNR, INO, Via Campi Flegrei,34 Comprensorio A Olivetti, I-80078 Naples, Italy
[2] Inst Biostructure & Bioimaging, Consiglio Nazl Ric, I-80131 Naples, Italy
[3] Univ Napoli Federico II, Dipartimento Diagnost Immagini Radioterapia, I-80131 Naples, Italy
来源
23RD INTERNATIONAL CONFERENCE ON OPTICAL FIBRE SENSORS | 2014年 / 9157卷
关键词
OPTICAL-FIBERS; CHAMBER; PHASE;
D O I
10.1117/12.2059067
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The measurement and monitoring of radiation dose delivered in patient tissues is a critical aspect in radiation therapy. Various dosimeters have proven effective in measuring radiations at low doses. However, there is a growing demand for new dosimeters based on small, non-invasive and high resolution devices. Here we report on a miniature dosimeter based on an optical fiber cavity. We demonstrate an ultimate detection limit of 160 mGy with an effective interaction region of 6 x 10(-4) mm(3). Due to its reliability, compactness and biomedical dose level sensitivity, our system shows itself suitable for applications in radiation therapy dosimetry.
引用
收藏
页数:4
相关论文
共 15 条
[1]   A new formalism for reference dosimetry of small and nonstandard fields [J].
Alfonso, R. ;
Andreo, P. ;
Capote, R. ;
Huq, M. Saiful ;
Kilby, W. ;
Kjaell, P. ;
Mackie, T. R. ;
Palmans, H. ;
Rosser, K. ;
Seuntjens, J. ;
Ullrich, W. ;
Vatnitsky, S. .
MEDICAL PHYSICS, 2008, 35 (11) :5179-5186
[2]   Diamond detector versus silicon diode and ion chamber in photon beams of different energy and field size [J].
Bucciolini, M ;
Buonamici, FB ;
Mazzocchi, S ;
De Angelis, C ;
Onori, S ;
Cirrone, GAP .
MEDICAL PHYSICS, 2003, 30 (08) :2149-2154
[3]   PI-PHASE-SHIFTED PERIODIC DISTRIBUTED STRUCTURES IN OPTICAL FIBERS BY UV POST-PROCESSING [J].
CANNING, J ;
SCEATS, MG .
ELECTRONICS LETTERS, 1994, 30 (16) :1344-1345
[4]   The Italian affair: The employment of parallel-plate ionization chambers for dose measurements in high dose-per-pulse IORT electron beams [J].
Cella, L. ;
Liuzzi, R. ;
Salvatore, M. .
MEDICAL PHYSICS, 2010, 37 (06) :2918-2924
[5]  
Cella L, 2001, PHYS MEDICA, V17, P100
[6]   Investigation of the use of MOSFET for clinical IMRT dosimetric verification [J].
Chuang, CF ;
Verhey, LJ ;
Xia, P .
MEDICAL PHYSICS, 2002, 29 (06) :1109-1115
[7]   LASER PHASE AND FREQUENCY STABILIZATION USING AN OPTICAL-RESONATOR [J].
DREVER, RWP ;
HALL, JL ;
KOWALSKI, FV ;
HOUGH, J ;
FORD, GM ;
MUNLEY, AJ ;
WARD, H .
APPLIED PHYSICS B-PHOTOPHYSICS AND LASER CHEMISTRY, 1983, 31 (02) :97-105
[8]   Probing the Ultimate Limit of Fiber-Optic Strain Sensing [J].
Gagliardi, G. ;
Salza, M. ;
Avino, S. ;
Ferraro, P. ;
De Natale, P. .
SCIENCE, 2010, 330 (6007) :1081-1084
[9]   Remote optical fiber dosimetry [J].
Huston, AL ;
Justus, BL ;
Falkenstein, PL ;
Miller, RW ;
Ning, H ;
Altemus, R .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2001, 184 (1-2) :55-67
[10]  
International Commission on Radiation Units and Measurements (ICRU), 1980, 33 ICRU