Tissue Dielectric Constant and Water Displacement Method Can Detect Changes of Mild Breast Cancer-Related Arm Lymphedema

被引:11
作者
Karlsson, Katarina [1 ,2 ]
Johansson, Karin [1 ]
Nilsson-Wikmar, Lena [3 ]
Brogardh, Christina [1 ,4 ]
机构
[1] Lund Univ, Fac Med, Dept Hlth Sci, Lund, Sweden
[2] Karolinska Univ Hosp, Physiotherapy Canc, S-17176 Stockholm, Sweden
[3] Karolinska Inst, Dept Neurobiol, Div Physiotherapy, Care Sci & Soc, Stockholm, Sweden
[4] Skane Univ Hosp, Dept Neurol Rehabil Med Memory Disorders & Geriat, Lund, Sweden
关键词
breast cancer; arm lymphedema; tissue dielectric constant; water displacement method; treatment; VOLUME; INSTRUMENTS; VALIDITY; WOMEN; TDC;
D O I
10.1089/lrb.2021.0010
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background: Most commonly, volume measurements are used to evaluate the effect of lymphedema treatment, but as the accumulation of lymph fluid can be local, this method may not always be the best. Tissue dielectric constant (TDC) can be applied to identify local lymphedema changes, but has not been used before when evaluating treatment in mild arm lymphedema. Thus, the overall aim of this study was to examine if TDC and water displacement method (WDM) can measure changes in mild breast cancer-related lymphedema during the 6-month standard treatment. More specifically, we examined changes within and between three defined groups based on lymphedema thresholds of TDC and WDM at start of treatment, as well as changes of the highest TDC ratio and site. Methods and Results: Forty-six women with mild arm lymphedema, received treatment with compression sleeves, mostly ccl 1, and instructions about self-care. Local tissue water was measured by TDC at six defined sites and lymphedema relative volume (LRV) by WDM before treatment and at first, second, third, and sixth month. There was a significant decrease in the site with the highest TDC ratio, as well as LRV at all follow-up visits. At 6 months, TDC ratio had decreased mean 0.26 (p < 0.001) and LRV mean - 3.3% (p < 0.001). There was a significant difference between the groups in change of TDC ratio, but not in LRV. Sixty percent changed the overall highest TDC ratio to another site during 6 months. Conclusion: Both TDC and WDM could detect changes in mild arm lymphedema but should be interpreted separately.
引用
收藏
页码:325 / 334
页数:10
相关论文
共 23 条
[1]   Early Detection of Lymphatic Disorder and Treatment for Lymphedema following Breast Cancer [J].
Akita, Shinsuke ;
Nakamura, Rikiya ;
Yamamoto, Naohito ;
Tokumoto, Hideki ;
Ishigaki, Tatsuya ;
Yamaji, Yoshihisa ;
Sasahara, Yoshitaro ;
Kubota, Yoshitaka ;
Mitsukawa, Nobuyuki ;
Satoh, Kaneshige .
PLASTIC AND RECONSTRUCTIVE SURGERY, 2016, 138 (02) :192E-202E
[2]   Can tissue dielectric constant measurement aid in differentiating lymphoedema from lipoedema in women with swollen legs? [J].
Birkballe, S. ;
Jensen, M. R. ;
Noerregaard, S. ;
Gottrup, F. ;
Karlsmark, T. .
BRITISH JOURNAL OF DERMATOLOGY, 2014, 170 (01) :96-102
[3]   Reliability of the MoistureMeterD Compact Device and the Pitting Test to Evaluate Local Tissue Water in Subjects with Breast Cancer-Related Lymphedema [J].
De Vrieze, Tessa ;
Gebruers, Nick ;
Nevelsteen, Ines ;
De Groef, An ;
Tjalma, Wiebren A. A. ;
Thomis, Sarah ;
Dams, Lore ;
Van der Gucht, Elien ;
Penen, Frauke ;
Devoogdt, Nele .
LYMPHATIC RESEARCH AND BIOLOGY, 2020, 18 (02) :116-128
[4]   Volumetric evaluation of upper extremities in 250 healthy persons [J].
Gebruers, N. ;
Truijen, S. ;
Engelborghs, S. ;
De Deyn, P. P. .
CLINICAL PHYSIOLOGY AND FUNCTIONAL IMAGING, 2007, 27 (01) :17-22
[5]   Measurement Properties of Instruments for Measuring of Lymphedema: Systematic Review [J].
Hidding, Janine T. ;
Viehoff, Peter B. ;
Beurskens, Carien H. G. ;
van Laarhoven, Hanneke W. M. ;
Sanden, Maria W. G. Nijhuis-van der ;
van der Wees, Philip J. .
PHYSICAL THERAPY, 2016, 96 (12) :1965-1981
[6]  
Johansson K., 2001, ADV PHYSIOTHERAPY, V3, P55, DOI DOI 10.1080/14038190119371
[7]   Arm lymphoedema in a cohort of breast cancer survivors 10 years after diagnosis [J].
Johansson, Karin ;
Branje, Elin .
ACTA ONCOLOGICA, 2010, 49 (02) :166-173
[8]   Concurrent validity of upper-extremity volume estimates: Comparison of calculated volume derived from girth measurements and water displacement volume [J].
Karges, JR ;
Mark, BE ;
Stikeleather, SJ ;
Worrell, TW .
PHYSICAL THERAPY, 2003, 83 (02) :134-145
[9]   Palpation of Increased Skin and Subcutaneous Thickness, Tissue Dielectric Constant, and Water Displacement Method for Diagnosis of Early Mild Arm Lymphedema [J].
Karlsson, Katarina ;
Nilsson-Wikmar, Lena ;
Brogardh, Christina ;
Johansson, Karin .
LYMPHATIC RESEARCH AND BIOLOGY, 2020, 18 (03) :219-225
[10]   Experimental and Analytical Comparisons of Tissue Dielectric Constant (TDC) and Bioimpedance Spectroscopy (BIS) in Assessment of Early Arm Lymphedema in Breast Cancer Patients after Axillary Surgery and Radiotherapy [J].
Lahtinen, Tapani ;
Seppala, Jan ;
Viren, Tuomas ;
Johansson, Karin .
LYMPHATIC RESEARCH AND BIOLOGY, 2015, 13 (03) :176-185