Metal artifact reduction using virtual monochromatic images for patients with pedicle screws implants on CT

被引:62
作者
Dong, Yue [1 ,2 ]
Shi, Ai Jun [1 ]
Wu, Jian Lin [3 ]
Wang, Ru Xin [1 ]
Sun, Li Fei [1 ]
Liu, Ai Lian [1 ]
Liu, Yi Jun [1 ]
机构
[1] Dalian Med Univ, Affiliated Hosp 1, Dept Radiol, Dalian 116011, Liaoning, Peoples R China
[2] DaLian Med Univ, Clin Oncol Coll, LiaoNing Canc Hosp, Dept Radiol, Shenyang 110042, Liaoning, Peoples R China
[3] Dalian Univ, Affiliated ZhongShan Hosp, Dept Radiol, Dalian 116001, Liaoning, Peoples R China
关键词
Spectral imaging; Virtual monochromatic image; Computed tomography; Metal artifacts; Pedicle screw; DUAL-ENERGY CT; SPINE; PLACEMENT; LUMBAR;
D O I
10.1007/s00586-015-4053-4
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
To evaluate the effect of spectral CT for metal artifact reduction in patients with pedicle screw. 45 patients with 119 pairs of pedicle screws underwent spectral CT examination. One set of conventional (140 kVp) polychromatic image and nine sets of virtual monochromatic images (60-140 keV) were obtained. The standard deviation (SD) of CT number in 12 locations around the implant and 1 on homogenous fat was measured to generate artifact index (AI). Objective assessment including AI, CT number and SD value was performed with independent t test and paired sample t test. Two radiologists independently reviewed the image quality, and Wilcoxon signed-rank test and kappa analysis were used for the subjective scores of image quality. The artifact index (AI) of all the regions decreased as keV increased. AIs of 100-140 keV were lower than that of 140 kVp images. At 120 keV there was no significant difference in CT numbers of psoas major muscle and vertebral canal between pedicle screw level and pedicle level, but a significant difference in SD value was determined between the two levels. The subjective scores at 100-140 keV were higher than the images at 140 kVp, and the highest subjective score of two observers and excellent interobserver agreement were found at 120 keV (kappa = 0.889). Virtual monochromatic images at high-energy levels have a well-concordant effect of removing metal artifacts, and 120 keV monochromatic images provided an accurate CT number and good subjective score.
引用
收藏
页码:1754 / 1763
页数:10
相关论文
共 22 条
  • [1] Accuracy of pedicle screw placement in the lumbosacral spine using conventional technique: computed tomography postoperative assessment in 102 consecutive patients Clinical article
    Amato, Vincenzo
    Giannachi, Luigi
    Irace, Claudio
    Corona, Claudio
    [J]. JOURNAL OF NEUROSURGERY-SPINE, 2010, 12 (03) : 306 - 313
  • [2] Comparative results between conventional and computer-assisted pedicle screw installation in the thoracic, lumbar, and sacral spine
    Amiot, LP
    Lang, K
    Putzier, M
    Zippel, H
    Labelle, H
    [J]. SPINE, 2000, 25 (05) : 606 - 614
  • [3] Artifacts in CT: Recognition and avoidance
    Barrett, JF
    Keat, N
    [J]. RADIOGRAPHICS, 2004, 24 (06) : 1679 - 1691
  • [4] Evaluation of Two Iterative Techniques for Reducing Metal Artifacts in Computed Tomography
    Boas, F. Edward
    Fleischmann, Dominik
    [J]. RADIOLOGY, 2011, 259 (03) : 894 - 902
  • [5] Spectral CT with Metal Artifacts Reduction Software for Improvement of Tumor Visibility in the Vicinity of Gold Fiducial Markers
    Brook, Olga R.
    Gourtsoyianni, Sofia
    Brook, Alexander
    Mahadevan, Anand
    Wilcox, Carol
    Raptopoulos, Vassilios
    [J]. RADIOLOGY, 2012, 263 (03) : 696 - 705
  • [6] Multichannel CT: Evaluating the spine in postoperative patients with orthopedic hardware
    Douglas-Akinwande, Annette C.
    Buckwalter, Kenneth A.
    Rydberg, Jonas
    Rankin, James L.
    Choplin, Robert H.
    [J]. RADIOGRAPHICS, 2006, 26 : S97 - U107
  • [7] Metallic artefact reduction with monoenergetic dual-energy CT: systematic ex vivo evaluation of posterior spinal fusion implants from various vendors and different spine levels
    Guggenberger, R.
    Winklhofer, S.
    Osterhoff, G.
    Wanner, G. A.
    Fortunati, M.
    Andreisek, G.
    Alkadhi, H.
    Stolzmann, P.
    [J]. EUROPEAN RADIOLOGY, 2012, 22 (11) : 2357 - 2364
  • [8] [惠萍 HUI Ping], 2011, [中华放射学杂志, Chinese Journal of Radiology], V45, P740
  • [9] Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multidetector CT
    Lee, Mi-Jung
    Kim, Sungjun
    Lee, Sung-Ah
    Song, Ho-Taek
    Huh, Yong-Min
    Kim, Dae-Hong
    Han, Seung Hwan
    Suh, Jin-Suck
    [J]. RADIOGRAPHICS, 2007, 27 (03) : 791 - 803
  • [10] Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software
    Lee, Young Han
    Park, Kwan Kyu
    Song, Ho-Taek
    Kim, Sungjun
    Suh, Jin-Suck
    [J]. EUROPEAN RADIOLOGY, 2012, 22 (06) : 1331 - 1340