VSImG: A high frame rate bitmap based display system for neuroscience research

被引:0
作者
Baptista de Almeida, Lirio Onofre [1 ]
Willem Slaets, Jan Frans [1 ]
Koeberle, Roland [1 ]
机构
[1] Univ Sao Paulo, Inst Phys Sao Carlos, Dept Phys & Informat, BR-13560970 Sao Carlos, SP, Brazil
基金
巴西圣保罗研究基金会;
关键词
Visual stimulus generation; Electrostatic deflection displays; Real-time control; High frame rate display; Neuroscience; INTEGRATION; ARTIFACTS; TIME;
D O I
10.1016/j.neucom.2011.02.016
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
This paper describes a visual stimulus generator (VSImG) capable of displaying a gray-scale, 256 x 256 x 8 bitmap image with a frame rate of 500 Hz using a boustrophedonic scanning technique. It is designed for experiments with motion-sensitive neurons of the fly's visual system, where the flicker fusion frequency of the photoreceptors can reach up to 500 Hz. Devices with such a high frame rate are not commercially available, but are required, if sensory systems with high flicker fusion frequency are to be studied. The implemented hardware approach gives us complete real-time control of the displacement sequence and provides all the signals needed to drive an electrostatic deflection display. With the use of analog signals, very small high-resolution displacements, not limited by the image's pixel size can be obtained. Very slow image displacements with visually imperceptible steps can also be generated. This can be of interest for other vision research experiments. Two different stimulus files can be used simultaneously, allowing the system to generate X-Y displacements on one display or independent movements on two displays as long as they share the same bitmap image. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:1762 / 1768
页数:7
相关论文
共 28 条
[1]  
AGNES A, 2008, PERCEPT PSYCHOPHYS, V70, P697, DOI DOI 10.3758/PP.70.4.697
[2]   Raster-scan cathode-ray tubes for vision research - Limits of resolution in space, time and intensity, and some solutions [J].
Bach, M ;
Meigen, T ;
Strasburger, H .
SPATIAL VISION, 1997, 10 (04) :403-414
[3]   Dynamically multilayered visual system of the multifractal fly [J].
Baptista, M. S. ;
Grebogi, Celso ;
Koberle, Roland .
PHYSICAL REVIEW LETTERS, 2006, 97 (17)
[4]   A complex biological system:: the fly's visual module [J].
Baptista, Murilo S. ;
de Almeida, Lirio O. B. ;
Slaets, Jan F. W. ;
Koberle, Roland ;
Grebogi, Celso .
PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 2008, 366 (1864) :345-357
[5]  
Brainard DH., 2002, Encyclopedia of imaging science and technology, P172, DOI DOI 10.1002/0471443395.IMG011
[6]  
ELZE T, 2007, J VISION, V7, P399
[7]   Recording from Two Neurons: Second-Order Stimulus Reconstruction from Spike Trains and Population Coding [J].
Fernandes, N. M. ;
Pinto, B. D. L. ;
Almeida, L. O. B. ;
Slaets, J. F. W. ;
Koeberle, R. .
NEURAL COMPUTATION, 2010, 22 (10) :2537-2557
[8]  
FOTOWAT H, 2007, J NEUROSCI, V27, P1122
[9]   Time-dependent activation of feed-forward inhibition in a looming-sensitive neuron [J].
Gabbiani, F ;
Cohen, I ;
Laurent, G .
JOURNAL OF NEUROPHYSIOLOGY, 2005, 94 (03) :2150-2161
[10]   Luminance artifacts of cathode-ray tube displays for vision research [J].
García-Pérez, MA ;
Peli, E .
SPATIAL VISION, 2001, 14 (02) :201-215