Client preferences for compression threshold in single-channel wide dynamic range compression hearing aids

被引:9
作者
Barker, C
Dillon, H
机构
[1] Australian Hearing, Maroochydore, Qld 4558, Australia
[2] Natl Acoust Labs, Chatswood, NSW 2067, Australia
关键词
D O I
10.1097/00003446-199904000-00004
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objective: Compression in hearing aids can be applied with low compression ratios over a wide range of input levels, but reverts to linear amplification below the compression threshold (CT). In this study, we aimed to determine which of two CTs was preferred by subjects as they used their hearing aids in their own environments, and whether they would prefer to have no low ratio compression at all. Design: Subjects were fitted with a multimemory hearing aid incorporating input controlled compression with a 2:1 compression ratio and output controlled compression limiting. The two memories contained identical programs except that they differed in CT. Sixteen mild to moderately sensorineurally hearing-impaired subjects compared low (approximate to 40 dB SPL) and moderate (approximate to 65 dB SPL) CTs over 2 mo of field trials using hand held remote controls to switch between the alternatives. In a third month's trial, the preferred option (which also included output controlled compression limiting) was compared with compression limiting alone, Results: The higher CT was preferred by 14 of the subjects. The combination of input compression and output compression limiting was preferred to compression limiting alone by 14 of the subjects. Conclusions: Several real world advantages of frequency independent 2:1 compression with a CT of about 65 dB SPL were demonstrated over linear amplification. Extending the compression to much lower input levels appear's to carry more disadvantages than advantages, at least for clients with mild and moderate hearing losses, when fitted with single-channel compression aids with a 2:1 compression ratio.
引用
收藏
页码:127 / 139
页数:13
相关论文
共 22 条
[1]  
Byrne D, 1996, J Am Acad Audiol, V7, P377
[2]   THE NATIONAL-ACOUSTIC-LABORATORIES (NAL) NEW PROCEDURE FOR SELECTING THE GAIN AND FREQUENCY-RESPONSE OF A HEARING-AID [J].
BYRNE, D ;
DILLON, H .
EAR AND HEARING, 1986, 7 (04) :257-265
[3]   THE INPUT/OUTPUT FORMULA - A THEORETICAL APPROACH TO THE FITTING OF PERSONAL AMPLIFICATION DEVICES [J].
CORNELISSE, LE ;
SEEWALD, RC ;
JAMIESON, DG .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1995, 97 (03) :1854-1864
[4]  
CORNELISSE LE, 1994, HEARING J, V47, P23
[5]  
De Gennaro S, 1986, J Rehabil Res Dev, V23, P17
[6]   ALLOWING FOR REAL EAR VENTING EFFECTS WHEN SELECTING THE COUPLER GAIN OF HEARING-AIDS [J].
DILLON, H .
EAR AND HEARING, 1991, 12 (06) :406-416
[7]   Compression? Yes, but for low or high frequencies, for low or high intensities, and with what response times? [J].
Dillon, H .
EAR AND HEARING, 1996, 17 (04) :287-307
[8]  
Dillon H., 1998, Australian Journal of Audiology, V20, P33
[9]   The national acoustic laboratories' procedure for selecting the saturation sound pressure level of hearing aids: Theoretical derivation [J].
Dillon, H ;
Storey, L .
EAR AND HEARING, 1998, 19 (04) :255-266
[10]  
DILLON H, 1988, HDB HEARING AID AMPL, P121