Reliability of vessel diameter measurements with a retinal oximeter

被引:48
作者
Blondal, Renata [2 ]
Sturludottir, Margret Kara [2 ]
Hardarson, Sveinn Hakon [1 ,2 ]
Halldorsson, Gisli Hreinn [2 ]
Stefansson, Einar [1 ]
机构
[1] Univ Iceland, Landspitali Natl Univ Hosp Iceland, Dept Ophthalmol, IS-101 Reykjavik, Iceland
[2] Oxymap Ehf, Reykjavik, Iceland
关键词
Retina; Artery; Vein; Arteriole; Venule; Diameter; Reliability; Oxygen; Oximeter; Oximetry; PROLIFERATIVE DIABETIC-RETINOPATHY; BLOOD-FLOW; PANRETINAL PHOTOCOAGULATION; FUNDUS PHOTOGRAPHS; MICRODENSITOMETRIC METHODS; VASCULAR REACTIVITY; ARTERIOLAR DIAMETER; LASER TREATMENT; GLAUCOMA EYES; PLUS DISEASE;
D O I
10.1007/s00417-011-1680-2
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The purpose of this study was to test the reliability of vessel diameter measurements with a newly developed retinal oximeter. Twelve healthy individuals participated in the study. Retinal images were taken with the Oxymap Retinal Oximeter. Diameters of retinal vessels were measured automatically with the Oxymap Analyzer software. Repeated measurements on the same vessel segments were compared. The automatic measurements were also compared with semi-automatic measurements with a plug-in for the ImageJ software. Variance coefficient (standard deviation/mean) from the repeatability test was 2.8 and 4.0% for first- and second-degree venules and 3.5 and 5.4% for first- and second-degree arterioles, respectively. ImageJ measured larger mean diameters than Oxymap Analyzer in all cases. Means of differences were 5.1 +/- 2.2, 2.9 +/- 1.3, and 2.7 +/- 1.6 pixels for first-, second-, and third-degree venules and 3.1 +/- 1.2, 2.7 +/- 0.9, and 2.9 +/- 1.4 pixels for first-, second-, and third-degree arterioles. Vessel diameter measurements with the oximeter are repeatable and comparison with an established method demonstrates a relatively stable offset where the standard deviation of the difference is rather small. Different definitions of vessel borders may be the cause of this difference.
引用
收藏
页码:1311 / 1317
页数:7
相关论文
共 53 条
[1]  
Arnarsson A, 2000, INVEST OPHTH VIS SCI, V41, P877
[2]   Oximetry of retinal vessels by dual-wavelength imaging: calibration and influence of pigmentation [J].
Beach, JM ;
Schwenzer, KJ ;
Srinivas, S ;
Kim, D ;
Tiedeman, JS .
JOURNAL OF APPLIED PHYSIOLOGY, 1999, 86 (02) :748-758
[3]  
Bland M., 1995, INTRO MED STAT, V2nd
[4]   MEASUREMENT OF VESSEL WIDTH ON FUNDUS PHOTOGRAPHS [J].
BRACHER, D ;
DOZZI, M ;
LOTMAR, W .
ALBRECHT VON GRAEFES ARCHIV FUR KLINISCHE UND EXPERIMENTELLE OPHTHALMOLOGIE, 1979, 211 (01) :35-48
[5]  
BRINCHMANN-HANSEN O, 1989, Eye (London), V3, P768
[6]  
BRINCHMANNHANSEN O, 1990, AVIAT SPACE ENVIR MD, V61, P112
[7]   Acute effects of oxygen and carbon dioxide on retinal vascular network geometry in hypertensive and normotensive subjects [J].
Chapman, N ;
Haimes, G ;
Stanton, AV ;
Thom, SA ;
Hughes, AD .
CLINICAL SCIENCE, 2000, 99 (06) :483-488
[8]   EVALUATION OF MICROMETRIC AND MICRODENSITOMETRIC METHODS FOR MEASURING THE WIDTH OF RETINAL VESSEL IMAGES ON FUNDUS PHOTOGRAPHS [J].
DELORI, FC ;
FITCH, KA ;
FEKE, GT ;
DEUPREE, DM ;
WEITER, JJ .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1988, 226 (04) :393-399
[9]  
FEKE GT, 1982, OPHTHALMOLOGY, V89, P757
[10]  
FEKE GT, 1989, INVEST OPHTH VIS SCI, V30, P58