Study of the impact of the object's location on the image on the results of distance measurement by the DFD technique

被引:0
作者
Palys, Tomasz [1 ]
Arciuch, Artur [1 ]
机构
[1] Mil Univ Technol, Inst Teleinformat & Automat, Urbanowicza Str 2, PL-00908 Warsaw, Poland
来源
13TH CONFERENCE ON INTEGRATED OPTICS: SENSORS, SENSING STRUCTURES, AND METHODS | 2018年 / 10830卷
关键词
Distance Measurement; Depth Form Defocus; Information Extraction; FOCUS;
D O I
10.1117/12.2503570
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The aim of the paper is to determine the impact of the object's location in relation to camera's optical axis on the distance measurement using the method belonging to the Depth From Defocus (DFD). The method has been developed to determining the shape of flaccid membrane used in Ventricular Assist Device (VAD). The tests were carried out in laboratory conditions. The research were carried out using a fixed miniature camera and a fixed focal length lens. The camera observed a marker with a diameter of 0.003 m, in various position relative to the optical axis perpendicular to the plane of the marker. The position of the marker was changed in three dimensions, with a 0.001 m step, in the range from 0 m to 0.010 m in directions according to X and Y. The Z direction was researched in the range from 0.100 m to 0.140 m with a step equal 0.001 m. The measurable effects of work are the recommendations regarding the correction of the distance measurement depending on the marker shift relative to the camera optical axis.
引用
收藏
页数:4
相关论文
共 1 条
  • [1] A new technique for distance measurement of between vehicles to vehicles by plate car using image processing
    Phelawan, J.
    Kittisut, P.
    Pornsuwancharoen, N.
    ISEEC, 2012, 32 : 348 - 353