Suprasternal Doppler ultrasound for assessment of stroke distance

被引:12
|
作者
Childs, C
Goldring, S
Tann, W
Hillier, VF
机构
[1] Booth Hall Childrens Hosp, Reg Paediat Burns Unit, Manchester M9 7AA, Lancs, England
[2] Univ Manchester, MRC, Trauma Grp, Manchester, Lancs, England
[3] Univ Manchester, Dept Med Biophys, Manchester, Lancs, England
关键词
Doppler ultrasound; stroke distance; measurement error;
D O I
10.1136/adc.79.3.251
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
An assessment of a non-invasive technique for measurement of stroke distance was made using a portable Doppler ultrasound machine. The aim was to determine the measurement error of repeated stroke distance measurements (within-observer variability) and to assess measurement agreement between two operators (between-observer variability), The measurement error (within-observer variability) for both operators was similar at approximately 2 cm. :However, the measurements of the two operators (between-observer variability) did not agree well. Using the mean (SD) of three readings by each operator, the mean difference between the operators was -0.21 cm (1.96) giving a 95% confidence interval for the differences of -4.0 to +3.6 cm. There were significant positive and negative correlations between stroke distance and a variety of variables (age, height, weight, heart rate), but the relations were weak. The results indicate that the Doppler ultrasound technique for measurement of stroke distance would best be used to study trend changes in an individual patient, or subject, by a single operator.
引用
收藏
页码:251 / 255
页数:5
相关论文
共 50 条
  • [1] Effect of table tilt on maternal cardiac output using suprasternal Doppler ultrasound
    Selvan, PT
    Bray, J
    Fernando, R
    Columb, MO
    BRITISH JOURNAL OF ANAESTHESIA, 2005, 94 (03) : 399P - 399P
  • [2] Color Doppler Ultrasound Imaging in the Assessment of Iliac Endofibrosis
    D'Abate, Fabrizio
    Paraskevas, Kosmas I.
    Oates, Crispian
    Palfreeman, Roger
    Hinchliffe, Robert J.
    ANGIOLOGY, 2017, 68 (03) : 225 - 232
  • [3] Risk stratification after acute myocardial infarction by Doppler stroke distance measurement
    Trent, RJ
    Rawles, JM
    HEART, 1999, 82 (02) : 187 - 191
  • [4] An assessment of the accuracy of renal blood flow estimation by Doppler ultrasound
    Wan, Li
    Yang, Natalie
    Hiew, Chee-Yan
    Schelleman, Anthony
    Johnson, Lynne
    May, Clive
    Bellomo, Rinaldo
    INTENSIVE CARE MEDICINE, 2008, 34 (08) : 1503 - 1510
  • [5] A DOPPLER ULTRASOUND DEVICE FOR ASSESSMENT OF BLOOD-VOLUME FLOW
    MOSER, U
    VIELI, A
    SCHUMACHER, P
    PINTER, P
    BASLER, S
    ANLIKER, M
    ULTRASCHALL IN DER MEDIZIN, 1992, 13 (02): : 77 - 79
  • [6] An automated carotid pulse assessment approach using Doppler ultrasound
    Yu, Alfred C. H.
    Cohen-Solal, Eric
    Raju, Balasundar I.
    Ayati, Shervin
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (03) : 1072 - 1081
  • [7] NONINVASIVE DOPPLER ULTRASOUND TECHNIQUE FOR THE INVIVO ASSESSMENT OF AORTIC COMPLIANCE
    LEHMANN, ED
    GOSLING, RG
    FATEMILANGROUDI, B
    TAYLOR, MG
    JOURNAL OF BIOMEDICAL ENGINEERING, 1992, 14 (03): : 250 - 256
  • [8] Intraobserver variation in Doppler ultrasound assessment of pulmonary artery pressure
    Subhedar, NV
    Shaw, NJ
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1996, 75 (01): : F59 - F61
  • [9] An assessment of the accuracy of renal blood flow estimation by Doppler ultrasound
    Li Wan
    Natalie Yang
    Chee-Yan Hiew
    Anthony Schelleman
    Lynne Johnson
    Clive May
    Rinaldo Bellomo
    Intensive Care Medicine, 2008, 34 : 1503 - 1510
  • [10] Doppler Ultrasound for Heart Rate Assessment in a Porcine Model of Neonatal Asphyxia
    Morina, Nicolo
    Johnson, Peter A.
    O'Reilly, Megan
    Lee, Tze-Fun
    Yaskina, Maryna
    Cheung, Po-Yin
    Schmolzer, Georg M.
    FRONTIERS IN PEDIATRICS, 2020, 8