Design and Control of a Pneumatic Artificial Muscle Actuated Above-Knee Prosthesis

被引:35
|
作者
Waycaster, Garrett [1 ]
Wu, Sai-Kit [1 ]
Shen, Xiangrong [1 ]
机构
[1] Univ Alabama, Dept Mech Engn, Tuscaloosa, AL 35487 USA
来源
JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME | 2011年 / 5卷 / 03期
关键词
LEVEL;
D O I
10.1115/1.4004417
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents the authors' investigation results of applying the pneumatic artificial muscle actuation to above-knee prostheses. As a well-known muscle actuator, the pneumatic artificial muscle actuator features a number of unique advantages, including high power density, and similar elastic characteristics to biological muscles. Despite multiple applications in related areas, the application of pneumatic artificial muscle in above-knee prostheses has not been explored. Inspired by this fact, the research presented in this paper aims to develop a pneumatic artificial muscle-actuated above-knee prosthesis, with three specific objectives: (1) demonstrate the pneumatic artificial muscle actuation's capability in generating sufficient torque output to meet the locomotive requirements; (2) develop an effective control approach to enable the restoration of locomotive functions; (3) conduct preliminary testing of the prosthesis prototype on a healthy subject through a specially designed able-body adaptor. In the prosthesis design, an agonist-antagonist layout is utilized to obtain a bidirectional motion. To minimize the radial profile, an open-frame structure is used, with the purpose of allowing the expansion of the muscle actuators into the center space without interference. Also, the muscle actuator parameters are calculated to provide sufficient torque capacity (up to 140 N m) to meet the requirements of level walking. According to this design, the fabricated prototype weighs approximately 3 kg, with a range of motion of approximately 100 degrees. For the control of the prosthesis, a model-based torque control algorithm is developed based on the sliding mode control approach, which provides robust torque control for this highly nonlinear system. Combining this torque control algorithm with an impedance-based torque command generator (higher-level control algorithm), the fabricated prosthesis prototype has demonstrated a capability of providing a natural gait during treadmill walking experiments. [DOI: 10.1115/1.4004417]
引用
收藏
页数:9
相关论文
共 50 条
  • [1] A PNEUMATIC ARTIFICIAL MUSCLE ACTUATED ABOVE-KNEE PROSTHESIS
    Waycaster, Garrett
    Wu, Sai-Kit
    Shen, Xiangrong
    PROCEEDINGS OF THE ASME DYNAMIC SYSTEMS AND CONTROL CONFERENCE 2010, VOL 1, 2010, : 793 - 800
  • [2] A PNEUMATIC ARTIFICIAL MUSCLE ACTUATED KNEE PROSTHESIS
    Waycaster, Garrett
    Wu, Sai-Kit
    Shen, Xiangrong
    PROCEEDINGS OF THE ASME SUMMER BIOENGINEERING CONFERENCE, 2010, 2010, : 735 - 736
  • [3] ANALYSIS OF A NEW POLYCENTRIC ABOVE-KNEE PROSTHESIS WITH A PNEUMATIC SWING PHASE-CONTROL
    PATIL, KM
    CHAKRABORTY, JK
    JOURNAL OF BIOMECHANICS, 1991, 24 (3-4) : 223 - 233
  • [4] Tracking method for the control of above-knee prosthesis
    Kalanovich, V.D.
    Popovich, B.D.
    Avtomatika i Telemekhanika, 1993, (01): : 127 - 141
  • [5] CONTROL ASPECTS OF ACTIVE ABOVE-KNEE PROSTHESIS
    POPOVIC, D
    TOMOVIC, R
    TEPAVAC, D
    SCHWIRTLICH, L
    INTERNATIONAL JOURNAL OF MAN-MACHINE STUDIES, 1991, 35 (06): : 751 - 767
  • [6] TRACKING METHOD FOR THE CONTROL OF THE ABOVE-KNEE PROSTHESIS
    KALANOVIC, VD
    POPOVIC, DB
    AUTOMATION AND REMOTE CONTROL, 1993, 54 (01) : 119 - 132
  • [7] Knee and Ankle Powered Above-Knee Prosthesis Design and Development
    Rupar, Miljan
    Vucina, Adisa
    Dedic, Remzo
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 2, 2019, 68 (02): : 625 - 629
  • [8] MODULAR ABOVE-KNEE PROSTHESIS
    SOKOLOW, J
    GRYNBAUM, BB
    ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1973, 54 (06): : 278 - 280
  • [9] ALIGNMENT OF THE ABOVE-KNEE PROSTHESIS
    FOORT, J
    PROSTHETICS AND ORTHOTICS INTERNATIONAL, 1979, 3 (03) : 137 - 139
  • [10] Parametric Design and Hybrid Fabrication of Above-Knee Prosthesis
    Amrutsagar, Lalitrao
    Parit, Gaurav
    Ghyar, Rupesh
    Bhallamudi, Ravi
    INDIAN JOURNAL OF ORTHOPAEDICS, 2020, 54 (03) : 381 - 390