Undesirable nuclear reactions and induced radioactivity as a result of the use of the high-energy therapeutic beams generated by medical linacs

被引:35
作者
Konefal, Adam [1 ]
Polaczek-Grelik, Kinga [1 ]
Zipper, Wiktor [1 ]
机构
[1] Silesian Univ, Inst Phys, Dept Nucl Phys & Its Applicat, PL-40007 Katowice, Poland
关键词
D O I
10.1093/rpd/ncm318
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
In this paper, the problem of undesirable photonuclear, electronuclear and neutron capture reactions taking place in the treatment room during emission of the typical high-energy therapeutic beams from two different medical accelerators, i.e. Primus Siemens and Varian Clinac-2300, is presented. The radioisotopes W-187, Mn-56, Al-28, Ni-57, Cl-38, Co-57 and Au-19 and the neutron activation of H-1 were identified as a consequence of these reactions. Moreover, the increased photon fluence rate behind the door of the accelerator bunker in the operator console room was observed during emission of the 20 MV X-rays from the Varian Clinac-2300 as well as in the case of the 15 MV X-ray beam from the Primus Siemens. No increased radiation was observed during the 6 MV X-ray beam emission. The performed measurements produced evidences on the presence of neutrons in the operator console room during emission of the 15 MV X-ray beam from the Primus Siemens as well as the 20 MV X-rays and the 22 MeV electrons from the Varian Clinac-2300 accelerator.
引用
收藏
页码:133 / 145
页数:13
相关论文
共 17 条
[1]  
[Anonymous], 1964, NUCL RAD DETECTION
[2]  
Beckurc K, 1968, NIEJTRONNAJA FIZIKA
[3]  
DIETRICH SS, 1988, ATOM DATA NUCL DATA, V38, P199, DOI [10.1016/0092-640X(88)90033-2, 10.1016/0092-640X(75)90010-8]
[4]   A study of neutron spectra from medical linear accelerators [J].
Facure, A ;
Falcao, RC ;
Silva, AX ;
Crispim, VR ;
Vitorelli, JC .
APPLIED RADIATION AND ISOTOPES, 2005, 62 (01) :69-72
[5]  
Firestone R. B., 1997, Table of Isotopes
[6]  
Giacri ML, 2006, NUCL SCI ENG, V153, P1
[7]  
Gudowska I., 1996, Nukleonika, V41, P105
[8]  
Konefal A, 2005, NUKLEONIKA, V50, P73
[9]  
Konefal A, 2001, POL J MED PHYS, V7, P291
[10]  
KONEFAL A, 2006, POL J ENVIRON STUD, V15, P176